FIREBIRD 44-2001Recalls & reported problems
Orthofix US LLC · Set Screw · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 18257200009695), received between Sep 4, 2015 and Mar 8, 2016 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K081684); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Migration or Expulsion of Device1
Patient problems
- Fall2
- No Code Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 8, 2016Migration or Expulsion of Device
INFORMATION PROVIDED STATES THAT DOCTOR PERFORMED PL FUSION FROM T10-L2 ON (B)(6) 2015. DURING FOLLOW UP VISIT IT WAS NOTICED THAT THREE SET SCREWS, (2) AT L2 AND (1) AT L3, HAD BACKED OUT OF THE CONSTRUCT ALLOWING THE RODS OUT OF THE TULIPS. REVISION SURGERY WAS PERFORMED ON (B)(6) 2016 TO REMOVE AND REPLACE 10 SET SCREWS.
Patient problems: No Code Available
MAUDE report 5486788 ↗MalfunctionSep 4, 2015Device Dislodged or Dislocated
INFORMATION PROVIDED STATES THAT THE PATIENT HAD SURGERY AT L5-S1 ON (B)(6) 2015. THE PATIENT FELL THREE MONTHS POST-OP. X-RAYS TAKEN SHOW THAT TWO OF THE SET SCREWS HAD BECOME DISLODGED FROM HARDWARE ON RIGHT AND LEFT SIDE AT L5. REVISION SURGERY TOOK PLACE ON (B)(6) 2015 TO REMOVE AND REPLACE TWO SET SCREWS. THE PATIENT IS IN GOOD CONDITION.
Patient problems: Fall
MAUDE report 5058397 ↗MalfunctionSep 4, 2015Device Dislodged or Dislocated
INFORMATION PROVIDED STATES THAT THE PATIENT HAD SURGERY AT L4-L5 ON (B)(6) 2015. THE PATIENT FELL ON (B)(6) 2015. XSET -RAYS TAKEN ON (B)(6) 2015 SHOWED THAT THE LEFT L4 PEDICLE SCREW HAD COME OUT OF THE HARDWARE. REVISION SURGERY TOOK PLACE ON (B)(6) 2015 TO REMOVE AND REPLACE THE LOOSE SCREW. THE PATIENT IS IN GOOD CONDITION.
Patient problems: Fall
MAUDE report 5056731 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081684. One submission often covers many models, so check the product and lot details in each record.
- Z-0855-2014 · Nov 26, 2013 · TerminatedOrthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver, Product Usage: The Modular Screw Driver is used for inserting a modular bone screw into the pedic
- Z-1608-2011 · Oct 22, 2010 · TerminatedORTHOFIX Firebird Spinal Fixation System, Pedicle Screw Spinal System, Reduction Case, Part #68-0111, Lot #16518-LC18. Intended for use in conjunction with Multi-Axial Reduction S
- Z-1674-2011 · Oct 22, 2010 · TerminatedBLACKSTONE Medical Inc. Firebird Spinal Fixation System 6.5mm x 30mm Bone Screw, Self Tapping, Cannulated. BLACKSTONE Medical Inc. Firebird Spinal Fixation System 6.5mm x 30mm Mul
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page