Skip to content

Halyard 34164Recalls & reported problems

O&M Halyard, Inc. · QUICK CHECK,H600,STRWRP,-,45X45,48 · Class 2

Adverse event reports
11
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 20680651341642), received between May 30, 2018 and Apr 25, 2025 — classified as 7 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Material Discolored” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 6 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Discolored6
  • Material Integrity Problem2
  • Appropriate Device Problem Term/Code Not Available1
  • Hole In Material1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2025Material Integrity Problem; Material Integrity Problem

    THE CUSTOMER REPORTED RUST STAINING ON WRAP WHERE THE TRAY WAS SITTING. ADDITIONAL INFORMATION REPORTED ON APRIL 15, 2025, STATES, THERE WERE A COUPLE OF PROCEDURE CANCELLATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21911174 ↗
  • MalfunctionApr 25, 2025Material Integrity Problem; Material Integrity Problem

    THE CUSTOMER REPORTED RUST STAINING ON WRAP WHERE THE TRAY WAS SITTING. ADDITIONAL INFORMATION REPORTED ON (B)(6) 2025 STATES, THERE WERE A COUPLE OF PATIENT PROCEDURE CANCELLATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21911169 ↗
  • InjuryJun 28, 2024Material Discolored; Material Discolored

    INCIDENT SIX. STERILIZATION WRAP EXHIBITED STAINING. FACILITY HAS USED PRODUCT FOR SEVERAL YEARS. THE FACILITY HAS NOT CHANGED THEIR STERILIZATION PROCESS OR PARAMETERS. TRAYS WERE REJECTED BY THE OR AND SIX SURGICAL CASES WERE CANCELLED. ADDITIONALLY, SEVERAL CASES WERE DELAYED UP TO 2-3 HOURS. NO INSTRUMENTS FROM THE TRAYS WERE USED ON PATIENTS. IN TWO SURGICAL CASES, 2 PATIENTS RECEIVED A BLOCK PRIOR TO THE SURGERY BEING CANCELLED. THE PATIENTS WERE PARALYZED FOR SIX HOURS FROM THE WAIST DOWN UNTIL THE MEDICATION WORE OFF. THE CASES WERE RE-SCHEDULED. THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED FOR THE PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19635022 ↗
  • InjuryJun 28, 2024Material Discolored; Material Discolored

    INCIDENT THREE: STERILIZATION WRAP EXHIBITED STAINING. FACILITY HAS USED PRODUCT FOR SEVERAL YEARS. THE FACILITY HAS NOT CHANGED THEIR STERILIZATION PROCESS OR PARAMETERS. TRAYS WERE REJECTED BY THE OR AND SIX SURGICAL CASES WERE CANCELLED. ADDITIONALLY, SEVERAL CASES WERE DELAYED UP TO 2-3 HOURS. NO INSTRUMENTS FROM THE TRAYS WERE USED ON PATIENTS. IN TWO SURGICAL CASES, 2 PATIENTS RECEIVED A BLOCK PRIOR TO THE SURGERY BEING CANCELLED. THE PATIENTS WERE PARALYZED FOR SIX HOURS FROM THE WAIST DOWN UNTIL THE MEDICATION WORE OFF. THE CASES WERE RE-SCHEDULED. THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED FOR THE PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19634470 ↗
  • InjuryJun 28, 2024Material Discolored; Material Discolored

    INCIDENT TWO: STERILIZATION WRAP EXHIBITED STAINING. FACILITY HAS USED PRODUCT FOR SEVERAL YEARS. THE FACILITY HAS NOT CHANGED THEIR STERILIZATION PROCESS OR PARAMETERS. TRAYS WERE REJECTED BY THE OR AND SIX SURGICAL CASES WERE CANCELLED. ADDITIONALLY, SEVERAL CASES WERE DELAYED UP TO 2-3 HOURS. NO INSTRUMENTS FROM THE TRAYS WERE USED ON PATIENTS. IN TWO SURGICAL CASES, 2 PATIENTS RECEIVED A BLOCK PRIOR TO THE SURGERY BEING CANCELLED. THE PATIENTS WERE PARALYZED FOR SIX HOURS FROM THE WAIST DOWN UNTIL THE MEDICATION WORE OFF. THE CASES WERE RE-SCHEDULED. THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED FOR THE PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19634457 ↗
  • InjuryJun 28, 2024Material Discolored; Material Discolored

    INCIDENT FIVE. STERILIZATION WRAP EXHIBITED STAINING. FACILITY HAS USED PRODUCT FOR SEVERAL YEARS. THE FACILITY HAS NOT CHANGED THEIR STERILIZATION PROCESS OR PARAMETERS. TRAYS WERE REJECTED BY THE OR AND SIX SURGICAL CASES WERE CANCELLED. ADDITIONALLY, SEVERAL CASES WERE DELAYED UP TO 2-3 HOURS. NO INSTRUMENTS FROM THE TRAYS WERE USED ON PATIENTS. IN TWO SURGICAL CASES, 2 PATIENTS RECEIVED A BLOCK PRIOR TO THE SURGERY BEING CANCELLED. THE PATIENTS WERE PARALYZED FOR SIX HOURS FROM THE WAIST DOWN UNTIL THE MEDICATION WORE OFF. THE CASES WERE RE-SCHEDULED. THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED FOR THE PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19634190 ↗
  • InjuryJun 28, 2024Material Discolored; Material Discolored

    INCIDENT FOUR. STERILIZATION WRAP EXHIBITED STAINING. FACILITY HAS USED PRODUCT FOR SEVERAL YEARS. THE FACILITY HAS NOT CHANGED THEIR STERILIZATION PROCESS OR PARAMETERS. TRAYS WERE REJECTED BY THE OR AND SIX SURGICAL CASES WERE CANCELLED. ADDITIONALLY, SEVERAL CASES WERE DELAYED UP TO 2-3 HOURS. NO INSTRUMENTS FROM THE TRAYS WERE USED ON PATIENTS. IN TWO SURGICAL CASES, 2 PATIENTS RECEIVED A BLOCK PRIOR TO THE SURGERY BEING CANCELLED. THE PATIENTS WERE PARALYZED FOR SIX HOURS FROM THE WAIST DOWN UNTIL THE MEDICATION WORE OFF. THE CASES WERE RE-SCHEDULED. THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED FOR THE PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19634149 ↗
  • InjuryJun 28, 2024Material Discolored; Material Discolored

    INCIDENT ONE. STERILIZATION WRAP EXHIBITED STAINING. FACILITY HAS USED PRODUCT FOR SEVERAL YEARS. THE FACILITY HAS NOT CHANGED THEIR STERILIZATION PROCESS OR PARAMETERS. TRAYS WERE REJECTED BY THE OR AND SIX SURGICAL CASES WERE CANCELLED. ADDITIONALLY, SEVERAL CASES WERE DELAYED UP TO 2-3 HOURS. NO INSTRUMENTS FROM THE TRAYS WERE USED ON PATIENTS. IN TWO SURGICAL CASES, 2 PATIENTS RECEIVED A BLOCK PRIOR TO THE SURGERY BEING CANCELLED. THE PATIENTS WERE PARALYZED FOR SIX HOURS FROM THE WAIST DOWN UNTIL THE MEDICATION WORE OFF. THE CASES WERE RE-SCHEDULED. THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED FOR THE PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19634039 ↗
  • InjurySep 14, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    THE FACILITY HAS EXPERIENCED HOLES IN THE WRAP POST STERILIZATION. THEY ARE USING CORNER PROTECTORS, AND THE HOLES ARE OCCURRING ON THE MIDDLE AND SIDES OF THE WRAPS ON THE WHITE SIDE. THE CUSTOMER CONFIRMED THE WEIGHT OF THE TRAYS ARE 17-25 POUNDS AND THEY ARE USING THE WRAPS ACCORDING TO THE INSTRUCTIONS FOR USE. THE CUSTOMER STATED THEY EXPERIENCED A DELAY IN A PROCEDURE THAT LAST LESS THAN A HALF HOUR. IT WAS RESOLVED BY TAKING A TRAY FROM ANOTHER CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17749067 ↗
  • MalfunctionMay 23, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    INCIDENT ONE. CUSTOMER REPORTED THEY HAVE EXPERIENCED WET PACKS. THIS RESULTED IN PATIENT DELAY IN PROCEDURE. ADDITIONAL INCIDENT DETAILS HAVE BEEN REQUESTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 16989423 ↗
  • MalfunctionMay 30, 2018Hole In Material; Hole In Material

    IT WAS REPORTED BY THE CUSTOMER THAT A HOLE WAS DISCOVERED IN A STERILIZATION WRAP POST OR USE. ADDITIONAL INFORMATION WAS RECEIVED (B)(6) 2018 FROM CUSTOMER THAT STATES, "THERE WAS NO PATIENT HARM AND THE HOLE WAS SEEN BEFORE PATIENT CARE STARTED."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7553837 ↗

Other Wrap, sterilization devices

All Wrap, sterilization devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.