DLP® 30155Recalls & reported problems

Medtronic, Inc. · CANNULA 30155 CORONARY 90DEG SOFT 10 17L · Class 2

All models of DLP (151)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 20613994575962), received between Mar 3, 2025 and Mar 4, 2025 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Material Separation” (16 reports).

1 other recall cover devices cleared under the same FDA submission (K131269, K821149); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 16 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation16

Patient problems

  • No Clinical Signs, Symptoms or Conditions16

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT THE ISSUE WITH THE UNUSED DEVICE WAS DISCOVERED UPON CHECKING THE STOCK AFTER THE PROCEDURE. IT WAS REPORTED THAT THE JOINT AT THE TIP BETWEEN THE ADVANCED PLASTIC AND METAL ROTATES. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21515126 ↗
  • MalfunctionMar 4, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT THE ISSUE WITH THE UNUSED DEVICE WAS DISCOVERED UPON CHECKING THE STOCK AFTER THE PROCEDURE. IT WAS REPORTED THAT THE JOINT AT THE TIP BETWEEN THE ADVANCED PLASTIC AND METAL ROTATES. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21515097 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21504030 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503970 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503864 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503762 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503614 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503261 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503230 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503190 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503148 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503079 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503035 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503026 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21503009 ↗
  • MalfunctionMar 3, 2025Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE, AT INVENTORY CHECK OF A DLP HIGH FLOW CORONARY ARTERY OSTIAL CANNULA, IT WAS REPORTED THAT THE PLASTIC-METAL BOND AT THE TIP WAS TWIRLED. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP WAS NOT OBSTRUCTED. THE TIP DID NOT DETACH FROM THE CANNULA BODY. THE CANNULA WAS NOT HEATED OR COOLED PRIOR TO USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21502951 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131269, K821149. One submission often covers many models, so check the product and lot details in each record.

  • Z-1795-2024 · Mar 19, 2024 · Open, Classified
    Medtronic DLP Coronary Ostial Cannulae, Model Number 30050

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page