McKesson 5000Recalls & reported problems

Mckesson Medical-Surgical Inc. · TEST KIT, PREGNANCY HCG DIPSTI · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This McKesson Consult hCG pregnancy test dipstick appears in 90 adverse event reports, with no deaths and 2 injuries. Reports mostly describe false positive pregnancy results. There were no reports in the past 12 months, down from 14 in the year before. The 2 open recalls concern possible false positives at very low hormone levels, and kits exposed to higher temperatures before delivery.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
91
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 91 adverse event reports that list this exact device (DI 20612479202751), received between Jun 19, 2017 and Jun 2, 2026 — classified as 2 injuries, 89 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (70 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 6 reportsJul 2024: 0 reportsAug 2024: 5 reportsSep 2024: 2 reportsOct 2024: 5 reportsNov 2024: 4 reportsDec 2024: 0 reportsJan 2025: 5 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 20, 2026Class IIOpen, Classified

    hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.

    Z-3273-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. May 25, 2022Class IIOpen, Classified

    Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

    Z-1587-2022 · Root cause: Environmental control · Recall details · FDA record ↗

Most reported problems

Device problems

  • False Positive Result70
  • False Negative Result20
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions66
  • No Consequences Or Impact To Patient18
  • Insufficient Information5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Code Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 91 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 2, 2026False Positive Result

