FIREBIRD NXG 36-2001Recalls & reported problems

Orthofix US LLC · SET SCREW · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 18257200084487), received between Oct 14, 2020 and Mar 11, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information4
  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Appropriate Device Problem Term/Code Not Available1
  • Loose or Intermittent Connection1

Patient problems

  • Insufficient Information4
  • Failure of Implant3

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT 36-2001 SET SCREW BACKED OUT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 24575009 ↗
  • MalfunctionMar 11, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT 36-2001 SET SCREW BACKED OUT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 24569223 ↗
  • MalfunctionMar 11, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT 36-2001 SET SCREW BACKED OUT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 24569222 ↗
  • MalfunctionFeb 6, 2025Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth; Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A PATIENT WENT TO A FOLLOW UP VISIT 3-4 WEEKS AFTER SURGERY AND A SET SCREW WAS FOUND TO BE LOOSE ON POST OP IMAGING AT L2-L3 LEVEL. THE PATIENT REQUIRED A REVISION SURGERY THAT EXTENDED UP ONE LEVEL FROM PREVIOUS HARDWARE. PATIENT DID NOT REPORT ANY PAIN. FURTHER DETAILS WERE REQUESTED. NO FURTHER DETAILS HAVE BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 21327513 ↗
  • MalfunctionFeb 6, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A PATIENT WENT TO A ROUTINE FOLLOW UP VISIT AFTER A DECOMPRESSION AND FUSION WITH THE USE OF FENESTERATED SCREWS THAT WERE IMPLANTED (B)(6) 2024. THE FOLLOW UP EXAM CONSISTED OF IMAGING THAT SHOWED SET SCREW BACKED OUT ON AN L2 -PELVIS JANUSUS FENESTRATED CASE. THE PATIENT REQUIRED A REVISION SURGERY THAT TOOK PLACE ON (B)(6)2025. NO SYMPTOMS WERE REPORTED BY PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 21326995 ↗
  • MalfunctionFeb 3, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A PATIENT WENT TO A ROUTINE FOLLOW UP VISIT AFTER A DECOMPRESSION AND FUSION WITH THE USE OF FENESTERATED SCREWS THAT WERE IMPLANTED (B)(6)2024. THE FOLLOW UP EXAM CONSISTED OF IMAGING THAT SHOWED SET SCREW BACKED OUT ON AN L2 -PELVIS JANUSUS FENESTRATED CASE. THE PATIENT REQUIRED A REVISION SURGERY THAT TOOK PLACE ON (B)(6)2025. NO SYMPTOMS WERE REPORTED BY PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 21293671 ↗
  • MalfunctionOct 14, 2020Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available

    INFORMATION PROVIDED STATES THAT PATIENT HAD L3-S1 POSTERIOR LUMBAR INTERBODY FUSION (PLIF) SURGERY ON (B)(6) 2020. IN WAS FOUND VIA X-RAYS DURING FOLLOW-UP VISIT THAT TWO OF THE LOCKING CAPS FROM S1 HAD BECOME LOOSE. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2020 TO REMOVE AND REPLACE THE HARDWARE AT S1.

    Patient problems: Insufficient Information

    MAUDE report 10680951 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →