FastLoad™ CT Syringe Pack 017344Recalls & reported problems
Bracco Injeneering SA · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 17630039300309), received between Feb 13, 2017 and Oct 27, 2017 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (4 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 12, 2018Class IITerminated
Insufficient seal on the sterile barrier of the device, compromising the sterility of the device
Z-0654-2019 · Root cause: Packaging process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Integrity Problem4
Patient problems
- No Patient Involvement4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 27, 2017Material Integrity Problem; Material Integrity Problem
ON 28-SEP-2017 A HEALTHCARE PROFESSIONAL REPORTED TO BRACCO INJENEERING A DEVICE MALFUNCTION WITH THE USE OF FASTLOAD CT SYRINGE PACK. THE REPORT WAS FORWARDED TO BRACCO DRUG SAFETY ON 20-OCT-2017. ON (B)(6) 2017 AT A BEGINNING OF A CT INJECTION, THE SYRINGE BROKE. NO PATIENT OR USER/OPERATOR INJURY. MOST RECENT FOLLOW-UP INFORMATION INCORPORATED ABOVE INCLUDES: 06-NOV-2017: FOLLOW UP: 3500A MEDWATCH COMPLETED BY MANUFACTURER OF FASTLOAD CT SYRINGE PACK ATTACHED TO THE CASE. NO NEW INFORMATION WAS RECEIVED FOR THIS REPORT. 02-JUL-2018: BINJ RECEIVED FOLLOW-UP INFORMATION WHICH WAS SENT TO BRACCO ON 27-JUL-2018. DETAILED DESCRIPTION OF THE DEVICE MALFUNCTION WAS ADDED TO THE CASE. THIS CASE IS MEDICALLY CLOSED. UDI NUMBER: (B)(4). BRACCO WORLDWIDE CASE ID#: (B)(4). COMPANY COMMENT: BASED ON TWO DECADES OF EXPERIENCE WITH THE EMPOWER INJECTOR, IT IS THE OPINION OF BRACCO MEDICAL ADVISORY BOARD MEMBER THAT THE LIKELIHOOD OF A SYRINGE BURST IS VERY SMALL AND IF ONE SHOULD OCCUR, THE LIKELIHOOD OF SERIOUS HARM OR PERMANENT INJURY OR DISABILITY TO PATIENTS OR HEALTHCARE PERSONNEL IS EXTREMELY LOW WHEN COMPARED TO THE TOTAL OF DEVICES SOLD ON THE MARKET (2 PPM).
Patient problems: No Patient Involvement
MAUDE report 6983183 ↗MalfunctionApr 14, 2017Material Integrity Problem; Material Integrity Problem
ON 23-FEB-2017, A TECHNOLOGIST REPORTED TO BRACCO INJENEERING (BINJ) A DEVICE MALFUNCTION WITH THE USE OF FASTLOAD CTA SYRINGE. THE REPORT WAS FORWARDED TO BRACCO DRUG SAFETY ON 13-APR-2017. ON (B)(6) 2017, DURING A CT INJECTION, THE SYRINGE POPPED OUT OF THE CT INJECTOR AND LANDED ON A PILLOW NEXT TO THE PATIENT. NO PATIENT OR PERSON HAS BEEN REPORTED INJURED IN THIS EVENT. THERE WAS ONLY A POTENTIAL FOR INJURY. MOST RECENT FOLLOW-UP INFORMATION INCORPORATED ABOVE INCLUDES: 02-JUL-2018: BINJ RECEIVED FOLLOW-UP INFORMATION WHICH WAS SENT TO BRACCO ON 27-JUL-2018. DETAILED DESCRIPTION OF THE DEVICE MALFUNCTION WAS ADDED TO THE CASE. THIS CASE IS MEDICALLY CLOSED. BRACCO WORLDWIDE CASE ID#: (B)(4). (PREVIOUSLY REPORTED AS (B)(4). COMPANY COMMENT: BASED ON TWO DECADES OF EXPERIENCE WITH THE EMPOWER INJECTOR, IT IS THE OPINION OF BRACCO MEDICAL ADVISORY BOARD MEMBER THAT THE LIKELIHOOD OF A SYRINGE BURST IS VERY SMALL AND IF ONE SHOULD OCCUR, THE LIKELIHOOD OF SERIOUS HARM OR PERMANENT INJURY OR DISABILITY TO PATIENTS OR HEALTHCARE PERSONNEL IS EXTREMELY LOW WHEN COMPARED TO THE TOTAL OF DEVICES SOLD ON THE MARKET (2 PPM).
