CHIMAERA Hip Fracture System 1 (REF 99-T93790)Recalls & reported problems

Orthofix Srl · CHIMAERA HFS LAG SCREW SLIDING L90MM STERILE · Class 2

All models of CHIMAERA Hip Fracture System (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 18054242514749), received between Nov 30, 2017 and Mar 11, 2026 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Mechanical Problem2
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1
  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 11, 2026Mechanical Problem

    IT WAS REPORTED THAT A PORTION OF A CHIMAERA SCREW IMPLANTED ON (B)(6) 2025, DISENGAGED DURING TREATMENT. PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24567807 ↗
  • InjuryJan 14, 2021Unintended Movement; Unintended Movement

    THE INFORMATION PROVIDED BY LOCAL DISTRIBUTOR INDICATES: HOSPITAL NAME: (B)(6) HOSPITAL. SURGEON'S NAME: (B)(6). DATE OF INITIAL SURGERY: (B)(6) 2020. BODY PART TO WHICH DEVICE WAS APPLIED: TROCHANTERIC FRACTURE. SURGERY DESCRIPTION: FRACTURE TREATMENT. PATIENT'S INFORMATION: (B)(6) YEAR-OLD, FEMALE, WEIGHT (B)(6) KG., HEIGHT 1.60 CM, PREVIOUS HEALTH CONDITION: DIABETES, HIGH BLOOD PRESSURE, SMOKING. PROBLEM OBSERVED DURING: INTO TREATMENT/POSTOPERATIVE. TYPE OF PROBLEM: DEVICE FUNCTIONAL PROBLEM. EVENT DESCRIPTION: "ON THE (B)(6) 2020, A PATIENT WAS OPERATED FOR A TROCHANTERIC FRACTURED, TREATED WITH CHIMAERA TROCHANTERIC NAIL. ON DECEMBER 16TH, IT WAS REPORTED TO OUR SALES REPRESENTATIVE THAT A COMPLICATION AROSE AS THE BARREL OF THE SLIDING LAG SCREW APPEARED TOTALLY APART LYING INTO THE SOFT TISSUE OF THE FEMUR." THE COMPLAINT REPORT FORM ALSO INDICATED: THE DEVICE FAILURE HAD ADVERSE EFFECTS ON PATIENT: LOSS OF ACHIEVED CORRECTION, LOSS OF STABILITY. THE INITIAL SURGERY WAS COMPLETED WITH THE DEVICE. THE EVENT DID NOT LEAD TO A DELAY IN THE DURATION OF THE SURGICAL PROCEDURE. AN ADDITIONAL SURGERY WAS REQUIRED: A REVISION SURGERY IS SCHEDULED TO TAKE PLACE IN THE FOLLOWING DAY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11168821 ↗
  • MalfunctionMar 9, 2020Break; Break

    THE INFORMATION PROVIDED BY THE HOSPITAL INDICATES: HOSPITAL NAME: (B)(6). SURGEON NAME: DR. (B)(6). DATE OF INITIAL SURGERY: (B)(6) 2020 . BODY PART TO WHICH DEVICE WAS APPLIED: FEMUR. SURGERY DESCRIPTION: FRACTURE TREATMENT. PATIENT INFORMATION: (B)(6) YEAR-OLD, FEMALE, WEIGHT (B)(6) KG., HEIGHT 140 CM; PREVIOUS HEALTH CONDITION: GOOD CONDITIONS. INDEPENDENT. PROBLEM OBSERVED DURING: CLINICAL USE ON PATIENT/INTRAOPERATIVE. TYPE OF PROBLEM: DEVICE FUNCTIONAL PROBLEM. EVENT DESCRIPTION: "WHEN INSERTING THE CEPHALIC SCREW IT WAS NOT POSSIBLE TO SCREW IT AS IT DID NOT LOCK ON NAIL. THE SCREW WAS EXTRACTED AND IT WAS OBSERVED THE BREAKAGE OF 3 OUT OF 6 OF THE TEETH. TWO OF THEM HAVE NOT BEEN RETRIEVED." THE COMPLAINT REPORT FORM ALSO INDICATES: THE DEVICE FAILURE HAD NO ADVERSE EFFECTS ON PATIENT. THE INITIAL SURGERY WAS NOT COMPLETED WITH THE DEVICE. A REPLACEMENT DEVICE WAS IMMEDIATELY AVAILABLE TO COMPLETE SURGERY. THE EVENT LED TO AN INCREASE IN THE DURATION OF THE SURGICAL PROCEDURE OF 5 MINUTES. AN ADDITIONAL SURGERY WAS NOT REQUIRED. A MEDICAL INTERVENTION (OUTPATIENT CLINIC) WAS NOT REQUIRED. COPY OF OPERATIVE REPORTS IS AVAILABLE. COPY OF X-RAY IMAGES IS AVAILABLE. PATIENT'S

