Ankle Compression Nailing System 1 (REF 99-T775100)Recalls & reported problems
Orthofix Srl · TITANIUM CALCANEUS THREADED HEAD SCREW D 5MM L 100MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 18033509858727), received between Mar 6, 2019 and Mar 6, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Mechanical Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Mechanical Problem1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 6, 2019Mechanical Problem; Mechanical Problem
THE INFORMATION PROVIDED BY THE LOCAL DISTRIBUTOR INDICATES: - PRODUCT CODE: AHN CALCANEAL SCREW CODE UNKNOWN. - BATCH NUMBER: UNKNOWN. - QUANTITY: 1. - HOSPITAL NAME: UNKNOWN. - SURGEON NAME: DR. (B)(6). - DATE OF INITIAL SURGERY: UNKNOWN. - BODY PART TO WHICH DEVICE WAS APPLIED: FOOT. - SURGERY DESCRIPTION: CORRECTION. - PATIENT INFORMATION: UNAVAILABLE. - PROBLEM OBSERVED DURING: INTO TREATMENT/POST-OPERATIVE. - TYPE OF PROBLEM: DEVICE FUNCTIONAL PROBLEM. - EVENT DESCRIPTION: "DR (B)(6) REPORTED THAT HE HAD A PATIENT IN THE OFFICE ON (B)(6) 2018 WHERE THE CALCANEAL SCREW HAD BACKED OUT OF THE BONE AND THROUGH THE SKIN. HE PULLED IT OUT REST OF THE WAY OUT". THE COMPLAINT REPORT FORM ALSO INDICATES: - THE DEVICE FAILURE DID NOT HAVE ANY ADVERSE EFFECTS ON PATIENT. - THE INITIAL SURGERY WAS COMPLETED WITH THE DEVICE. - THE EVENT DID NOT LEAD TO A CLINICALLY RELEVANT INCREASE IN THE DURATION OF THE SURGICAL PROCEDURE. - AN ADDITIONAL SURGERY WAS NOT REQUIRED. - A MEDICAL INTERVENTION (OUTPATIENT CLINIC) WAS REQUIRED: PERFORMED ON (B)(6) 2018. - COPIES OF THE OPERATIVE REPORTS ARE NOT AVAILABLE. - COPIES OF THE X-RAY IMAGES ARE NOT AVAILABLE. - PATIENT CURRENT HEALTH CONDITION: PATI
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8394188 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page