McKesson 5001Recalls & reported problems
Mckesson Medical-Surgical Inc. · TEST KIT, PREG HCG CASSETTE WA · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This McKesson Consult hCG urine pregnancy test has 168 reports, with no deaths and 5 injuries. Many reports describe false positive results, and some describe false negatives. There are 2 open recalls: one says the test may detect very low hCG levels below its stated sensitivity, which can produce false positives, and the other concerns high temperatures before delivery that may have reduced effectiveness.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 177
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 177 adverse event reports that list this exact device (DI 20612479202744), received between Nov 17, 2017 and Aug 31, 2026 — classified as 5 injuries, 172 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (126 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 20, 2026Class IIOpen, Classified
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Z-3271-2026 · Root cause: Device Design · Recall details · FDA record ↗
- May 25, 2022Class IIOpen, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Z-1587-2022 · Root cause: Environmental control · Recall details · FDA record ↗
Most reported problems
Device problems
- False Positive Result126
- False Negative Result46
- Incorrect, Inadequate or Imprecise Result or Readings5
- Incorrect Or Inadequate Test Results1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions109
- No Consequences Or Impact To Patient44
- Insufficient Information7
- No Information7
- No Known Impact Or Consequence To Patient6
- No Code Available3
- Miscarriage1
Recent adverse event reports
The 25 most recent of 177 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50532702 ↗MalfunctionAug 31, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50532498 ↗MalfunctionJun 9, 2026False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT WITH THE MCKESSON CONSULT DIAGNOSTICS HCG URINE CASSETTE. A RETEST WAS PERFORMED WITH AN ALTERNATE LOT NUMBER (0001134798), WHICH YIELDED A NEGATIVE RESULT. QUANTITATIVE TESTING WAS PERFORMED USING A SIEMENS DIMENSION ANALYZER AND YIELDED A RESULT OF <1 MIU/ML. NO ADDITIONAL PATIENT INFORMATION WAS PROVIDED. NO ADVERSE EVENTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25450657 ↗MalfunctionJun 9, 2026False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT WITH THE MCKESSON CONSULT DIAGNOSTICS HCG URINE CASSETTE. A RETEST WAS PERFORMED WITH AN ALTERNATE LOT NUMBER (0001134798), WHICH YIELDED A NEGATIVE RESULT. QUANTITATIVE TESTING WAS PERFORMED USING A SIEMENS DIMENSION ANALYZER AND YIELDED A RESULT OF <1 MIU/ML. NO ADDITIONAL PATIENT INFORMATION WAS PROVIDED. NO ADVERSE EVENTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25450521 ↗MalfunctionJun 9, 2026False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT WITH THE MCKESSON CONSULT DIAGNOSTICS HCG URINE CASSETTE. QUANTITATIVE TESTING WAS PERFORMED USING A SIEMENS DIMENSION ANALYZER AND YIELDED A RESULT OF <1 MIU/ML. NO ADDITIONAL PATIENT INFORMATION WAS PROVIDED. NO ADVERSE EVENTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25450397 ↗MalfunctionJun 9, 2026False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT WITH THE MCKESSON CONSULT DIAGNOSTICS HCG URINE CASSETTE. QUANTITATIVE TESTING WAS PERFORMED USING A SIEMENS DIMENSION ANALYZER AND YIELDED A RESULT OF <1 MIU/ML. NO ADDITIONAL PATIENT INFORMATION WAS PROVIDED. NO ADVERSE EVENTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25450390 ↗MalfunctionJun 5, 2026False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT FOR PAT-013636 WHEN TESTING WITH THE MCKESSON CONSULT DIAGNOSTICS HCG URINE CASSETTE. THE PATIENT WAS BEING SCREENED FOR PREGNANCY PRIOR TO THE PRESCRIPTION OF BIRTH CONTROL. THE PATIENT HAD A PREGNANCY TEST PERFORMED AT ANOTHER FACILITY TEN DAYS LATER WITH A NEGATIVE RESULT. THE PRESCRIPTION OF BIRTH CONTROL WAS DELAYED DUE TO THE POSITIVE RESULT. NO ADVERSE HEALTH OUTCOMES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25408969 ↗MalfunctionJun 5, 2026False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT FOR (B)(6) WHEN TESTING WITH THE MCKESSON CONSULT DIAGNOSTICS HCG URINE CASSETTE. THE PATIENT WAS BEING SCREENED FOR PREGNANCY PRIOR TO THE PRESCRIPTION OF BIRTH CONTROL. NO CONFIRMATORY TESTING WAS PERFORMED; HOWEVER, PREGNANCY WAS RULED OUT BASED ON LAST MENSTRUAL PERIOD AND SEXUAL HISTORY. THE PRESCRIPTION OF BIRTH CONTROL WAS DELAYED DUE TO THE POSITIVE RESULT. NO ADVERSE HEALTH OUTCOMES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25408968 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED 18MAR2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 10. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 2027969