    THE CUSTOMER REPORTED THREE FALSE POSITIVE RESULTS WHEN TESTING A PATIENT WITH THE CONSULT HCG DIPSTICK. THE PATIENT PRESENTED WITH ABDOMINAL PAIN. A SERUM SAMPLE WAS COLLECTED AND CONFIRMATORY TESTING WAS PERFORMED AT LABCORP WITH A NEGATIVE RESULT. THE PATIENT WAS NOT SEXUALLY ACTIVE. ADMINISTRATION OF AN UNSPECIFIED VACCINE WAS DELAYED DUE TO THE POSITIVE RESULTS, HOWEVER THE CUSTOMER CONFIRMED THAT THERE WAS NO NEGATIVE IMPACT TO THE PATIENT'S HEALTH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25366136 ↗
  • MalfunctionMar 10, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING A FAINT FALSE POSITIVE HCG RESULT WHILE USING THE CONSULT HCG DIPSTICK TEST ON A NON-PREGNANT PATIENT. SPECIFICS WERE NOT PROVIDED AS TO WHY THE PATIENT PRESENTED, HOWEVER THE PATIENT PROVIDED A FRESH URINE SAMPLE WHICH WAS TESTED USING THE CONSULT HCG DIPSTICK. THE RESULTS WERE READ AT 3 MINUTES AND A FAINT FALSE POSITIVE RESULT WAS OBTAINED. CONFIRMATORY TESTING WAS PERFORMED USING A CONSULT HCG URINE CASSETTE WHICH YIELDED A NEGATIVE HCG RESULT. NO ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21568394 ↗
  • MalfunctionMar 5, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING DISCREPANT RESULTS FOR 2 DIFFERENT INDIVIDUALS WHEN USING CONSULT HCG DIPSTICK TESTS AND CONSULT HCG URINE CASSETTES. THE CUSTOMER REPORTED AFTER RECEIVING DISCREPANT RESULTS FOR THE FIRST PATIENT, THE CUSTOMER DECIDED TO TEST A COLLEAGUE. THE CUSTOMER'S COLLEAGUE PROVIDED A FRESH URINE SAMPLE WHICH YIELDED A FAINT POSITIVE HCG RESULT WITH THE DIPSTICK TEST AND A NEGATIVE HCG RESULT FOR THE CASSETTE TEST. THE CUSTOMER INDICATED THEY CONSIDERED THE FAINT POSITIVE HCG RESULT TO BE FALSE FOR THE COLLEAGUE. NO ADVERSE EVENT WAS REPORTED. NO FURTHER INFORMATION WAS PROVIDED. SEE MDR 2027969-2025-00057 FOR THE DISCREPANT RESULT ON PATIENT 1.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21527958 ↗
  • MalfunctionFeb 21, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING TWO FALSE POSITIVE HCG RESULTS WHILE USING CONSULT HCG DIPSTICK TESTS ON TWO PATIENTS. THE FIRST PATIENT PRESENTED FOR PREGNANCY TESTING PRIOR TO UNDERGOING AN ENDOMETRIAL BIOPSY AND SURGERY DUE TO PAIN. THE PATIENT PROVIDED A FRESH URINE SAMPLE WHICH YIELDED A POSITIVE HCG RESULT. A CONFIRMATORY BLOOD TEST WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. HOWEVER, AS A RESULT OF THE POSITIVE HCG RESULT, THE ENDOMETRIAL BIOPSY PROCEDURE WAS RESCHEDULED FOR (B)(6) 2025. ALTHOUGH THE CUSTOMER REPORTED A DELAY IN UNDERGOING THE ENDOMETRIAL BIOPSY PROCEDURE, THERE WAS NO DELAY IN THE SURGICAL PROCEDURE AS THE SURGERY WAS SCHEDULED FOR (B)(6) 2025 DUE TO OR AVAILABILITY. THE CUSTOMER MENTIONED THE PATIENT EXPRESSED FRUSTRATION DUE TO THE PAIN THEY WERE IN, HOWEVER THE FALSE POSITIVE HCG RESULT DID NOT RESULT IN A DELAY IN TREATMENT. NO ADVERSE EVENT REPORTED. SEE MDR 2027969-2025-00052 REGARDING THE FALSE POSITIVE EVENT ON THE SECOND PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21438768 ↗
  • MalfunctionFeb 21, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING TWO FALSE POSITIVE HCG RESULTS WHILE USING CONSULT HCG DIPSTICK TESTS ON TWO PATIENTS. THE SECOND PATIENT PRESENTED FOR AN UNSPECIFIED REASON. THE PATIENT PROVIDED A FRESH URINE SAMPLE WHICH YIELDED A POSITIVE HCG RESULT. A SECOND TEST WAS PERFORMED USING THE SAME URINE SAMPLE WITH A CONSULT HCG DEVICE FROM A DIFFERENT KIT (SAME LOT) WHICH YIELDED A NEGATIVE RESULT. NO ADVERSE EVENTS WERE REPORTED. SEE MDR 2027969-2025-00051 REGARDING THE FALSE POSITIVE EVENT ON THE FIRST PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21437973 ↗
