Permobil F5Recalls & reported problems
Permobil, Inc. · F5 Corpus · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Permobil F5 Corpus is a powered wheelchair. Reports mostly describe falls from the seat after sudden stops, unexpected movement or loss of control on ramps and curbs, often mentioning bone fractures. FDA records include 26 reports naming this device identifier, including 1 death and 19 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 1
- Injuries reported
- 20
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 17330818229844), received between Oct 20, 2016 and Sep 24, 2026 — classified as 1 death, 20 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem12
- Mechanical Problem6
- Detachment of Device or Device Component4
- Unintended Movement3
- Device Fell1
- Device Operational Issue1
- Device Tipped Over1
- Electrical /Electronic Property Problem1
Patient problems
- Bone Fracture(s)14
- No Known Impact Or Consequence To Patient3
- Hip Fracture2
- Limb Fracture2
- Multiple Fractures2
- Pressure Sore/Ulcer2
- Abrasion1
- Bruise/Contusion1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
InjurySep 24, 2026Mechanical Problem; Unintended System Motion; Unintended Movement
Reports the end-user had used the chair's tilt and recline functions for positional relief. As the end-user attempted to return the chair to an upright position, an inadvertent activation/mode-selection sequence allegedly occurred. While they attempted to return the chair to seating-function mode, a blanket or shirt sleeve allegedly contacted or became caught on the joystick, causing the wheelchair to travel backward and resulting in the end-users' head and neck striking a wall.
Patient problems: Nervous System Injury; Spinal Cord Injury
MAUDE report 50889313 ↗InjuryJul 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RECEIVED REPORT CLAIMING AS THE END-USER WAS EXITING A HOME VIA A RAMP THAT TRANSITIONED ONTO GRAVEL PATHWAY. REPORTS CLAIMS THE DEVICE AT SOME POINT LOST TRACTION WHICH CAUSED THE END-USER TO LOSE CONTROL WHERE THE DEVICE REPORTEDLY SLID DOWN AN EMBANKMENT INTO A HOLE. REPORT INDICATES THE END-USER REMAINED SEATED IN THE DEVICE AND IT REMAINED UPRIGHT THROUGHOUT THE EVENT, BUT THE DROP INTO THE HOLE REPORTEDLY RESULTED IN INJURIES THAT REQUIRED MEDICAL INTERVENTION TO ADDRESS.
Patient problems: Bone Fracture(s); Multiple Fractures
MAUDE report 22630684 ↗InjurySep 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REPORTS WHILE LEANING FORWARD IN SEATING TO PICK UP THEIR GLASSES, THE END-USER INADVERTENTLY HIT THE "TRIGGER" WHICH CAUSED THE DEVICE TO MOVE ALLOWING THE END-USER TO LOSE POSITIONING AND FALL OUT OF SEATING RESULTING IN AN INJURY.
Patient problems: Bone Fracture(s); Hip Fracture
MAUDE report 20176118 ↗InjuryJun 10, 2024Mechanical Problem; Device Tipped Over; Unintended Movement; Mechanical Problem; Device Tipped Over; Unintended Movement
RECEIVED REPORT OF WHILE THE END-USER WAS STOPPED IN FRONT OF A RESTAURANT ENTRANCE, THE F5 CORPUS DEVICE REPORTEDLY STARTED DRIVING IN REVERSE WITH NO TOUCHING OF JOYSTICK. THIS REPORTEDLY RESULTED IN THE DEVICE'S CASTER WHEELS DROPPING OFF A CURB EDGE WHICH ALLOWED THE DEVICE TO LOSE STABILITY AND FALL TO THE SIDE RESULTING IN INJURY TO THE END-USER.
Patient problems: Bone Fracture(s); Limb Fracture
MAUDE report 19497296 ↗InjuryMar 1, 2024Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
REPORT RECEIVED WHILE END-USER WAS DRIVING THEIR F5 CORPUS, THE FOOTPLATES HIT THE GROUND WHICH CAUSED A SUDDEN STOPPAGE OF THE DEVICE. THIS REPORTEDLY CAUSED THE END-USER TO LOSE POSITIONING AND FALL FROM THE SEATING TO THE GROUND, RESULTING IN AN INJURY REQUIRING MEDICAL INTERVENTION TO ADDRESS.
