Halyard 13904Recalls & reported problems
Avanos Medical, Inc. · HALYARD* Oral Care Kit · Class 2
All models of Halyard (232)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 20680651139041), received between Apr 11, 2017 and Dec 28, 2017 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Detachment Of Device Component7
- Detachment of Device or Device Component3
- Break2
Patient problems
- No Known Impact Or Consequence To Patient12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 28, 2017Detachment Of Device Component; Detachment Of Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT PATIENTS. THIS IS THE SECOND OF FIVE REPORTS. REFER TO 8030647-2017-00191 FOR THE FIRST PATIENT. REFER TO 8030647-2017-00193 FOR THE THIRD PATIENT. REFER TO 8030647-2017-00194 FOR THE FOURTH PATIENT. REFER TO 8030647-2017-00195 FOR THE FIFTH PATIENT. IT WAS REPORTED THE SPONGE SWAB TIP DETACHED FROM THE SUCTION CATHETER DURING ORAL CARE IN PATIENT¿S MOUTH. THE EVENT OCCURRED WITHOUT TRAUMA TO SWAB TIP (IE. NO BITING FROM PATIENT). THE SWAB TIP HAD TO BE RETRIEVED FROM PATIENT¿S MOUTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7151200 ↗MalfunctionDec 28, 2017Detachment Of Device Component; Detachment Of Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT PATIENTS. THIS IS THE FIFTH OF FIVE REPORTS. REFER TO 8030647-2017-00191 FOR THE FIRST PATIENT. REFER TO 8030647-2017-00192 FOR THE SECOND PATIENT. REFER TO 8030647-2017-00193 FOR THE THIRD PATIENT. REFER TO 8030647-2017-00194 FOR THE FOURTH PATIENT. IT WAS REPORTED THE SPONGE SWAB TIP DETACHED FROM THE SUCTION CATHETER DURING ORAL CARE IN PATIENT¿S MOUTH. THE EVENT OCCURRED WITHOUT TRAUMA TO SWAB TIP (IE. NO BITING FROM PATIENT). THE SWAB TIP HAD TO BE RETRIEVED FROM PATIENT¿S MOUTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7150889 ↗MalfunctionDec 28, 2017Detachment Of Device Component; Detachment Of Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT PATIENTS. THIS IS THE FIRST OF FIVE REPORTS. REFER TO 8030647-2017-00192 FOR THE SECOND PATIENT. REFER TO 8030647-2017-00193 FOR THE THIRD PATIENT. REFER TO 8030647-2017-00194 FOR THE FOURTH PATIENT. REFER TO 8030647-2017-00195 FOR THE FIFTH PATIENT. IT WAS REPORTED THE SPONGE SWAB TIP DETACHED FROM THE SUCTION CATHETER DURING ORAL CARE IN PATIENT¿S MOUTH. THE EVENT OCCURRED WITHOUT TRAUMA TO SWAB TIP (IE. NO BITING FROM PATIENT). THE SWAB TIP HAD TO BE RETRIEVED FROM PATIENT¿S MOUTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7150887 ↗MalfunctionDec 28, 2017Detachment Of Device Component; Detachment Of Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT PATIENTS. THIS IS THE THIRD OF FIVE REPORTS. REFER TO 8030647-2017-00191 FOR THE FIRST PATIENT. REFER TO 8030647-2017-00192 FOR THE SECOND PATIENT. REFER TO 8030647-2017-00194 FOR THE FOURTH PATIENT. REFER TO 8030647-2017-00195 FOR THE FIFTH PATIENT. IT WAS REPORTED THE SPONGE SWAB TIP DETACHED FROM THE SUCTION CATHETER DURING ORAL CARE IN PATIENT¿S MOUTH. THE EVENT OCCURRED WITHOUT TRAUMA TO SWAB TIP (IE. NO BITING FROM PATIENT). THE SWAB TIP HAD TO BE RETRIEVED FROM PATIENT¿S MOUTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7150885 ↗MalfunctionDec 28, 2017Detachment Of Device Component; Detachment Of Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT PATIENTS. THIS IS THE FOURTH OF FIVE REPORTS. REFER TO 8030647-2017-00191 FOR THE FIRST PATIENT. REFER TO 8030647-2017-00192 FOR THE SECOND PATIENT. REFER TO 8030647-2017-00193 FOR THE THIRD PATIENT. REFER TO 8030647-2017-00195 FOR THE FIFTH PATIENT. IT WAS REPORTED THE SPONGE SWAB TIP DETACHED FROM THE SUCTION CATHETER DURING ORAL CARE IN PATIENT¿S MOUTH. THE EVENT OCCURRED WITHOUT TRAUMA TO SWAB TIP (IE. NO BITING FROM PATIENT). THE SWAB TIP HAD TO BE RETRIEVED FROM PATIENT¿S MOUTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7150874 ↗MalfunctionNov 30, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