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9805962 ↗
  • InjuryDec 10, 2018Break; Break

    THE INFORMATION PROVIDED BY THE LOCAL DISTRIBUTOR INDICATES: HOSPITAL NAME: (B)(6); SURGEON'S NAME: DR. (B)(6); DATE OF INITIAL SURGERY: (B)(6) 2018; BODY PART TO WHICH DEVICE WAS APPLIED: N/A; SURGERY DESCRIPTION: FRACTURE TREATMENT; PATIENT INFORMATION: N/A; PROBLEM OBSERVED DURING: INTO TREATMENT/POST-OPERATIVE; EVENT DESCRIPTION: FEMORAL NECK WAS NOT PROPERLY REDUCED AND FIXED IN A NECK ANGLE GREATER THAN WHAT IS RECOMMENDED. THE MAIN LAG SCREW BROKE AS PER ATTACHED PICTURES. NAIL WAS EXTRACTED, AND CHIMAERA WAS USED FOR RE FIXATION. THE INITIAL PROCEDURE WAS DONE AT (B)(6) HOSPITAL. THE 2ND PROCEDURE WAS DONE AT (B)(6) ON (B)(6) 2018 THE COMPLAINT REPORT FORM ALSO INDICATES: THE DEVICE FAILURE HAD ADVERSE EFFECTS ON PATIENT: LOSS OF FRACTURE REDUCTION AND LOSS OF ACHIEVED CORRECTION. THE INITIAL SURGERY WAS COMPLETED WITH THE DEVICE. AN ADDITIONAL SURGERY WAS REQUIRED FOLLOWING DEVICE FAILURE: PERFORMED ON (B)(6) 2018. COPIES OF THE OPERATIVE REPORTS ARE NOT AVAILABLE COPIES OF THE X-RAY IMAGES ARE AVAILABLE. PATIENT CURRENT HEALTH CONDITION: N/A. ON (B)(6) 2018, ORTHOFIX (B)(4) RECEIVED COPY OF X-RAY IMAGES AND THE FOLLOWING FURTHER INFORMATION: "PLEASE FIND ATTACHED FORM AND

    Patient problems: No Code Available

    MAUDE report 8145793 ↗
  • InjuryNov 30, 2017Mechanical Problem; Mechanical Problem

    THE INFORMATION PROVIDED BY THE SALES MANAGER INDICATES: HOSPITAL NAME: (B)(6); SURGEON NAME: (B)(6); DATE OF INITIAL SURGERY: (B)(6) 2017; BODY PART TO WHICH DEVICE WAS APPLIED: FEMUR; SURGERY DESCRIPTION: FRACTURE TREATMENT; PATIENT INFORMATION: (B)(6), FEMALE; PROBLEM OBSERVED DURING TREATMENT; TYPE OF PROBLEM: DEVICE FUNCTIONAL PROBLEM; EVENT DESCRIPTION: "DURING A CHIMAERA IMPLANT (CODE OF THE NAIL WAS 99-T93125), A LAG SCREW (DEVICE CODE 99-T93790) WAS INSERTED. DURING THE COMPRESSION PHASE, THE SCREW CAME OUT OF THE NAIL. FROM THE INFORMATION AVAILABLE, IT SEEMS TO ME THAT THE WINGS OF THE LAG SCREW DID NOT HOOK-UP TO THE NAIL. ONCE THE SCREW WAS REMOVED, THE SURGEON WAS NOT ABLE TO RELEASE THE SCREW FROM THE SCREWDRIVER THEREFORE THE PROPER INSTRUMENTATION TO RE-IMPLANT THE CEPHALIC SCREW WAS NOT MORE AVAILABLE. TO CONCLUDE THE SURGERY, THE SURGEON DECIDED TO USE THE RETENTION ROD (DEVICE CODE 193282). IN THIS CONDITION HE WAS NOT ABLE TO DO A SUFFICIENT FORCE TO COMPLETE THE SCREW INSERTION. THESE MANEUVERS CAUSED THE BREAKAGE OF THE RETENTION ROD. THE INTERVENTION ENDED WITH THE LAG SCREW IMPLANTED (BUT A PORTION OF THE RETENTION ROD IS STUCK IN IT) ". THE COMPLAINT REPOR

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7070035 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →