Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20199150023282), received between Sep 30, 2026 and Sep 30, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K031037); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 24, 2026Open, Classified

    Medtronic received reports of pinhole leaks in the wire-wound body of 22 Fr EOPA 3D Arterial Cannulae and EOPA Elongated One-Piece Arterial Cannulae used during cardiopulmonary bypass procedures.

    Z-2807-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Material Deformation1
  • Packaging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Deformation; Packaging Problem

    It was reported that prior to the operation, an EOPA arterial cannula was identified upon opening the package with a squashed and flattened tip. The packaging heat-seal area was damaged and showed blue ink and lines corresponding to the cannula wire-wound tip section. It was alleged that the cannula tip was positioned in the packaging heat-seal area during manufacturing and was crushed by heat-sealing equipment. The cannula was unusable and was replaced with another EOPA cannula without delay to the operation. There was no patient involved.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942847 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K031037. One submission often covers many models, so check the product and lot details in each record.

  • Z-2808-2026 · Jun 24, 2026 · Open, Classified
    EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622
  • Z-2806-2026 · Jun 24, 2026 · Open, Classified
    EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422
  • Z-2809-2026 · Jun 24, 2026 · Open, Classified
    EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77722
  • Z-0985-2025 · Dec 12, 2024 · Open, Classified
    EOPA Arterial Cannula, Model Numbers 77422 (22 Fr) and 77418 (18 Fr)
  • Z-2420-2015 · Jul 24, 2015 · Terminated
    Medtronic EOPA Arterial Cannula, Model # 77620. Lot: 2015060544 intended to deliver oxygenated blood to the heart during cardiopulmonary bypass procedures for up to 6 hours in du