  • MalfunctionFeb 18, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING A FALSE POSITIVE HCG RESULT WHILE USING THE CONSULT HCG DIPSTICK TEST ON A PATIENT. THE PATIENT PROVIDED A FRESH URINE SAMPLE WHICH HAD NO APPEARANCE ISSUES. THE SAMPLE WAS TESTED, THE RESULTS WERE READ BETWEEN 3-4 MINUTES, AND A FALSE POSITIVE HCG RESULT WAS OBSERVED. THE CUSTOMER REPORTED RETESTING WAS PERFORMED USING A DEVICE FROM THE SAME LOT NUMBER AND A NEGATIVE HCG RESULT WAS OBTAINED. ADDITIONALLY, A CONFIRMATORY BLOOD TEST WAS PERFORMED WHICH ALSO YIELDED A NEGATIVE HCG RESULT. NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21398123 ↗
  • MalfunctionJan 20, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WHEN TESTING TWO PATIENTS WITH THE CONSULT DIAGNOSTICS HCG URINE TESTS DIPSTICK. THE CUSTOMER WAS UNABLE TO PROVIDE ANY INFORMATION ON THE SECOND PATIENT, AS THE EVENT HAD OCCURRED SOME TIME AGO. NO ADVERSE EVENTS WERE REPORTED. SEE H10 FOR MDR NUMBER RELATED TO THE OTHER PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21194240 ↗
  • MalfunctionJan 20, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WHEN TESTING TWO PATIENTS WITH THE CONSULT DIAGNOSTICS HCG URINE TESTS DIPSTICK. PATIENT (B)(6) WAS TESTED TWICE, BOTH TIMES WITH POSITIVE RESULTS. THE PATIENT'S PRESCRIPTION WAS CHANGED FROM SUBOXONE (BUPRENORPHINE/NALOXONE) TO BUPRENORPHINE DUE TO THE FALSE POSITIVE RESULTS. THE PATIENT STATED THAT IT WAS IMPOSSIBLE THAT SHE WAS PREGNANT AND PERFORMED HER OWN PREGNANCY TEST THE FOLLOWING DAY WITH A NEGATIVE RESULT. THE PATIENT HAD NOT TAKEN ANY OF THE NEW MEDICATION, AND THE PRESCRIPTION WAS CHANGED BACK FOLLOWING THE NEGATIVE RESULT. THE CUSTOMER CONFIRMED THAT THERE WAS NO NEGATIVE IMPACT DUE TO THE CHANGE IN MEDICATION. SEE H10 FOR MDR NUMBER RELATED TO THE OTHER PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21194231 ↗
  • MalfunctionJan 8, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING SEVERAL FALSE POSITIVE RESULTS WHILE USING THE CONSULT HCG URINE TESTS ON FOUR DIFFERENT PATIENTS. ON (B)(6) 2024, THE CUSTOMER INDICATED ONE PATIENT RECEIVED FALSE POSITIVE HCG RESULTS ON THAT DAY AND THREE PATIENTS RECEIVED FALSE POSITIVE RESULTS ONE MONTH PRIOR. IT WAS REPORTED THAT ALL PATIENTS HAD CONFIRMATORY TESTING PERFORMED WHICH YIELDED NEGATIVE HCG RESULTS. THE CUSTOMER INDICATED THE MOST RECENT PATIENT FROM (B)(6) 2024 PROVIDED A FRESH URINE SAMPLE, THE RESULTS WERE READ AT 5 MINUTES AND 3 POSITIVE HCG RESULTS WERE OBSERVED. TECHNICAL SERVICES ADVISED PER THE PACKAGE INSERT, READ THE RESULT AT 3-4 MINUTES. DO NOT INTERPRET RESULTS AFTER THE APPROPRIATE READ TIME. NO ADVERSE EVENTS WERE REPORTED, THEREFORE THIS EVENT IS NOT CONSIDERED REPORTABLE. IT IS CURRENTLY UNCLEAR IF THIS READ TIME DEVIATION APPLIES TO THE 3 PATIENTS FROM THE MONTH PRIOR, THEREFORE 3 MDRS WILL BE CONSERVATIVELY FILED. SEE MDR 2027969-2025-00008 AND 2027969-2025-00009 REGARDING THE FALSE POSITIVE RESULTS FROM PATIENT 1 AND 2 RESPECTIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21107791 ↗
  • MalfunctionJan 8, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING SEVERAL FALSE POSITIVE RESULTS WHILE USING THE CONSULT HCG URINE TESTS ON FOUR DIFFERENT PATIENTS. ON (B)(6) 2024, THE CUSTOMER INDICATED ONE PATIENT RECEIVED FALSE POSITIVE HCG RESULTS ON THAT DAY AND THREE PATIENTS RECEIVED FALSE POSITIVE RESULTS ONE MONTH PRIOR. IT WAS REPORTED THAT ALL PATIENTS HAD CONFIRMATORY TESTING PERFORMED WHICH YIELDED NEGATIVE HCG RESULTS. THE CUSTOMER INDICATED THE MOST RECENT PATIENT FROM (B)(6) 2024 PROVIDED A FRESH URINE SAMPLE, THE RESULTS WERE READ AT 5 MINUTES AND 3 POSITIVE HCG RESULTS WERE OBSERVED. TECHNICAL SERVICES ADVISED PER THE PACKAGE INSERT, READ THE RESULT AT 3-4 MINUTES. DO NOT INTERPRET RESULTS AFTER THE APPROPRIATE READ TIME. NO ADVERSE EVENTS WERE REPORTED, THEREFORE THIS EVENT IS NOT CONSIDERED REPORTABLE. IT IS CURRENTLY UNCLEAR IF THIS READ TIME DEVIATION APPLIES TO THE 3 PATIENTS FROM THE MONTH PRIOR, THEREFORE 3 MDRS WILL BE CONSERVATIVELY FILED. SEE MDR 2027969-2025-00008 AND 2027969-2025-00010 REGARDING THE FALSE POSITIVE RESULTS FROM PATIENT 1 AND 3 RESPECTIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21107675 ↗
