Skip to content

Empower FastLoad CT Syringe Pack -sterile disposable… recallZ-0654-2019 · Bracco Injeneering S.A. · initiated Nov 12, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bracco Injeneering S.A. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Insufficient seal on the sterile barrier of the device, compromising the sterility of the device

FDA root cause category: Packaging process control

What the company is doing

The firm, Bracco Injeneering, issued "URGENT: MEDICAL DEVICE RECALL" letter via third party recall firm, on 11/12/18 to customers and distributors identifying the problem, and requesting return of unused product., Complete and return the Removal Recall Response Form even if you have no recalled stock on hand. Questions: contact Stericycle at 888-759-6907 (Monday- Friday 8 a.m. to 5 p.m. Eastern Time) or [email protected].

Recall details

Recall number
Z-0654-2019
Status
Terminated
Initiated by the company
Nov 12, 2018
Recalling firm
Bracco Injeneering S.A.
Quantity
33,350 packs (667 cases)
Product code
DXT
FDA submissions
K933846
Distribution
US Nationwide Distribution

Product description

Empower FastLoad CT Syringe Pack -sterile disposable syringe kit for use with the EmpowerCT, EmpowerCTA and EmpowerCTA+ Injector Systems SKU 017344/6720 Empower FastLoad" CT Syringe Pack is a sterile, disposable syringe kit designed for use with EmpowerCT, EmpowerCTA and EmpowerCTA+ Injector Systems. It is used to intravenously facilitate the administration of contrast media and flushing solutions into the human vascular systems.

Code information (models, lots, serial numbers)

Lot # IANBL-1804 UDI: (01) 17630039300309 (17) 210420 (11) 180420 (10) IANBL-1804

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

5 devices cleared under K933846 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates