FITBONE TAA SYSTEM 1 (REF 60001383)Recalls & reported problems
Orthofix Srl · FITBONE INTRAMEDULLARY LENGTHENING NAIL TAA1140-F-205 · Class 2 · Implant
All models of FITBONE TAA SYSTEM (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 18053340343930), received between Nov 28, 2024 and Dec 8, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Mechanical Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Mechanical Problem1
Patient problems
- Loss of fracture reduction2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryDec 8, 2024Mechanical Problem; Mechanical Problem
THE INFORMATION PROVIDED BY THE LOCAL DISTRIBUTOR INDICATES: HOSPITAL NAME: (B)(6), SURGEON'S NAME: (B)(6), PRODUCT CODE: 60001445, BATCH NUMBER: 6217512, DATE OF INITIAL SURGERY: ON (B)(6) 2023, APPLICATION SITE: RIGHT TIBIA, SURGERY DESCRIPTION: CORRECTION AND LENGTHENING, PATIENT INFORMATION: 45 YEARS, MALE, 98 KG, 186 CM, STATUS POST SPINA BIFIDA OCCULTA, PROBLEM OBSERVED DURING: INTO TREATMENT/POST-OPERATIVE, EVENT DESCRIPTION: "NAIL BREAKAGE, ACKNOWLEDGED AT FOLLOW UP EXAMINATION; LOSS OF CORRECTION + PAIN; REQUIRED SURGICAL INTERVENTION." THE REPORTING FORM ALSO INDICATED: THE DEVICE FAILURE CAUSED ADVERSE EFFECTS ON PATIENT: LOSS OF ACHIEVED CORRECTION. AN ADDITIONAL SURGERY WAS REQUIRED PERFORMED ON (B)(6) 2024. COPY OF THE OPERATIVE REPORTS IS AVAILABLE (NOT RECEIVED BY ORTHOFIX). COPY OF X-RAY IMAGES IS AVAILABLE. PATIENT CURRENT HEALTH CONDITION: PATIENT RECOVERING WELL; CORRECTION RESTORED + FITBONE EXCHANGED TO TRAUMA NAIL. FURTHER INFORMATION RECEIVED ON DECEMBER 3, 2024: THE NAIL BREAKAGE WAS DUE TO WRONG WEIGHT BEARING OF THE PATIENT AND NOT DUE TO A TECHNICAL PROBLEM OF THE NAIL. THE SURGEON USED A TRAUMA NAIL TO RECOVER THE CORRECTION. THE PATIENT IS DOING WELL,
Patient problems: Loss of fracture reduction
MAUDE report 20875180 ↗InjuryNov 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE INFORMATION PROVIDED BY THE HOSPITAL INVOLVED INDICATES: HOSPITAL NAME: (B)(6). SURGEON'S NAME: (B)(6). PRODUCT CODE: 60001383. BATCH NUMBER: B4636190. APPLICATION SITE: FEMUR. PATIENT INFORMATION: MALE. PATIENT NUMBER: (B)(6). DATE OF SURGERY: (B)(6) 2024. SURGERY DESCRIPTION: LENGTHENING. DATE OF EVENT ONSET: 21-NOV-2024. DATE OF EVENT AWARENESS: (B)(6) 2024, ADMISSION TO OUR EMERGENCY DEPARTMENT. EVENT DESCRIPTION: "LEFT FEMUR FRACTURE (ABOVE THE LENGTHENING NAIL) FOLLOWING A MECHANICAL FALL - SLIDE IN A SHOPPING MALL. THE NAIL DID NOT BREAK BUT THERE IS A LINK WITH THE LOCATION OF THE FEMORAL FRACTURE". THE REPORTING FORM ALSO INDICATED: COPY OF X-RAY IMAGES IS AVAILABLE. PATIENT CURRENT HEALTH CONDITION: HOSPITALIZATION IN THE PEDIATRICS SURGERY DEPARTMENT - SURGERY PLANNED ON THE (B)(6) 2024. ON (B)(6) 2025, ORTHOFIX RECEIVED A COPY OF THE COMPLETED COMPLAINT FORM WHICH INCLUDES THE FOLLOWING ADDITIONAL DETAILS: PATIENT'S INFORMATION: 15 YEARS, MALE, WITH WEAK BONE DENSITY DURING THE RE-INTERVENTION PERFORMED ON (B)(6) 2024, A PLATE WAS ADDED OVER THE NAIL. THE NAIL WAS NOT REMOVED. BONE CONSOLIDATION IS GOING WELL, PATIENT IS OK. ORTHOFIX SRL REF: (B)(4). DISTRIBUTOR REF
Patient problems: Loss of fracture reduction
MAUDE report 20804046 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.