  • MalfunctionJan 8, 2025False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING SEVERAL FALSE POSITIVE RESULTS WHILE USING THE CONSULT HCG URINE TESTS ON FOUR DIFFERENT PATIENTS. ON (B)(6) 2024, THE CUSTOMER INDICATED ONE PATIENT RECEIVED FALSE POSITIVE HCG RESULTS ON THAT DAY AND THREE PATIENTS RECEIVED FALSE POSITIVE RESULTS ONE MONTH PRIOR. IT WAS REPORTED THAT ALL PATIENTS HAD CONFIRMATORY TESTING PERFORMED WHICH YIELDED NEGATIVE HCG RESULTS. THE CUSTOMER INDICATED THE MOST RECENT PATIENT FROM (B)(6) 2024 PROVIDED A FRESH URINE SAMPLE, THE RESULTS WERE READ AT 5 MINUTES AND 3 POSITIVE HCG RESULTS WERE OBSERVED. TECHNICAL SERVICES ADVISED PER THE PACKAGE INSERT, READ THE RESULT AT 3-4 MINUTES. DO NOT INTERPRET RESULTS AFTER THE APPROPRIATE READ TIME. NO ADVERSE EVENTS WERE REPORTED, THEREFORE THIS EVENT IS NOT CONSIDERED REPORTABLE. IT IS CURRENTLY UNCLEAR IF THIS READ TIME DEVIATION APPLIES TO THE 3 PATIENTS FROM THE MONTH PRIOR, THEREFORE 3 MDRS WILL BE CONSERVATIVELY FILED. SEE MDR 2027969-2025-00009 AND 2027969-2025-00010 REGARDING THE FALSE POSITIVE RESULTS FROM PATIENT 2 AND 3 RESPECTIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21107398 ↗
  • MalfunctionNov 13, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT URINE SAMPLE WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THE PATIENT HAD EXPERIENCED A MISCARRIAGE APPROXIMATELY 6 WEEKS PRIOR. LAB TESTING WAS PERFORMED WITH A RESULT OF <5 MIU/ML. NO ADVERSE EVENTS WERE REPORTED. THE CUSTOMER ALSO REPORTED FALSE POSITIVE RESULTS FOR THREE OTHER PATIENTS. SEE H10 FOR THE RELATED REPORT NUMBERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20677821 ↗
  • MalfunctionNov 13, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT URINE SAMPLE WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THE PATIENT WAS 6 WEEKS POSTPARTUM, AND WAS BEING TESTED AS PART OF A ROUTINE FOLLOW-UP. LAB TESTING WAS PERFORMED TWICE WITH A RESULT OF <1 MIU/ML EACH TIME. NO ADVERSE EVENTS WERE REPORTED. THE CUSTOMER ALSO REPORTED FALSE POSITIVE RESULTS FOR THREE OTHER PATIENTS. SEE H10 FOR THE RELATED REPORT NUMBERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20676563 ↗
  • MalfunctionNov 13, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT URINE SAMPLE WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THE PATIENT WAS PERI-MENOPAUSAL, AND WAS BEING TESTED AS PART OF AN ANNUAL APPOINTMENT. LAB TESTING WAS PERFORMED WITH A RESULT OF 5 MIU/ML. NO ADVERSE EVENTS WERE REPORTED. THE CUSTOMER ALSO REPORTED FALSE POSITIVE RESULTS FOR THREE OTHER PATIENTS. SEE H10 FOR THE RELATED REPORT NUMBERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20676297 ↗
  • MalfunctionNov 13, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED TWO FAINT FALSE POSITIVE RESULTS WHEN TESTING A PATIENT URINE SAMPLE WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THE SAMPLE WAS TESTED A THIRD TIME WITH A NEGATIVE RESULT. THERE WAS NO NEGATIVE IMPACT TO THE PATIENT'S CARE AS A RESULT OF THE TWO ALLEGED FALSE POSITIVES. THE CUSTOMER WAS UNABLE TO PROVIDE ANY OTHER INFORMATION REGARDING THIS EVENT. THE CUSTOMER ALSO REPORTED FALSE POSITIVE RESULTS FOR THREE OTHER PATIENTS. SEE H10 FOR THE RELATED REPORT NUMBERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20676077 ↗
  • MalfunctionOct 22, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING A TOTAL OF 3 FALSE POSITIVE RESULTS WHILE USING CONSULT HCG DIPSTICK TESTS ON 3 PATIENTS. THE CUSTOMER REPORTED THE THIRD PATIENT PRESENTED FOR A L-SPINE X-RAY PROCEDURE DUE TO BACK PAIN. THE PATIENT PROVIDED A FRESH URINE SAMPLE THAT WAS DIPPED FOR 5 SECONDS. THE RESULTS WERE READ AT 3 MINUTES AND A FALSE POSITIVE HCG RESULT WAS OBSERVED. AS A RESULT OF THE POSITIVE HCG RESULT, THE PROCEDURE WAS NOT PERFORMED, AND THE PATIENT WAS PROVIDED "SUPPORTIVE OVER-THE-COUNTER MEDICINES (TYLENOL EVERY 4-6 HOURS/IBUPROFEN EVERY 6-8 HOURS)" FOR THE BACK PAIN. THE CUSTOMER REPORTED THE PATIENT WAS STABLE. SEE MDR 2027969-2024-00177 AND 2027969-2024-00178 FOR THE FIRST AND SECOND PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20510408 ↗
