Halyard 34145Recalls & reported problems

O&M Halyard, Inc. · QUICK CHECK,H400,STRWRP,-,48X48,48 · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20680651341451), received between Sep 21, 2018 and Jan 30, 2026 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Puncture/Hole” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Puncture/Hole4
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 30, 2026Appropriate Device Problem Term/Code Not Available

    WRAPPING SHEETS EXHIBITED MICRO-PERFORATIONS AFTER STERILIZATION AND WERE ALREADY DAMAGED UPON REMOVAL FROM THE PACKAGE PRIOR TO USE. AS A RESULT, MULTIPLE SURGICAL PROCEDURES WERE DELAYED AFTER PATIENTS WERE ALREADY POSITIONED ON THE OPERATING TABLE. ADDITIONAL DETAILS (INCLUDING PATIENT DETAILS FOR THOSE AFFECTED) HAVE BEEN REQUESTED; HOWEVER, NO FURTHER DETAILS HAVE BEEN PROVIDED TO-DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24222485 ↗
  • InjuryJul 1, 2024Material Puncture/Hole; Material Puncture/Hole

    HOLES WERE FOUND ON THE STERILIZATION WRAP WHICH CAUSED A DELAY IN PROCEDURE. THE PROCEDURE WAS POSTPONED TO LATER IN THE DAY. THERE WERE NO HOLES OBSERVED BEFORE OR AFTER WRAPPING. THE TRAYS HAD BEEN STORED FOR A WEEK AND WERE PLACED AND STORED CORRECTLY ACCORDING TO CUSTOMER. THEY WERE ALSO OBSERVED TO HAVE NO HOLES OR DEFECTS BEFORE SENDING THEM TO THEATRE FOR THE SCHEDULED OPERATION. THE PATIENT HAD NOT BEEN SEDATED AT THE TIME OF DISCOVERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19646210 ↗
  • InjuryMay 29, 2024Material Puncture/Hole; Material Puncture/Hole

    THERE WERE HOLES IN THE WRAP NOTED AFTER STERILIZATION. SURGICAL PROCEDURE WAS CANCELLED. THIS WAS A PODIATRY PROCEDURE. THE CUSTOMER STATED THAT BOTH THE INITIAL TRAY FOR USE IN THE PROCEDURE AND THE BACKUP TRAY EXHIBITED HOLES. THE ISSUE WAS NOTICED PRIOR TO USE WITH THE PATIENT; HOWEVER, THE PROCEDURE WAS DELAYED AND CANCELLED FOR THE DAY. THE PATIENT WAS TRANSFERRED TO ANOTHER FACILITY TO COMPLETE THE PROCEDURE. THE HOSPITAL INCURRED THE COST OF THE PROCEDURE. THERE WAS NO INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19415677 ↗
  • InjuryApr 11, 2024Material Puncture/Hole; Material Puncture/Hole

    HIP REPLACEMENT PROCEDURE WAS SCHEDULED FOR PATIENT ON (B)(6) 2024. CLINICIAN WAS UNABLE TO PERFORM PROCEDURE AT SCHEDULED TIME BECAUSE HOLES WERE FOUND IN THE STERILIZATION WRAP COVERING SURGICAL TRAY. PATIENT WAS NOT IMPACTED OTHER THAN ANNOYED BY HAVING TO RESCHEDULE PROCEDURE. THE PATIENT WAS NOT SEDATED AT THE TIME OF DISCOVERY. THE COMPLAINT SITE INDICATED THE HOLES WERE NOT NOTED PRIOR TO WRAPPING INSTRUMENT TRAY IN CSSD. THE TRAYS HAD BEEN STORED BY THEM FOR A WEEK PRIOR TO CASE. O&M SALES REP IS CONDUCTING TRAINING WITH THE SITE ON BEST PRACTICES FOR MANAGING STERILIZATION TRAYS TO MINIMIZE RISK OF CAUSING HOLES IN WRAP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19094107 ↗
  • MalfunctionSep 21, 2018Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED ON 25-SEPT-2018 THAT THE SAMPLE WAS NOT AVAILABLE FOR RETURN. IT WAS REPORTED THAT THERE WERE HOLES FOND ON THE WHITE SIDE OF THE STERILIZATION WRAP. THE HOLES WERE DETECTED AFTER USE. ADDITIONAL INFORMATION WAS REPORTED ON 30-AUG-2018 THAT STATES, "THE OBJECTS WHICH WRAPPED WITH THE PRODUCT IN QUESTION AND STERILIZED WERE USED TO THE PATIENT BUT NO PATIENT INJURY REPORTED"

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7897689 ↗

Other Wrap, sterilization devices

All Wrap, sterilization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page