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267383 ↗MalfunctionMay 24, 2026False Positive Result
The Distributor reported that a customer received a total of 15-20 false positive hCG results over the past 6 weeks for 10-15 patients while using the McKesson CONSULT® hCG Urine Test Cassettes. The customer reported several false positives were observed on lot 0001134798 and lots 0001176452 and 0001140630 yielded a few false positives. The customer reported the first event occurred 18MAR2026. Each patient provided urine and 2oz of urine was applied to each test. Due to the volume of sample being indicated, attempts were made to clarify this information however, they were unsuccessful. The results were read at 1 to 2 minutes and a false positive result was obtained. In some instances, repeat testing yielded an additional positive hCG result. Confirmatory blood testing was performed which yielded a negative result. It is currently unclear if every patient had confirmatory testing performed. The customer indicated both 10 patients were \"impacted\" and \"no patient impact\". No specific adverse health consequences were reported. Although requested, no further information was provided.\\Please note, this report is for patient 11.\\See MDRs 2027969-2026-00057, 2027969-2026-00058, 20279
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267382 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED 18MAR2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 15. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 2027969
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267381 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED (B)(6) 2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 12. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 20279
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267380 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED (B)(6) 2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 13. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 20279
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267379 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED 18MAR2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 14. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 2027969
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267377 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED (B)(6) 2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 3. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 202796
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267365 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED (B)(6) 2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 1. SEE MDRS 2027969-2026-00058, 2027969-2026-00059, 202796
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267364 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT: 0001134798 AND LOTS: 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED 18MAR2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 4. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 202796
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267363 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED (B)(6) 2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 9. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 202796
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267362 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED (B)(6) 2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 7. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 202796
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267361 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED 18MAR2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 6. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 2027969-
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267360 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED (B)(6) 2026. EACH PATIENT PROVIDED URINE AND 2 OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 5. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 20279
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267359 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED (B)(6) 2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 2. SEE MDRS 2027969-2026-00057, 2027969-2026-00059, 202796