  • MalfunctionOct 22, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING A TOTAL OF 3 FALSE POSITIVE RESULTS WHILE USING CONSULT HCG DIPSTICK TESTS ON 3 PATIENTS. THE CUSTOMER REPORTED THE SECOND PATIENT HAD A PREGNANCY TEST CONDUCTED AS SHE COULD NOT RECALL HER LAST MENSTRUATION. THE PATIENT PROVIDED A FRESH URINE SAMPLE THAT WAS DIPPED FOR 5 SECONDS. THE RESULTS WERE READ AT 3 MINUTES AND A FALSE POSITIVE HCG RESULT WAS OBSERVED. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE HCG RESULT. NO ADVERSE EVENT REPORTED. SEE MDR 2027969-2024-00177 AND 2027969-2024-00179 FOR THE FIRST AND THIRD PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20509829 ↗
  • MalfunctionOct 22, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING A TOTAL OF 3 FALSE POSITIVE RESULTS WHILE USING CONSULT HCG DIPSTICK TESTS ON 3 PATIENTS. THE CUSTOMER REPORTED THE FIRST PATIENT TESTED POSITIVE FOR COVID-19 AND PRIOR TO ADMINISTERING A STEROID TREATMENT (4MG DEXAMETHASONE AND 40MG DEPO-MEDROL INJECTION), A PREGNANCY TEST WAS CONDUCTED. THE PATIENT PROVIDED A FRESH URINE SAMPLE THAT WAS DIPPED FOR 5 SECONDS. THE RESULTS WERE READ AT 3 MINUTES AND A FAINT FALSE POSITIVE HCG RESULT WAS OBSERVED. AS A RESULT OF THE POSITIVE HCG RESULT, THE STEROID TREATMENT WAS NOT PROVIDED. THE CUSTOMER REPORTED THE PATIENT WAS STABLE. SEE MDR 2027969-2024-00178 AND 2027969-2024-00179 FOR THE SECOND AND THIRD PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20509213 ↗
  • MalfunctionOct 15, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WHEN TESTING PATIENTS WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THREE ATTEMPTS HAVE BEEN MADE TO REQUEST ADDITIONAL DETAILS; HOWEVER THE CUSTOMER HAS BEEN UNRESPONSIVE. NO ADVERSE EVENTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20457610 ↗
  • MalfunctionOct 14, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WHEN TESTING PATIENTS WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THREE ATTEMPTS HAVE BEEN MADE TO REQUEST ADDITIONAL DETAILS, HOWEVER THE CUSTOMER HAS BEEN UNRESPONSIVE. NO ADVERSE EVENTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20447373 ↗
  • MalfunctionSep 6, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT URINE SAMPLE WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THE PATIENT WAS BEING TESTED AS PART OF A ROUTINE MEDICAL CHECKUP PRIOR TO THE PRESCRIPTION OF BIRTH CONTROL. FOLLOWING THE POSITIVE RESULT, A URINE SAMPLE WAS TESTED AT THE LAB WITH AN HCG CASSETTE (BRAND AND LOT NUMBER UNKNOWN) WITH A NEGATIVE RESULT. THERE WAS NO NEGATIVE IMPACT TO THE PATIENT'S MEDICAL CARE. THE CUSTOMER REPORTED FALSE POSITIVE RESULTS FOR TWO ADDITIONAL PATIENTS. SEE H10 FOR THE RELATED MDR NUMBERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20161472 ↗
  • MalfunctionSep 6, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT URINE SAMPLE WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THE PATIENT WAS BEING TESTED AS PART OF A ROUTINE MEDICAL CHECKUP PRIOR TO THE PRESCRIPTION OF BIRTH CONTROL. FOLLOWING THE POSITIVE RESULT, A URINE SAMPLE WAS TESTED AT THE LAB WITH AN HCG CASSETTE (BRAND AND LOT NUMBER UNKNOWN) WITH A NEGATIVE RESULT. THERE WAS NO NEGATIVE IMPACT TO THE PATIENT'S MEDICAL CARE. THE CUSTOMER REPORTED FALSE POSITIVE RESULTS FOR TWO ADDITIONAL PATIENTS. SEE H10 FOR THE RELATED MDR NUMBERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20161372 ↗
  • MalfunctionAug 23, 2024False Positive Result; False Positive Result

