DLP® 69331 (UDI 20199150021998)Recalls & reported problems
Medtronic, Inc. · CANNULA 69331 VEN RT METAL 31FR 10PK 17L · Class 2
All models of DLP (151)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20199150021998), received between Sep 29, 2026 and Sep 29, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Puncture/Hole” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K120988); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Material Puncture/Hole
It was reported that prior to use, the DLP Single Stage Venous Cannula with Right Angle Metal Tip was opened and its tip appeared sharp where the holes were drilled. The cannula was not used and was replaced with another product. There was no patient involved.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928661 ↗MalfunctionSep 29, 2026Material Puncture/Hole
It was reported that, upon opening the DLP Single Stage Venous Cannula with Right Angle Metal Tip prior to use, the cannula tip at the drilled holes felt sharp. The cannula was not used and was replaced with a Medtronic product. There was no patient involved.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928436 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K120988. One submission often covers many models, so check the product and lot details in each record.
- Z-3128-2024 · Jul 9, 2024 · Open, ClassifiedMedtronic DLP Single Stage Venous Cannulae with right angle metal tip, REF 67312
- Z-0908-2013 · Feb 19, 2013 · TerminatedMedtronic DPL Single Stage Venous Cannula with Metal Tip, used in cardiopulmonary bypass procedures. Model numbers: 67312; 67314; CB67314; 67316; 67318; CB67318; 67320; 69312;
Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY11R00R6Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page