Patient problems: Insufficient Information
MAUDE report 18818261 ↗MalfunctionSep 1, 2023Electrical /Electronic Property Problem; Unexpected Shutdown; Electrical /Electronic Property Problem; Unexpected Shutdown
RECEIVED REPORT CLAIMING WHILE THE END-USER WAS DRIVING OUT OF DOORS IN THEIR COMMUNITY, THE POWER WHEELCHAIR ALLEGEDLY CAME TO AN ABRUPT STOP WHICH FORCED THE END-USER TO LOSE POSITIONING AND FALL FROM THE SEATING TO THE GROUND. THIS ACTION REPORTEDLY RESULTED IN MINOR INJURIES THAT DID NOT NECESSITATE THE NEED FOR MEDICAL INTERVENTION.
Patient problems: Abrasion; Bruise/Contusion; Swelling/ Edema
MAUDE report 17673544 ↗InjurySep 27, 2022Mechanical Problem; Unintended Movement; Device Fell; Mechanical Problem; Unintended Movement; Device Fell
RECEIVED REPORT CLAIMING AS THE END-USER WAS EXITING THEIR TRANSPORT VEHICLE VIA THE INCORPORATED RAMP, ONE OF THE DRIVE MOTORS WAS ALLEGED TO HAVE "LOCKED-UP" WHICH REPORTEDLY CAUSED THE DEVICE TO TURN AND DRIVE OFF THE SIDE OF THE RAMP. THE ENSUING FALL RESULTED IN AN INJURY TO THE END-USER REQUIRING MEDICAL INTERVENTION TO ADDRESS.
Patient problems: Bone Fracture(s); Limb Fracture
MAUDE report 15490701 ↗InjuryAug 31, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
RECEIVED REPORT CLAIMING DUE TO THE INABILITY TO REPLACE DEFECTIVE COMPONENT ON THE WHEELCHAIR IN A TIMELY MANNER, THE END-USER DEVELOPED A PRESSURE ULCER REQUIRING HOSPITALIZATION TO ADDRESS.
Patient problems: Pressure Sore/Ulcer; Tissue Breakdown
MAUDE report 15329226 ↗InjuryJan 7, 2021Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
RECEIVED REPORT CLAIMING AS END-USER WAS IN PROCESS OF PERFORMING A STAND FUNCTION, THEY LOST THEIR POSITIONING AND FELL FORWARD OUT OF THE SEATING TO THE FLOOR RESULTING IN SERIOUS INJURIES REQUIRING MEDICAL INTERVENTION.
Patient problems: Bone Fracture(s); Multiple Fractures; Skull Fracture
MAUDE report 11130850 ↗InjuryJun 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RECEIVED REPORT CLAIMING END-USER HAVING RECEIVED TISSUE DAMAGE TO THEIR LEFT LEG, REPORTEDLY DUE TO THEIR LEG PRESSING INTO THE THIGH SUPPORT.
Patient problems: Pressure Sore/Ulcer
MAUDE report 10202617 ↗InjuryFeb 13, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RECEIVED REPORT CLAIMING END-USER HAVING INADVERTENTLY MANEUVERED THEIR DEVICE OVER THE TOP EDGE OF A STAIRWELL FALLING APPROXIMATELY 15 STEPS TO THE BOTTOM. END-USER REPORTED TO HAVE SUFFERED SIGNIFICANT INJURIES REQUIRING MEDICAL ATTENTION.
Patient problems: Bone Fracture(s)
MAUDE report 9708586 ↗DeathFeb 5, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RECEIVED REPORT CLAIMING AS THE END-USER WAS ATTEMPTING TO LIGHT A WOOD BURNING STOVE WITH A TORCH, THEY INADVERTENTLY DROPPED THE TORCH IN THEIR LAP SUBSEQUENTLY IGNITING THEIR CLOTHING. END-USER WAS REPORTED TO HAVE SUSTAINED GRAVE INJURIES AS A RESULT.
Patient problems: Death
MAUDE report 9673483 ↗InjuryOct 14, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RECEIVED REPORT CLAIMING WHILE THE END-USER WAS AT A SCHOOL/DAY PROGRAM, THEY INITIATED A STANDING SEQUENCE. IT WAS REPORTED DURING THIS FUNCTION, THE END-USER SUFFERED A FRACTURE TO THEIR RIGHT LEG.