    THE CUSTOMER REPORTED RECEIVING A FALSE POSITIVE HCG RESULT WHILE USING THE CONSULT HCG DIPSTICK TEST. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. NO ADVERSE EVENTS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20057429 ↗
  • MalfunctionAug 21, 2024False Negative Result; False Negative Result

    THE DISTRIBUTOR REPORTED A FALSE NEGATIVE RESULT WHEN TESTING A PATIENT WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THE CUSTOMER STATED THAT THE RESULT WAS NOT CONSISTENT WITH THE LAB TESTS. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. NO ADVERSE EVENTS WERE REPORTED. THE CUSTOMER ALSO REPORTED A FALSE POSITIVE RESULT WITH THIS KIT. SEE H10 FOR THE RELATED MDR NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20014034 ↗
  • MalfunctionAug 18, 2024False Positive Result; False Positive Result

    THE DISTRIBUTOR REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT WITH THE MCKESSON CONSULT DIAGNOSTICS HCG DIPSTICK. THE CUSTOMER STATED THAT THE RESULT WAS NOT CONSISTENT WITH THE LAB TESTS. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. NO ADVERSE EVENTS WERE REPORTED. THE CUSTOMER ALSO REPORTED A FALSE NEGATIVE RESULT WITH THIS KIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19998904 ↗

FDA clearance and classification

The FDA regulates product code JHI (visual, pregnancy hcg, prescription use) under 21 CFR 862.1155 as a Class II device. 21 CFR 862.1155 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the McKesson 5000

Has the McKesson 5000 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-3273-2026 (Class II), began on Aug 20, 2026: hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the McKesson 5000?

The FDA's MAUDE database lists 91 adverse event reports for this device, received between Jun 19, 2017 and Jun 2, 2026. In those reports, the most common specific device problems are “False Positive Result” (70), “False Negative Result” (20) and “Incorrect, Inadequate or Imprecise Result or Readings” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the McKesson 5000?

None of the reports are classified as deaths, 2 as injuries and 89 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the McKesson 5000, and is it still sold?

The FDA's device database lists Mckesson Medical-Surgical Inc. as the company responsible for it (brand McKesson, model 5000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.