Patient problems: No Patient Involvement
MAUDE report 6494503 ↗MalfunctionFeb 15, 2017Material Integrity Problem; Material Integrity Problem
ON 16-JAN-2017, A TECHNOLOGIST REPORTED TO BRACCO INJENEERING (BINJ) A DEVICE MALFUNCTION WITH THE USE OF FASTLOAD CTA SYRINGE. THE REPORT WAS FORWARDED TO BRACCO DRUG SAFETY ON 10-FEB-2017. ON 16-JAN-2017, AFTER 10 ML OF A 50 ML ISOVUE CONTRAST MEDIA INJECTION, THE CM SYRINGE POPPED OUT OF THE INJECTOR AND LANDED ON THE TABLE. NO PATIENT OR PERSON HAS BEEN REPORTED INJURED IN THIS EVENT. THERE WAS ONLY A POTENTIAL FOR INJURY. MOST RECENT FOLLOW-UP INFORMATION INCORPORATED ABOVE INCLUDES: 02-JUL-2018: BINJ RECEIVED FOLLOW-UP INFORMATION WHICH WAS SENT TO BRACCO ON 27-JUL-2018. DETAILED DESCRIPTION OF THE DEVICE MALFUNCTION, RELEVANT TEST (CAPA) WAS ADDED TO THE CASE. THIS CASE IS MEDICALLY CLOSED. BRACCO WORLDWIDE CASE ID#: (B)(4) (PREVIOUSLY REPORTED AS (B)(4). COMPANY COMMENT: BASED ON TWO DECADES OF EXPERIENCE WITH THE EMPOWER INJECTOR, IT IS THE OPINION OF BRACCO MEDICAL ADVISORY BOARD MEMBER THAT THE LIKELIHOOD OF A SYRINGE BURST IS VERY SMALL AND IF ONE SHOULD OCCUR, THE LIKELIHOOD OF SERIOUS HARM OR PERMANENT INJURY OR DISABILITY TO PATIENTS OR HEALTHCARE PERSONNEL IS EXTREMELY LOW WHEN COMPARED TO THE TOTAL OF DEVICES SOLD ON THE MARKET (2 PPM).
Patient problems: No Patient Involvement
MAUDE report 6332861 ↗MalfunctionFeb 13, 2017Material Integrity Problem; Material Integrity Problem
ON (B)(6) 2017, A TECHNOLOGIST REPORTED TO BRACCO INJENEERING (BINJ) A DEVICE MALFUNCTION WITH THE USE OF THE FASTLOAD CT SYRINGE. THE REPORT WAS FORWARDED TO BRACCO DRUG SAFETY ON 09-FEB-2017. ON (B)(6) 2017, DURING AN INJECTION IN THE MORNING, THE TECH WAS DOING A ISOVUE 370 50ML INJECTION AND THE CM SYRINGE POPPED OUT OF THE INJECTOR AND LANDED ON TOP OF THE PATIENT (BEFORE INJECTION OF CONTRAST AND SCAN). PRESSURE WAS NORMAL. NO PATIENT OR PERSON HAS BEEN REPORTED INJURED IN THIS EVENT. THERE WAS ONLY A POTENTIAL FOR INJURY. MOST RECENT FOLLOW-UP INFORMATION INCORPORATED ABOVE INCLUDES: 02-JUL-2018: BINJ RECEIVED FOLLOW-UP INFORMATION WHICH WAS SENT TO BRACCO ON 27-JUL-2018. DETAILED DESCRIPTION OF THE DEVICE MALFUNCTION WAS ADDED TO THE CASE. THIS CASE IS MEDICALLY CLOSED. FDA#: 3004753774-2017-00001 BRACCO WORLDWIDE CASE ID#: (B)(4). (PREVIOUSLY REPORTED AS (B)(4)). COMPANY COMMENT: BASED ON TWO DECADES OF EXPERIENCE WITH THE EMPOWER INJECTOR, IT IS THE OPINION OF BRACCO MEDICAL ADVISORY BOARD MEMBER THAT THE LIKELIHOOD OF A SYRINGE BURST IS VERY SMALL AND IF ONE SHOULD OCCUR, THE LIKELIHOOD OF SERIOUS HARM OR PERMANENT INJURY OR DISABILITY TO PATIENTS OR HEALTHCARE PERSONNEL
Patient problems: No Patient Involvement
MAUDE report 6325463 ↗
Other Injector And Syringe, Angiographic devices
- Monarch® 00884450298022Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls37 reports
- Blue Diamond™ 00884450498743Merit Medical Systems, Inc. · Injector and syringe, angiographic2 recalls0 reports
- Monarch® 00884450298053Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- Monarch® 00884450298060Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- NAMIC 80085007Medline Industries, Inc. · Injector and syringe, angiographic1 recall36 reports
- NA 622510B. Braun Medical Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC85 reports
- NAMIC 80087007Medline Industries, Inc. · Injector and syringe, angiographic1 recall20 reports
- Blue Diamond™ 00884450381168Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page