Patient problems: Bone Fracture(s)
MAUDE report 9188715 ↗InjurySep 13, 2019Mechanical Problem; Structural Problem; Sharp Edges; Mechanical Problem; Structural Problem; Sharp Edges
RECEIVED REPORT CLAIMING WHILE THE END-USER WAS PERFORMING A TRANSFER, THEIR FOOT SLIPPED OFF THE FOOTPLATE ASSEMBLY WHICH REPORTEDLY CAUSED THE TOE NEXT TO THEIR LARGE TOE TO BECOME CUT. IT WAS REPORTED THE END-USER SOUGHT MEDICAL INTERVENTION TO RESOLVE.
Patient problems: Skin Tears
MAUDE report 9035533 ↗InjuryJun 5, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RECEIVED REPORT CLAIMING AS END-USER WAS DRIVING ALONG A SIDEWALK, THEY INADVERTENTLY HIT A BUMP CAUSING THEIR CHEST STRAP TO BECOME UNDONE. THIS REPORTEDLY CAUSED THEM TO LOSE POSITIONING AND SLUMP FORWARD, FORCING THE END-USER TO LOSE CONTROL OF THE DEVICE AND DRIVE OFF THE SIDEWALK DOWN AN EMBANKMENT. REPORTS INDICATE THE END-USER SUFFERED AN INJURY REQUIRING HOSPITALIZATION.
Patient problems: Bone Fracture(s); Hip Fracture
MAUDE report 8670763 ↗MalfunctionNov 19, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
REPORTS ARE THE LEFT DRIVE TIRE ASSEMBLY INCLUDING MOTOR HUB DETACHED FROM THE WHEELCHAIR. NO HARM WAS SUSTAINED TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8086457 ↗MalfunctionSep 25, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
REPORTS THAT THE LEFT DRIVE WHEEL ASSEMBLY DETACHED FROM THE WHEELCHAIR. IT WAS DISCOVERED THAT THE MAIN HUB BOLT HAD BACKED OUT. NO INJURIES HAVE OCCURRED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7905143 ↗InjuryJun 6, 2018Mechanical Problem; Device Operational Issue; Mechanical Problem; Device Operational Issue
RECEIVED REPORT THAT AS END-USER WAS IN PROCESS OF PERFORMING A TRANSFER OUT OF THE SEATING, THE DEVICE ALLEGEDLY ROLLED AWAY CAUSING THE END-USER TO LOSE THEIR BALANCE AND FALL TO THE GROUND RESULTING IN AN INJURY REQUIRING MEDICAL ATTENTION.
Patient problems: Wound Dehiscence
MAUDE report 7573565 ↗MalfunctionJun 6, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6), THE END USER WAS ON A MORNING DRIVE WITH HER WHEELCHAIR, WHEN THE FRONT LEFT DRIVE TIRE DETACHED FROM THE CHASSIS. NO INJURY WAS REPORTED TO HAVE OCCURRED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7573423 ↗InjuryApr 30, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RECEIVED REPORT THAT WHILE END-USER WAS DRIVING THE DEVICE, THEY HIT A BUMP IN THE TERRAIN WHICH CAUSED THEIR RIGHT FOOT TO COME OFF THE FOOTPLATE. WHEN THIS OCCURRED, IT WAS REPORTED THE END-USERS FOOT ROLLED UNDER THE RIGHT DRIVE WHEEL CAUSING THEM TO SUFFER A FRACTURE TO THEIR RIGHT TIBIA.
Patient problems: Bone Fracture(s)
MAUDE report 7472467 ↗InjuryJan 17, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6), THE WHEELCHAIR WAS DELIVERED TO THE CUSTOMER. DURING SETUP AND INITIAL USE OF THE STAND FEATURE THE PATIENT WAS ALLEGEDLY INJURED. THE FOLLOWING DAY THE PATIENT WHEN TO THE HOSPITAL WHERE BROKEN RIGHT FEMUR BONE WAS CONFIRMED.
Patient problems: Bone Fracture(s)
MAUDE report 7202373 ↗InjuryDec 26, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RECEIVED REPORT OF END-USER ALLEGING HAVING SUFFERED A FRACTURED TIBIA AFTER PERFORMING A STAND FUNCTION WITH THE CHAIR.
Patient problems: Bone Fracture(s)
MAUDE report 7146477 ↗InjuryDec 13, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT STATED THAT HE WAS TRAVELING FULL SPEED AND TO AVOID THE DOG TOOK A HARD LEFT TURN AND FELL OUT OF THE WHEELCHAIR, HAVING NOT BEEN WEARING A POSITIONAL BELT. END USER STATED THAT THE DEVICE DID NOT TIP OVER OR SUSTAIN DAMAGES DURING THE EVENT. THE PATIENT IS REPORTED TO HAVE BROKE HIS LEG.
