Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20613994878513), received between Jun 15, 2018 and Aug 16, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Material Puncture/Hole” (3 reports).
3 other recalls cover devices cleared under the same FDA submission (K845045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole3
- Air/Gas in Device1
- Crack1
- Material Deformation1
- Split1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 16, 2026Material Deformation
Medtronic received information that prior to use of a DLP Single Stage Venous Cannulae, it was reported that the distal tip of the venous cannula was ruptured. The device was replaced. There was no patient involvement, so no adverse effect occurred.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50268251 ↗MalfunctionSep 9, 2025Crack; Material Puncture/Hole; Crack; Material Puncture/Hole
IT WAS REPORTED THAT DURING USE THE SINGLE STAGE VENOUS CANNULA WAS CUT AT THE PLACE WHERE THE SURGEON LACQUERED THE VEIN AFTER BEING PLACED IN THE PATIENT'S INFERIOR VENA CAVA. THE USE OF THE DEVICE WAS UNKNOWN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23002865 ↗MalfunctionOct 18, 2021Material Puncture/Hole; Air/Gas in Device; Material Puncture/Hole; Air/Gas in Device
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A DLP SINGLE STAGE VENOUS CANNULA, THE CUSTOMER REPORTED THAT A BREAK/HOLE IN THE CANNULA WAS NOTED. THE CANNULA WAS SUTURED TO THE STERNAL RETRACTOR WITH THE 1ST SUTURE, APPROX. 5-7 CM UP FROM THE TIP. THIS CAUSED THE CANNULA TO SUCK AIR DURING USE. AFTER TWO HOURS OF PERFUSION AND AFTER WEANING THE PATIENT, THE STAFF INVOLVED SAW A LARGE HOLE IN THE VENOUS CANNULA. THE VENOUS CANNULA WAS ALMOST COMPLETELY DIVIDED INTO TWO PARTS. THE SUTURE WAS LOOSELY TIED TO THE CANNULA AND THE STERNUM RETRACTOR. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12647715 ↗InjuryOct 18, 2021Material Puncture/Hole; Material Puncture/Hole
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A DLP SINGLE STAGE VENOUS CANNULA, THE CUSTOMER REPORTED THAT A BREAK/HOLE IN THE CANNULA WAS NOTED. AFTER CANNULATION, WHEN THE CANNULA WAS ATTACHED TO THE STERNAL RETRACTOR WITH THE 1ST SUTURE, APPROX. 5-7 CM UP FROM THE TIP, A HOLE WAS IDENTIFIED IN THE CANNULA, WHICH CAUSED THE CANNULA TO SUCK AIR DURING USE.. THE DEVICE WAS NOT REPLACED, THE CANNULA WAS REATTACHED AND THE HOLE WAS COVERED WITH TAPE/PLASTIC TO PREVENT LEAKAGE. THE SUTURE WAS LOOSELY TIED TO THE CANNULA AND THE STERNUM RETRACTOR. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12647622 ↗MalfunctionJun 15, 2018Split; Split
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THE SINGLE STAGE VENOUS CANNULA, WHILE FIXING THE CANNULA WITH A SILK THREAD THE CANNULA BROKE-UP (SPLIT) NOT ALLOWING A SEALED ENVIRONMENT. THE PRODUCT WAS REPLACED. THERE WAS NO ADVERSE EFFECT ON THE PATIENT. THE PROCEDURE TOOK LONGER DUE THE CANNULA BEEN REPLACED.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 7604563 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K845045. One submission often covers many models, so check the product and lot details in each record.
- Z-0812-2012 · Dec 14, 2011 · TerminatedMedtronic, Single Stage Venous Cannula, Carmeda Bio-Active Surface Venous Cannula, Sterilized Using Ethylene Oxide, Rx Only, Nonpyrogenic. Model # 's CB66112, CB66114, CB66116,
- Z-0808-2012 · Dec 14, 2011 · TerminatedMedtronic DLP, DLP Single Stage Venous Cannulae. Rx Only, Sterile EO, Nonpyrogenic. Model #'s 66112, 66114, 66116, 66118, 66120, 66122, 66124, 66126, 66128, 66130, 66132, 66134,
- Z-1489-04 · Sep 15, 2004 · TerminatedMedtronic DLP VAD Venous Cannula for Ventricular Assist, Models 95036 and CB95036.
Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page