FITBONE TAA SYSTEM 1 (REF 60001348)Recalls & reported problems
Orthofix Srl · FITBONE INTRAMEDULLARY LENGTHENING NAIL TAA1180-T-245 · Class 2 · Implant
All models of FITBONE TAA SYSTEM (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 18053340343909), received between Jul 1, 2025 and Jul 1, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Mechanical Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem1
Patient problems
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 1, 2025Mechanical Problem; Mechanical Problem
RECEIVED INFORMATION THAT DURING TREATMENT, A FITBONE NAIL AFTER 4 MM LENGTHENING STOPPED DISTRACTING. PATIENT UNDERWENT A REVISION PROCEDURE. NAIL WAS PLACED BACK IN PATIENT AND WILL BE REPLACED WITH A NEW NAIL. AS PER REPORTER PATIENT IS IN GOOD HEALTH CONDITIONS. DISTRIBUTOR REF: (B)(4).
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 22372434 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page