Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 20680651481393), received between Jun 6, 2023 and Aug 20, 2026 — classified as 3 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Product Quality Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Product Quality Problem4
- Flaked2
- Material Integrity Problem2
- Material Discolored1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- No Patient Involvement1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 20, 2026Flaked
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50351744 ↗MalfunctionNov 6, 2025Product Quality Problem; Product Quality Problem
THE SURGERY DEPARTMENT AT COMMUNITY HOSPITAL REPORTED AN ISSUE INVOLVING HALYARD ABSORBENT TRAY LINERS LEAVING DEBRIS INSIDE INSTRUMENT KITS AFTER PROCESSING THROUGH THE STERILE PROCESSING DEPARTMENT (SPD). THE DEBRIS WAS DISCOVERED AT THE START OF A SURGICAL CASE, RESULTING IN THE PROCEDURE¿S CANCELLATION. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23493268 ↗MalfunctionNov 5, 2025Product Quality Problem; Product Quality Problem
THE STERILIZATION PROCESS APPEARS TO HAVE ACTIVATED ADHESIVE AND FIBER FRAGMENTS ORIGINATING FROM THE HALYARD ABSORBENT TRAY LINER. ADHESIVE AND FIBERS WERE FOUND ON SURGICAL INSTRUMENTS. NO INJURY WAS REPORTED. TWO PROCEDURES (CORONARY ARTERY BYPASS GRAFT AND SPINE) WERE CANCELED AS RESULT OF DISCOVERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23480402 ↗MalfunctionJul 11, 2025Product Quality Problem; Product Quality Problem
BROWN STAINS WERE FOUND ON TRAY LINER AND INSTRUMENT. IMPLANTS HAD BEEN PLACED ON THE LINER FOR PROCEDURE AT TIME OF DISCOVERY OF THE FOREIGN MATERIAL. PROCEDURE WAS INITIALLY DELAYED AN HOUR AND HALF AS A RESULT, THEN RESCHEDULED. THE BACK TABLE WAS CONSIDERED STERILITY COMPROMISED UPON DISCOVERY OF THE STAINS. IMPLANTS WERE SCRUBBED AND RESTERILIZED FOR THE NEXT CASE. PATIENT WAS UNDER ANESTHESIA AT THE TIME OF DISCOVERY, THERE WERE NO OTHER KNOWN PATIENT AFFECTS PER COMPLAINT REPORTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22478521 ↗InjuryMay 28, 2024Material Discolored; Material Discolored
BROWN SPOTS WERE REPORTED ON THE TRAY LINERS POST CUSTOMER STERILIZATION OF SURGICAL TRAYS. THE SITE REPORTED THERE WAS A DELAY IN SURGERY (VARYING) AND SOME CASES HAD TO BE CANCELLED AS A RESULT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19405169 ↗MalfunctionMay 13, 2024Product Quality Problem; Product Quality Problem
THE TRAY LINER EXHIBITS BROWN STAINS. THE CUSTOMER STATED THIS HAS RESULTED IN DELAYS UP TO (6) HOURS AND MULTIPLE CANCELLATIONS OF CASES. NO INSTRUMENTS FROM THE TRAYS WERE USED ON PATIENTS. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 19297124 ↗MalfunctionMar 13, 2024Flaked; Flaked
THE NURSING TEAM FILED A COMPLAINT REGARDING WHITE PARTICLES FROM THE TRAY LINER FALLING OFF INTO THE INSTRUMENTATION TRAY UPON OPENING. THE PARTICLES WERE REPORTED TO AFFECT THE SURGICAL FIELD STERILITY. ATTEMPTS WERE MADE TO GATHER MORE INFORMATION REGARDING INCIDENT IMPACT ON (B)(6) 2024. THE CUSTOMER HAS NOT RESPONDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18895638 ↗InjuryJun 15, 2023Material Integrity Problem; Material Integrity Problem
THE CUSTOMER NOTICED LIGHT BROWN STAINING ON THE HALYARD 48139 TRAY LINERS. THIS INCIDENT OCCURRED SEVERAL TIMES CONSISTENTLY FOR A FEW DAYS AT THE HOSPITAL AND CAUSED THEM TO STOP STERILIZING TRAYS FOR A PERIOD OF TIME. NO PATIENTS WERE INJURED; HOWEVER, SURGERY DELAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17140411 ↗InjuryJun 6, 2023Material Integrity Problem; Material Integrity Problem
THE CUSTOMER NOTICED LIGHT BROWN STAINING ON THE HALYARD 48139 TRAY LINERS. THIS INCIDENT OCCURRED SEVERAL TIMES CONSISTENTLY FOR A FEW DAYS AT THE HOSPITAL AND CAUSED THEM TO STOP STERILIZING TRAYS FOR A PERIOD OF TIME. NO PATIENTS WERE INJURED; HOWEVER, SURGERY DELAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17076256 ↗
Other Drape, surgical, exempt devices
- Medline DYNJP3102AMedline Industries, Inc. · Drape, surgical, exempt3 recalls2 reports
- Medline DYNJP2498Medline Industries, Inc. · Drape, Surgical, Exempt3 recalls1 report
- Medline DYNJP8005UGMedline Industries, Inc. · Drape, Surgical, Exempt3 recalls0 reports
- Medline DYNJP3103Medline Industries, Inc. · Drape, Surgical, Exempt3 recalls0 reports
- Medline DYNJP8304AMedline Industries, Inc. · Drape, surgical, exempt3 recalls0 reports
- Medline DYNJP9103Medline Industries, Inc. · Drape, surgical, exempt2 recalls7 reports
- Medline DYNJP8201Medline Industries, Inc. · Drape, surgical, exempt2 recalls6 reports
- Medline DYNJP6006Medline Industries, Inc. · Drape, surgical, exempt2 recalls4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page