Patient problems: Bone Fracture(s)
MAUDE report 7115281 ↗InjuryMar 16, 2017Inadequate User Interface; Inadequate User Interface
THE HUSBAND WAS MAKING ADJUSTMENTS TO THE WHEELCHAIR ONE MORNING, SO THE PATIENT COULD BE FIT TO "STAND" AND USE THE WHEELCHAIR. THE HUSBAND USED THE POWER ELEVATING LEGREST FEATURE TO GET PROPER PLACEMENT FOR PATIENTS FEET AND LEGS. THE HUSBAND THEN EQUIPPED THE PATIENT WITH ANKLE HUGGERS AND KNEE BLOCK PADS AND INSTALLED THE CHEST ROLL. IT WAS AT THIS TIME, WHEN THE PATIENT STARTED SCREAMING. THE PATIENT HAD ACCIDENTALLY RESTED HER HAND ON THE JOYSTICK CONTROLS ACTIVATING THE POWER LEGREST FUNCTION FROM SEATING MODE, WHICH WAS LEFT ON BY THE HUSBAND. THIS EVENT OVER EXTENDED HER LEGS CAUSING A HEEL FRACTURE. NOTE: INCIDENT DATE HAPPENED AROUND (B)(6) 2017, BUT AN EXACT DATE COULDN'T BE CONFIRMED.
Patient problems: Bone Fracture(s)
MAUDE report 6411828 ↗InjuryOct 20, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DEALER REPORTS HAVING BEEN CONTACTED BY END USERS FATHER REPORTING END USER HAVING BEEN TREATED FOR A FRACTURED LEFT LEG. FAMILY ALLEGES THIS INJURY OCCURED WHILE END USER WAS USING THE CHAIRS STAND FEATURE.
Patient problems: Bone Fracture(s)
MAUDE report 6044623 ↗
FDA clearance and classification
Cleared through 510(k) K143014 on Mar 25, 2015 (decision: Substantially Equivalent). Submitted by Permobil AB as “F5”. The FDA review took 156 days. FDA record ↗
The FDA regulates product code ITI (wheelchair, powered) under 21 CFR 890.3860 as a Class II device. 21 CFR 890.3860 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. FDA interprets mobility to persons restricted to a sitting position to mean the device type is intended to provide mobility to individuals who have mobility impairments and require an assistive device for mobility. NOTE: This type of device is not intended to climb stairs (see product code: IMK).
Common questions about the Permobil F5
Has the Permobil F5 been recalled?
No FDA recall record in our data names this exact device identifier (DI 17330818229844), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Permobil F5?
The FDA's MAUDE database lists 25 adverse event reports for this device, received between Oct 20, 2016 and Sep 24, 2026. In those reports, the most common specific device problems are “Mechanical Problem” (6), “Detachment of Device or Device Component” (4) and “Unintended Movement” (3); the most common patient problems are “Bone Fracture(s)” (14), “Hip Fracture” (2) and “Limb Fracture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Permobil F5?
1 report is classified as a death, 20 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Permobil F5 FDA approved?
It was cleared through 510(k) K143014 on Mar 25, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Permobil F5, and is it still sold?
The FDA's device database lists Permobil, Inc. as the company responsible for it (brand Permobil, model F5). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Wheelchair, Powered devices
- SmartDrive® MX2+Max Mobility, LLC · Wheelchair, Powered858 reports
- Permobil M3Permobil, Inc. · Wheelchair, Powered1 recall60 reports
- TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx TDXSP2-MCGInvacare Corporation · Wheelchair, powered2 recalls2 reports
- TDX SP2 with LiNX Electronics TDXSP2Invacare Corporation · Wheelchair, powered2 recalls0 reports
- Invacare® TDX SP2 with Captain's Seat TDXSP2VInvacare Corporation · Wheelchair, powered2 recalls0 reports
- Q6 Edge 3Pride Mobility Products Corp · Wheelchair, Powered4 deaths87 reports
- Q6 Edge 2.0 Q6 Edge 2.0 / Q6 Edge 2.0 with iLevelPride Mobility Products Corp · Wheelchair, Powered1 death83 reports
- Permobil M5Permobil, Inc. · Wheelchair, Powered1 recall7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.