DLP® 66128Recalls & reported problems

Medtronic, Inc. · CANNULA 66128 SS VENOUS 28FR 10PK 17L · Class 2

All models of DLP (151)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20613994878575), received between Aug 20, 2025 and Aug 20, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K845045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 19, 2024Open, Classified

    Potential for unsealed sterile packing.

    Z-1805-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Deformation2
  • Mechanical Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2025Mechanical Problem; Material Deformation; Mechanical Problem; Material Deformation

    IT WAS REPORTED THAT DURING USE OF DLP SINGLE STAGE VENOUS CANNULAE, THE DEVICES WERE USED FOR BI-CAVAL CANNULATION FOR A MITRAL VALVE OPERATION. AT THE END OF THE PROCEDURE, IT WAS NOTICED THAT THERE WERE SIGNIFICANT MARKS ON ONE OF THE CANNULA CONSISTENT TO WHERE THE TUBING CLAMP HAD BEEN APPLIED. THE MARKS APPEARED CONSISTENT WITH THE PATTERN ON THE TUBING CLAMP. AS EXPERIENCED USERS OF THIS CANNULA, THE CUSTOMER HAD NOT OBSERVED THIS DEGREE OF INDENTATION ON THE CANNULA POST-PROCEDURE BEFORE. THE CUSTOMER SUGGESTED, BUT CANNOT CONFIRM, THAT THE TUBING ON THE CANNULA WAS SOFTER THAN USUAL. THE CANNULA WAS NOT PERFORATED, JUST MARKED. THE SAME METAL TUBING CLAMP WAS USED ON BOTH CANNULAE AND THE OTHER CANNULA (OF SAME LOT NUMBER) WAS NOT MARKED. THE CLAMP HAD BEEN USED FOR PREVIOUS OPERATIONS WITH THE SAME DEVICES. NO DAMAGE WAS SEEN TO ANY COMPONENTS OR PACKAGING BEFORE THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING THE AFFECTED CANNULA SO THERE WAS NO INTERRUPTION OR DELAY TO THE OPERATION. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULAE WERE NOT VISIBLY SOFTER. THE CANNULAE WERE NOT HEATED OR COOLED PRIOR TO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22851450 ↗
  • MalfunctionAug 20, 2025Mechanical Problem; Material Deformation; Mechanical Problem; Material Deformation

    IT WAS REPORTED THAT DURING USE OF DLP SINGLE STAGE VENOUS CANNULAE, THE DEVICES WERE USED FOR BI-CAVAL CANNULATION FOR A MITRAL VALVE OPERATION. AT THE END OF THE PROCEDURE, IT WAS NOTICED THAT THERE WERE SIGNIFICANT MARKS ON ONE OF THE CANNULA CONSISTENT TO WHERE THE TUBING CLAMP HAD BEEN APPLIED. THE MARKS APPEARED CONSISTENT WITH THE PATTERN ON THE TUBING CLAMP. AS EXPERIENCED USERS OF THIS CANNULA, THE CUSTOMER HAD NOT OBSERVED THIS DEGREE OF INDENTATION ON THE CANNULA POST-PROCEDURE BEFORE. THE CUSTOMER SUGGESTED, BUT CANNOT CONFIRM, THAT THE TUBING ON THE CANNULA WAS SOFTER THAN USUAL. THE CANNULA WAS NOT PERFORATED, JUST MARKED. THE SAME METAL TUBING CLAMP WAS USED ON BOTH CANNULAE AND THE OTHER CANNULA (OF SAME LOT NUMBER) WAS NOT MARKED. THE CLAMP HAD BEEN USED FOR PREVIOUS OPERATIONS WITH THE SAME DEVICES. NO DAMAGE WAS SEEN TO ANY COMPONENTS OR PACKAGING BEFORE THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING THE AFFECTED CANNULA SO THERE WAS NO INTERRUPTION OR DELAY TO THE OPERATION. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CANNULAE WERE NOT VISIBLY SOFTER. THE CANNULAE WERE NOT HEATED OR COOLED PRIOR TO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22851320 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K845045. One submission often covers many models, so check the product and lot details in each record.

  • Z-0812-2012 · Dec 14, 2011 · Terminated
    Medtronic, Single Stage Venous Cannula, Carmeda Bio-Active Surface Venous Cannula, Sterilized Using Ethylene Oxide, Rx Only, Nonpyrogenic. Model # 's CB66112, CB66114, CB66116,
  • Z-0808-2012 · Dec 14, 2011 · Terminated
    Medtronic DLP, DLP Single Stage Venous Cannulae. Rx Only, Sterile EO, Nonpyrogenic. Model #'s 66112, 66114, 66116, 66118, 66120, 66122, 66124, 66126, 66128, 66130, 66132, 66134,
  • Z-1489-04 · Sep 15, 2004 · Terminated
    Medtronic DLP VAD Venous Cannula for Ventricular Assist, Models 95036 and CB95036.

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →