Skip to content

DLP® 31001Recalls & reported problems

Medtronic, Inc. · CANNULA 31001 IMA 1MM 1.8 20PK 17L · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20613994575313), received between Mar 5, 2026 and Mar 5, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K810820); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 19, 2024Open, Classified

    Potential for unsealed sterile packing.

    Z-1796-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Material Deformation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 5, 2026Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THE DLP I.M.A. CANNULAE, IT WAS REPORTED THAT INSTEAD OF HAVING A ROUNDED, ATRAUMATIC TIP, THE CANNULAE HAD A SHARP ELEMENT AT THE DISTAL END. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24517946 ↗
  • MalfunctionMar 5, 2026Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THE DLP I.M.A. CANNULAE, IT WAS REPORTED THAT INSTEAD OF HAVING A ROUNDED, ATRAUMATIC TIP, THE CANNULAE HAD A SHARP ELEMENT AT THE DISTAL END. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24517920 ↗
  • MalfunctionMar 5, 2026Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THE DLP I.M.A. CANNULAE, IT WAS REPORTED THAT INSTEAD OF HAVING A ROUNDED, ATRAUMATIC TIP, THE CANNULAE HAD A SHARP ELEMENT AT THE DISTAL END. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24517831 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K810820. One submission often covers many models, so check the product and lot details in each record.

  • Z-1799-2024 · Mar 19, 2024 · Open, Classified
    Medtronic Vessel Cannulae: a) DLP 2 mm Vessel Cannula Blunt Tip, Model Number 30004; b) DLP 3 mm Vessel Cannula Blunt Tip, Model Number 30003
  • Z-1316-2024 · Feb 7, 2024 · Open, Classified
    Medtronic DLP Vessel Cannula, Model Number REF 30000

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.