Visao® 3318606Recalls & reported problems
Medtronic Xomed, Inc. · COOLING SET 3318606 2PK IPC VISAO · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20613994436355), received between Nov 22, 2019 and Nov 22, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Blocked Connection” (1 report).
1 other recall cover devices cleared under the same FDA submission (K081277); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Blocked Connection1
- Loose or Intermittent Connection1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 22, 2019Loose or Intermittent Connection; Blocked Connection; Loose or Intermittent Connection; Blocked Connection
A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT THE TUBING BECAME LOOSE AND AIR BUBBLE KEPT BEING INTRODUCED IN THE COOLANT TUBING WHICH LEAD TO THE HANDPIECE OVERHEATING PRIOR A PROCEDURE. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S). THE TUBING WAS DISCONNECTED PRIOR TO DRILLING. THERE WAS NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9361247 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081277. One submission often covers many models, so check the product and lot details in each record.
- Z-0961-2022 · Dec 15, 2021 · TerminatedSoftware 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND E
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.