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267358 ↗MalfunctionMay 24, 2026False Positive Result
THE DISTRIBUTOR REPORTED THAT A CUSTOMER RECEIVED A TOTAL OF 15-20 FALSE POSITIVE HCG RESULTS OVER THE PAST 6 WEEKS FOR 10-15 PATIENTS WHILE USING THE MCKESSON CONSULT® HCG URINE TEST CASSETTES. THE CUSTOMER REPORTED SEVERAL FALSE POSITIVES WERE OBSERVED ON LOT 0001134798 AND LOTS 0001176452 AND 0001140630 YIELDED A FEW FALSE POSITIVES. THE CUSTOMER REPORTED THE FIRST EVENT OCCURRED 18MAR2026. EACH PATIENT PROVIDED URINE AND 2OZ OF URINE WAS APPLIED TO EACH TEST. DUE TO THE VOLUME OF SAMPLE BEING INDICATED, ATTEMPTS WERE MADE TO CLARIFY THIS INFORMATION HOWEVER, THEY WERE UNSUCCESSFUL. THE RESULTS WERE READ AT 1 TO 2 MINUTES AND A FALSE POSITIVE RESULT WAS OBTAINED. IN SOME INSTANCES, REPEAT TESTING YIELDED AN ADDITIONAL POSITIVE HCG RESULT. CONFIRMATORY BLOOD TESTING WAS PERFORMED WHICH YIELDED A NEGATIVE RESULT. IT IS CURRENTLY UNCLEAR IF EVERY PATIENT HAD CONFIRMATORY TESTING PERFORMED. THE CUSTOMER INDICATED BOTH 10 PATIENTS WERE \"IMPACTED\" AND \"NO PATIENT IMPACT\". NO SPECIFIC ADVERSE HEALTH CONSEQUENCES WERE REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION WAS PROVIDED. PLEASE NOTE, THIS REPORT IS FOR PATIENT 8. SEE MDRS 2027969-2026-00057, 2027969-2026-00058, 2027969-
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267357 ↗MalfunctionMay 22, 2026False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT WITH THE MCKESSON CONSULT DIAGNOSTICS HCG URINE CASSETTE. THE PATIENT WAS BEING SEEN FOR A BIRTH CONTROL CONSULTATION. THE PATIENT HAD AN IUD THAT WAS PLACED ON (B)(6) 2025. THE PATIENT DECIDED TO HAVE THE IUD REMOVED BASED ON THE POSITIVE RESULT, HOWEVER THE PROVIDER WAS UNABLE TO REMOVE THE IUD AS THE STRINGS WERE NOT VISIBLE. QUALITATIVE SERUM HCG TESTING WAS PERFORMED (METHOD NOT SPECIFIED) AND THE RESULT WAS NEGATIVE. NO ADVERSE HEALTH OUTCOMES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25260387 ↗MalfunctionMay 22, 2026False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WHEN TESTING A PATIENT WITH THE MCKESSON CONSULT DIAGNOSTICS HCG URINE CASSETTE. THE PATIENT WAS BEING SEEN FOR A BIRTH CONTROL CONSULTATION. QUALITATIVE SERUM HCG TESTING WAS PERFORMED (METHOD NOT SPECIFIED) AND THE RESULT WAS NEGATIVE. NO ADVERSE HEALTH OUTCOMES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25260379 ↗
FDA clearance and classification
The FDA regulates product code JHI (visual, pregnancy hcg, prescription use) under 21 CFR 862.1155 as a Class II device. 21 CFR 862.1155 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the McKesson 5001
Has the McKesson 5001 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-3271-2026 (Class II), began on Aug 20, 2026: hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the McKesson 5001?
The FDA's MAUDE database lists 177 adverse event reports for this device, received between Nov 17, 2017 and Aug 31, 2026. In those reports, the most common specific device problems are “False Positive Result” (126), “False Negative Result” (46) and “Incorrect, Inadequate or Imprecise Result or Readings” (5); the most common patient problem is “Miscarriage” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the McKesson 5001?
None of the reports are classified as deaths, 5 as injuries and 172 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the McKesson 5001, and is it still sold?
The FDA's device database lists Mckesson Medical-Surgical Inc. as the company responsible for it (brand McKesson, model 5001). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Visual, Pregnancy Hcg, Prescription Use devices
- CARDINAL HEALTH B1077-23Cardinal Health 200, LLC · Visual, pregnancy hCG, prescription use1 recall154 reports
- McKesson 5000Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls91 reports
- Sure-Vue 4581405021Fisher Scientific Company L.L.C. · Visual, pregnancy hCG, prescription use1 recall90 reports
- McKesson 5002Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls36 reports
- Henry Schein 9004072Henry Schein, Inc. · Visual, pregnancy hCG, prescription use1 recall79 reports
- AlereTM 92210Alere San Diego, Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls24 reports
- MEDLINE MPH12025HCGMedline Industries, Inc. · Visual, pregnancy hCG, prescription use1 recall69 reports
- Henry Schein 9004073Henry Schein, Inc. · Visual, pregnancy hCG, prescription use1 recall63 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page