AGILE NAIL 1 (REF 99-183281)Recalls & reported problems
Orthofix Srl · GUIDE WIRE WITH OLIVE L980MM D2.8MM · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 18054242519041), received between Jul 1, 2020 and Jul 1, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Mechanical Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Mechanical Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2020Mechanical Problem; Mechanical Problem
THE INFORMATION PROVIDED BY LOCAL DISTRIBUTOR INDICATES: PRODUCT CODE: 99183281 (GUIDE WIRE WITH OLIVE L980MM D2.8MM), BATCH NUMBER: B1256628, QUANTITY: 1, HOSPITAL NAME: (B)(6). DATE OF INITIAL SURGERY: (B)(6) 2020. BODY PART TO WHICH DEVICE WAS APPLIED: FEMUR. SURGERY DESCRIPTION: FRACTURE TREATMENT. PATIENT'S INFORMATION: NO RESPONSE. PROBLEM OBSERVED DURING: CLINICAL USE ON PATIENT/INTRAOPERATIVE. TYPE OF PROBLEM: DEVICE FUNCTIONAL PROBLEM. EVENT DESCRIPTION: "INTRODUCTION OF OLIVE GUIDE FOR IMPLANT OF AGILE NAIL. WHEN SORTING OUT THE GUIDE, OLIVE KEPT INSIDE FEMUR. IT REVEALS IMPOSSIBLE TO EXTRACT OLIVE WITHOUT OPENING FEMUR. THIS REPRESENTS AN INFECTIOUS RISK". THE COMPLAINT REPORT FORM ALSO INDICATED: THE DEVICE FAILURE HAD ADVERSE EFFECTS ON PATIENT (UNRETRIEVED DEVICE FRAGMENT), THE INITIAL SURGERY WAS COMPLETED WITH THE DEVICE, THE EVENT DID NOT LEAD TO A DELAY IN THE DURATION OF THE SURGICAL PROCEDURE, AN ADDITIONAL SURGERY WAS NOT REQUIRED, PRODUCT IS AVAILABLE FOR RETURN, NO INFORMATION WAS PROVIDED ABOUT PATIENT CURRENT HEALTH CONDITIONS. ON JUNE 11, 2020, ORTHOFIX SRL RECEIVED THE FOLLOWING ADDITIONAL INFORMATION: 1). NAME OF THE SURGEON : PR FRANCK ACCADBLED, DR (B)
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 10216634 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page