MPS Console 5001000Recalls & reported problems
Quest Medical Inc · Myocardial Protection System Console, 110v · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20634624501007), received between Sep 14, 2016 and May 18, 2017 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K953838); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message2
- Detachment Of Device Component1
- Device Operates Differently Than Expected1
- Improper Device Output1
- Intermittent Continuity1
Patient problems
- No Known Impact Or Consequence To Patient4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2017Improper Device Output
THE HOSPITAL PERFUSIONIST REPORTED AN ISSUE ENCOUNTERED WITH THE MPS CONSOLE DURING USE. THE REPORT STATED THAT THE CONSOLE DISPLAYED REALLY LOW, AND SOMETIMES ZERO, PRESSURE DURING DELIVERY. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE ALLEGED ISSUE. THE CONSOLE WILL BE RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6574542 ↗MalfunctionApr 4, 2017Detachment Of Device Component; Intermittent Continuity
THE HOSPITAL EQUIPMENT COORDINATOR REPORTED AN ISSUE ENCOUNTERED WITH THE MPS CONSOLE DURING USE. THE REPORT STATED THAT THE CONSOLE HAD BEEN INTERMITTENTLY POPPING THE VENT VALVE WHILE DELIVERY CARDIOPLEGIA. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE ALLEGED EVENT. A FIELD SERVICE ENGINEER WILL TRAVEL TO THE USER FACILITY TO EVALUATE THE CONSOLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6457761 ↗MalfunctionMar 2, 2017Device Displays Incorrect Message
THE HOSPITAL PERFUSIONIST REPORTED AN ISSUE WITH THE MPS CONSOLE DURING USE. SHE REPORTED THAT THE CONSOLE ALARMED AND DISPLAYED AN ERROR CODE FOR "INTERNAL TIMING ERROR" DURING THE INDUCTION DOSE OF ARREST AGENT. THE REPORT STATED THAT THE CONSOLE DISPLAYED THE ERROR FOUR TIMES BEFORE REMAINING ON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE ALLEGED ISSUE. THE CONSOLE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6372922 ↗MalfunctionSep 14, 2016Device Displays Incorrect Message; Device Operates Differently Than Expected
THE HOSPITAL BIOMEDICAL ENGINEER REPORTED AN ISSUE WITH THE MPS CONSOLE DURING USE. HE REPORTED FOR PERFUSION THAT THE CONSOLE WOULD REPEATEDLY DISPLAY AN ERROR CODE AND WOULD OPEN THE VENT VALVE DURING CARDIOPLEGIA DELIVERY. THE REPORT STATED THAT THE CONSOLE WAS USED TO COMPLETE THE PROCEDURE WITH THE PERFUSIONIST MANUALLY CLAMPING THE VENT LINE CLOSED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE ALLEGED ISSUE. THE CONSOLE WILL BE EVALUATED BY A FIELD SERVICE ENGINEER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5950956 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K953838. One submission often covers many models, so check the product and lot details in each record.
- Z-0361-2016 · Oct 28, 2015 · TerminatedModel 5001102 MPS Delivery Set w/ arrest agent and additive cassettes, heat exchanger &10 ft. delivery tubing; Model 5001102-AS MPS Delivery Set with 6 ft. delivery tubing; Model 7
- Z-1924-2011 · Jul 29, 2010 · TerminatedQuest MPS and MPS 2 Consoles, Quest Medical, Inc., An Atrion company. Single software controlled device that incorporates a pump, a heater/cooler, temperature monitoring, pressure
- Z-1037-03 · May 23, 2003 · TerminatedQuest Myocardial Protection System (MPS) Arrest/Additive Cassettes
Other Heat-Exchanger, Cardiopulmonary Bypass devices
- Capiox® FX Advance Hollow Fiber Oxygenator w Integrated AF w Hardshell Res. 3CX*FX25RECTerumo Cardiovascular Systems Corporation · Heat-Exchanger, Cardiopulmonary Bypass3 deaths267 reports
- QUADROX-i HMO 51000-USA #SQUADROX-i Small Ad.+Filter /Small Adult. FilterMaquet Cardiopulmonary AG · HEAT-EXCHANGER, CARDIOPULMONARY BYPASS2 recalls0 reports
- QUADROX-i Pediatric, diffusion, BIOLINEMaquet Cardiopulmonary AG · HEAT-EXCHANGER, CARDIOPULMONARY BYPASS2 recalls0 reports
- QUADROX-iD BEQ-HMOD70000-USA #SQUADROX-iD Ad./coated/ no.Fil.Maquet Cardiopulmonary AG · HEAT-EXCHANGER, CARDIOPULMONARY BYPASS2 recalls0 reports
- QUADROX-i BEQ-HMO 71000-USA #SQUADROX-i Adult+BIOLINE +FilterMaquet Cardiopulmonary AG · HEAT-EXCHANGER, CARDIOPULMONARY BYPASS2 recalls0 reports
- MPS2 Console 5201260Quest Medical Inc · Heat-Exchanger, Cardiopulmonary Bypass1 death67 reports
- Quantum PureFlow Standard Heat Exchanger High Flow 3/8 W HX55V-S0WSpectrum Medical Srl · Heat-Exchanger, Cardiopulmonary Bypass1 recall3 reports
- Sarns™ Cardioplegia Set 165720Terumo Cardiovascular Systems Corporation · Heat-Exchanger, Cardiopulmonary Bypass1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.