THE INITIAL REPORT CONTAINED STATEMENTS: REFER TO 8030647-2017-00151 FOR THE SECOND PATIENT. REFER TO 8030647-2017-00152 FOR THE THIRD PATIENT. HOWEVER, THE CORRECTED STATEMENTS SHOULD READ: REFER TO 8030647-2017-00150 FOR THE FIRST PATIENT. REFER TO 8030647-2017-00151 FOR THE SECOND PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7072429 ↗MalfunctionNov 30, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 8030647-2017-00150 FOR THE FIRST PATIENT. REFER TO 8030647-2017-00152 FOR THE THIRD PATIENT. IT WAS REPORTED THAT THE SWAB TIP ON THE SUCTION HANDLE CAME OFF IN THE PATIENT'S MOUTH. THIS EVENT HAPPENED THREE TIMES IN TWO DAYS. THE SWAB TIP WAS ABLE TO BE RETRIEVED FROM THE PATIENT'S MOUTHS WITHOUT INJURY TO THE PATIENTS. NO FURTHER INFORMATION WAS RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7072407 ↗MalfunctionNov 30, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 8030647-2017-00151 FOR THE SECOND PATIENT. REFER TO 8030647-2017-00152 FOR THE THIRD PATIENT. IT WAS REPORTED THAT THE SWAB TIP ON THE SUCTION HANDLE CAME OFF IN THE PATIENT'S MOUTH. THIS EVENT HAPPENED THREE TIMES IN TWO DAYS. THE SWAB TIP WAS ABLE TO BE RETRIEVED FROM THE PATIENT'S MOUTHS WITHOUT INJURY TO THE PATIENTS. NO FURTHER INFORMATION WAS RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7072399 ↗MalfunctionNov 28, 2017Break; Break
IT WAS REPORTED THAT A SWAB SPONGE ON THE END OF THE TOOTHETTES WERE COMING OFF THE SUCTION HANDLE IN THE PATIENT'S MOUTH. THE CLINICIAN SUCCESSFULLY RETRIEVED THE SWAB FROM THE PATIENT'S MOUTH. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7064622 ↗MalfunctionJun 21, 2017Detachment Of Device Component; Detachment Of Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 3006646024-2017-00013 FOR THE SECOND REPORT. IT WAS REPORTED THAT, "BRUSHES ARE BREAKING OFF IN THE PATIENT'S MOUTH. THE SUCTION HANDLE COMES APART WHILE IN USE AND THE PATIENT'S LIPS ARE MORE DRY THAN BEFORE. THE SUCTION TIP PROVIDES THE STAFF WITH DIFFICULTY IN SUCTIONING OUT THE ENTIRE ORALPHYRNAGEL AREA." NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6659104 ↗MalfunctionJun 21, 2017Detachment Of Device Component; Detachment Of Device Component
HALYARD RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 3006646024-2017-00012 FOR THE FIRST REPORT. IT WAS REPORTED THAT, "BRUSHES ARE BREAKING OFF IN THE PATIENT'S MOUTH. THE SUCTION HANDLE COMES APART WHILE IN USE AND THE PATIENT'S LIPS ARE MORE DRY THAN BEFORE. THE SUCTION TIP PROVIDES THE STAFF WITH DIFFICULTY IN SUCTIONING OUT THE ENTIRE ORALPHARYNGEAL AREA." NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6659082 ↗MalfunctionApr 11, 2017Break; Break
IT WAS REPORTED WHILE PERFORMING ORAL CARE ON A PATIENT WHO MOUTH WAS DRY AND STICKY THE SUCTION SWAB BROKE OFF ON THE CURVED AREA AND INTO THE PATIENT¿S MOUTH. THE SPEECH PATHOLOGIST WAS ABLE TO REMOVE THE SWAB FROM THE PATIENT¿S MOUTH. ADDITIONAL INFORMATION WAS RECEIVED ON (B)(6) 2017 STATED THE SWAB BROKE WITH THE ENTIRE STEM FALLING INTO THE PATIENT¿S ORAL CAVITY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6482352 ↗
FDA clearance and classification
The FDA regulates product code JOL (catheter and tip, suction) under 21 CFR 880.6740 as a Class II device. 21 CFR 880.6740 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Halyard 13904
Has the Halyard 13904 been recalled?
No FDA recall record in our data names this exact device identifier (DI 20680651139041), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Halyard 13904?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Apr 11, 2017 and Dec 28, 2017. In those reports, the most common specific device problems are “Detachment Of Device Component” (7), “Detachment of Device or Device Component” (3) and “Break” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Halyard 13904?
None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Halyard 13904 FDA approved?
Devices of this type (product code JOL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Avanos Medical, Inc.
Who makes the Halyard 13904, and is it still sold?
The FDA's device database lists Avanos Medical, Inc. as the company responsible for it (brand Halyard, model 13904). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER AND TIP, SUCTION devices
- J-VAC 2160Ethicon Inc. · CATHETER AND TIP, SUCTION152 reports
- Wayne G56537Cook Incorporated · CATHETER AND TIP, SUCTION2 recalls33 reports
- Bard® Suction Connector EN0066000C. R. Bard, Inc. · CATHETER AND TIP, SUCTION1 recall6 reports
- ARROW IPN037088Teleflex Incorporated · CATHETER AND TIP, SUCTION1 recall0 reports
- MEDLINE DYND41478Medline Industries, Inc. · CATHETER AND TIP, SUCTION1 recall0 reports
- MEDLINE DYND41476Medline Industries, Inc. · CATHETER AND TIP, SUCTION1 recall0 reports
- ARROW IPN039282Teleflex Incorporated · CATHETER AND TIP, SUCTION1 recall0 reports
- MEDLINE DYND41474Medline Industries, Inc. · CATHETER AND TIP, SUCTION1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page