Skip to content

MC2® 91236Recalls & reported problems

Medtronic, Inc. · CANNULA 91236 MC2 VEN 36/46FR 10PK 17L · Class 2

All models of MC2 (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20613994887355), received between Feb 28, 2023 and Nov 28, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Contamination” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination1
  • Material Deformation1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 28, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING THE MC2 TWO STAGE VENOUS CANNULA, THE SURGEON CANNULATED WITH THE CANNULA AND OBSERVED THAT IT WAS STUCK UPON ATTEMPTING TO REMOVE THE STYLET. IT WOULD NOT COME OUT OF THE CANNULA. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23670049 ↗
  • MalfunctionMay 12, 2025Contamination; Contamination

    IT WAS REPORTED THAT PRIOR TO USE OF AN MC2 TWO STAGE VENOUS CANNULA FOREIGN MATTER (FM) WAS OBSERVED IN THE DEVICE. THE FM APPEARS TO BE A HAIR. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22000571 ↗
  • InjuryFeb 28, 2023Material Deformation; Material Deformation

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A MC2 TWO STAGE VENOUS CANNULA, DURING CARDIOPULMONARY BYPASS, DRAINAGE DIFFICULTIES WERE OBSERVED AT THE LEVEL OF THE VENOUS RETURN. THE CUSTOMER STATED THAT CORRECTIVE MEASURES WERE UNDERTAKEN WHICH WERE OPTIMIZATION OF VASCULAR FILLING, TRANSFUSION AND SUPPLY OF PRESSOR AMINES. THE CUSTOMER OBSERVED THAT WHEN CARDIOPULMONARY BYPASS WAS STOPPED AND DURING DECANNULATION A BEND WAS OBSERVED ON THE CANNULA AT THE LEVEL OF THE ATRIAL POSITIONING EYELET. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO IMPACT TO THE PATIENT AFTER THE PROCEDURE. THERE WAS NO DAMAGE TO THE PACKAGING (OUTER BOX, INNER PACKAGING OR STERILE BARRIER). NO OTHER DEVICES WERE DAMAGED IN THE SAME BOX/SHIPPING CONTAINER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16452359 ↗

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page