DLP® 68112Recalls & reported problems
Medtronic, Inc. · CANNULA 68112 SS VEN MALL 12FR 10PK 17L · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20613994879473), received between Dec 6, 2024 and Dec 6, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Deformation” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K022272); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 6, 2024Obstruction of Flow; Material Deformation; Obstruction of Flow; Material Deformation
IT WAS REPORTED THAT PRIOR TO USE OF A DLP MALLEABLE SINGLE STAGE VENOUS CANNULA IT WAS FOUND TO BE DAMAGED/DEFORMED WITH INDENTATIONS IN THE BODY. THE LOCATION OF THE DAMAGE WAS SPECIFIED, AND IT WAS NOTED THAT THESE DENTS MIGHT HAVE IMPEDED BLOOD FLOW IF THE DEVICE HAD BEEN USED. THE PACKAGING WAS NOT DAMAGED, AND NO OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER WERE AFFECTED. THE DEVICE WAS NOT USED, AND IT WAS REPLACED TO COMPLETE THE CASE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20862081 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K022272. One submission often covers many models, so check the product and lot details in each record.
- Z-1805-2024 · Mar 19, 2024 · Open, ClassifiedMedtronic Venous Cannulae: a) DLP 20 Fr. Malleable Single Stage Venous Cannula, Model Number 68120, b) DLP 24 Fr. Malleable Single Stage Venous Cannula, Model Number 68124, c)
- Z-0810-2012 · Dec 14, 2011 · TerminatedMedtronic, DLP, DLP Malleable Single Stage Venous Cannulae. Rx Only, Sterilized Using Ethylene Oxide, Nonpyrogenic. Model #'s 68112, 68114, 68116, 68118, 68120, 68122, 68124, 68
- Z-0812-2012 · Dec 14, 2011 · TerminatedMedtronic, Single Stage Venous Cannula, Carmeda Bio-Active Surface Venous Cannula, Sterilized Using Ethylene Oxide, Rx Only, Nonpyrogenic. Model # 's CB66112, CB66114, CB66116,
Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page