DLP® Aortic Root Cannula with Flow-Guard™ Introducer 21012Recalls & reported problems

Medtronic, Inc. · CANNULA 21012 AR VENT FLOW 12GA 20PK 17L · Class 2

All models of DLP Aortic Root Cannula with Flow-Guard Introducer (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20613994495390), received between Aug 8, 2024 and Jan 6, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 5, 2025Class IOpen, Classified

    Unexpected loose material in the male luer used in the aortic root cannula has been identified. Potential harms when identified prior to use - procedure delay. Potential harm if not identified and used - stroke (reversible and irreversible).

    Z-1306-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Separation2
  • Crack1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 6, 2025Crack; Material Separation; Crack; Material Separation

    IT WAS REPORTED THAT DURING THE PRE-OPERATIVE INSPECTION OF A DLP AORTIC ROOT CANNULA WITH A VENT LINE, THE PLASTIC RETAINING NODULES THAT HOLD THE INSERTION NEEDLE INSIDE THE CANNULA WERE FOUND BROKEN OFF INSIDE THE BLUE CONNECTOR. THERE WAS NO DAMAGE TO ANY PACKAGING AND NO VISIBLE DAMAGE TO OTHER PARTS OF THE CANNULA. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21079264 ↗
  • MalfunctionAug 8, 2024Material Separation; Material Deformation; Material Separation; Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A DLP AORTIC ROOT CANNULA, IT WAS REPORTED THAT WHEN CANNULA WAS OPENED THE PLASTIC TIP WAS DAMAGED AND THE NEEDLE WAS LOOSE.THE DEVICE WAS NOT USED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19937146 ↗

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices