Change Healthcare Stratus Imaging PACS 1.4.0Recalls & reported problems
Change Healthcare Canada Company · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 17540262160093), received between Nov 22, 2023 and Nov 22, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Application Program Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Nov 1, 2023Class IIOpen, Classified
Under certain circumstances, information from HL7 messages received by Stratus PACS/Imaging Share systems has been incorrectly associated with another patient's study.
Z-0551-2024 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Application Program Problem1
- Computer Software Problem1
- Data Problem1
- Patient Data Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 22, 2023Computer Software Problem; Application Program Problem; Data Problem; Patient Data Problem; Computer Software Problem; Application Program Problem; Data Problem; Patient Data Problem
UNDER CERTAIN CIRCUMSTANCES, INFORMATION FROM HL7 MESSAGES RECEIVED BY THE STRATUS PACS/ IMAGING SHARE SYSTEM HAVE BEEN INCORRECTLY ASSOCIATED WITH ANOTHER PATIENT'S STUDY WHICH MAY RESULT IN THE FOLLOWING POTENTIAL SCENARIOS: 1. A SUBSET OF PATIENT STUDY METADATA (PATIENT NAME, MRN, ACCESSION NUMBER, DOB, GENDER, STUDY PROCEDURE) HAS BEEN UPDATED WITH ANOTHER PATIENT'S METADATA (PATIENT NAME, MRN, ACCESSION NUMBER, DOB, GENDER, STUDY PROCEDURE). THIS MAY RESULT IN PATIENT'S IMAGES BEING LABELED INCORRECTLY WITH ANOTHER PATIENT'S METADATA AND ASSOCIATED WITH THE WRONG PATIENT RECORD. 2. AS A RESULT OF INCORRECT IMAGE DEMOGRAPHICS, THE MISLABELED IMAGES MAY NOT BE ASSOCIATED WITH THE RECORD FOR THE PATIENT THAT HAD AN IMAGING STUDY. 3. AN ISSUED REPORT ON A STUDY THAT IS SENT TO THE STRATUS IMAGING SYSTEM VIA HL7, MAY GET ATTACHED TO ANOTHER PATIENT'S STUDY. THIS MAY RESULT IN A REPORT BEING ASSOCIATED WITH THE INCORRECT PATIENT STUDY. 4. AS A RESULT OF INCORRECT REPORT ASSOCIATION DESCRIBED ABOVE, SOME PATIENTS WILL HAVE MISSING REPORTS WITH NO REPORT BEING DISPLAYED IN THE STUDY. THE INCORRECT ASSOCIATION OF IMAGE AND REPORT METADATA ACROSS IMAGING STUDIES CAN OCCUR WITHIN YOUR FA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18193816 ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page