McKesson 32639Recalls & reported problems
Mckesson Medical-Surgical Inc. · Nasal Cannula ADULT, CONTOURED TIP 14’ (4.3 m) CRUSH RESISTANT TUBING | NO PAD DESIGN | SINGLE PATIENT USE · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20612479208500), received between Aug 3, 2026 and Aug 3, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Connection Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Connection Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 3, 2026Connection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50122867 ↗
FDA clearance and classification
The FDA regulates product code CAT (cannula, nasal, oxygen) under 21 CFR 868.5340 as a Class I device. 21 CFR 868.5340 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the McKesson 32639
Has the McKesson 32639 been recalled?
No FDA recall record in our data names this exact device identifier (DI 20612479208500), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the McKesson 32639?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Aug 3, 2026. That report lists device problem “Connection Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the McKesson 32639?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the McKesson 32639 FDA approved?
Devices of this type (product code CAT) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Mckesson Medical-Surgical Inc.
Who makes the McKesson 32639, and is it still sold?
The FDA's device database lists Mckesson Medical-Surgical Inc. as the company responsible for it (brand McKesson, model 32639). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Cannula, Nasal, Oxygen devices
- Fisher & Paykel Healthcare OPT844Fisher & Paykel Healthcare Limited · Cannula, Nasal, Oxygen39 reports
- Fisher and Paykel Healthcare OPT312Fisher & Paykel Healthcare Limited · Cannula, Nasal, Oxygen21 reports
- Fisher and Paykel Healthcare OPT314Fisher & Paykel Healthcare Limited · Cannula, Nasal, Oxygen17 reports
- O2 Cannula 001366Salter Labs · Cannula, Nasal, Oxygen9 reports
- N/A 16SOFT-4Salter Labs · CANNULA, NASAL, OXYGEN8 reports
- N/A 1053-7Salter Labs · CANNULA, NASAL, OXYGEN8 reports
- Fisher & Paykel Healthcare OPT842Fisher & Paykel Healthcare Limited · Cannula, Nasal, Oxygen7 reports
- O2 Cannula 001365Salter Labs · Cannula, Nasal, Oxygen7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page