RAD® 1884012Recalls & reported problems

Medtronic Xomed, Inc. · BLADE 1884012 RAD12 5PK CURVED 4MM · Class 1

All models of RAD (29)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 20681490047535), received between Jan 3, 2018 and Jan 28, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Vibration” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Vibration3
  • Break1
  • Dull, Blunt1
  • Material Fragmentation1
  • Material Separation1

Patient problems

  • No Consequences Or Impact To Patient5
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2025Vibration; Vibration

    IT WAS REPORTED THAT WHEN THE USER INSTALLED THE BLADE INTO THE HANDPIECE FOR TESTING, FOUND THAT THE BLADE HAD OBVIOUSLY SWAY. THE USER THEN TRIED TO RECONNECT THE BLADE BUT USELESS. FIANLLY, CHANGED A NEW BLADE TO PERFORM THE SURGERY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21249391 ↗
  • MalfunctionFeb 13, 2020Vibration; Dull, Blunt; Vibration; Dull, Blunt

    A HEALTH CARE PROVIDER REPORTED VIA MANUFACTURER REPRESENTATIVE THAT THE BLADE WAS CONNECTED TO THE M5 AND IT WASN¿T CUTTING WHEN HE TRIED TO USE IT. WHEN THE HCP TOOK IT OUT FROM THE PATIENT, THE BLADE WAS VIBRATING VIGOROUSLY. HE TRIED TO FIX IT ON AGAIN, BUT IT WAS STILL VIBRATING VIGOROUSLY. ANOTHER PIECE OF THE BLADE WAS OPENED, BUT THE BLADE WASN¿T CUTTING AS WELL. SO, THE HCP CHANGED THE HANDPIECE TO M4, AND IT STILL WASN¿T CUTTING AS WELL. HE DECIDED TO CHANGE TO ANOTHER CONSOLE, AFTER WHICH, THE PROCEDURE PROCEEDED SMOOTHLY. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9706002 ↗
  • MalfunctionOct 4, 2019Vibration; Vibration

    A HEALTH CARE PROVIDER (HCP) REPORTED VIA DISTRIBUTOR THAT A BLADE VIBRATED ABNORMALLY AFTER BEING USED FOR 10 MINUTES DURING A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS). THE DOCTOR REINSTALLED THE BLADE BUT STILL COULD NOT RESOLVE THE PROBLEM. THEN, THE DOCTOR USED ANOTHER NEW BLADE TO COMPLETE THE OPERATION. THE SECOND BLADE COULD FUNCTION PROPERLY WITHOUT VIBRATION WITH THE SAME HANDPIECE. IT WAS THE INITIAL USE OF THE BLADE. NO ERROR CODE WAS DISPLAYED ON THE CONSOLE. THERE WAS A 10 MINUTE DELAY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9156983 ↗
  • MalfunctionNov 20, 2018Material Separation; Material Separation

    A HEALTH CARE PROVIDER (HCP) REPORTED VIA A MANUFACTURER REPRESENTATIVE THAT A BLADE BROKE INTO TWO PIECES WITHIN THE PATIENT'S CAVITY DURING THE PROCEDURE. THE HCP NOTED THAT THE BLADE WAS BEING USED NORMALLY WITH A HANDPIECE. THERE WAS A DELAY IN THE PROCEDURE BECAUSE THE FRAGMENT NEEDED TO BE TAKEN OUT OF THE PATIENT'S CAVITY AND CHECK IF THERE WAS ANY INJURY. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS NO PATIENT IMPACT OR INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8089002 ↗
  • MalfunctionAug 30, 2018Break; Break

    THE CUSTOMER REPORTED THAT THE TIP WAS DAMAGED SHORTLY AFTER HE BEGAN USING IT. THERE WAS NO IMPACT TO THE PATIENT, AND A NEW BLADE WAS OPENED TO CONTINUE THE FESS PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7836115 ↗
  • MalfunctionJan 3, 2018Material Fragmentation; Material Fragmentation

    DURING VISUAL INSPECTION, SMALL FRAGMENTS WERE NOTED INSIDE THE OUTER TUBE MOUTH OF THE BLADE. THE SMALL FRAGMENTS WERE STUCK TO THE TIP WALL. THE FRAGMENTS APPEAR TO BE PORTIONS OF THE INNER CUTTER TIP. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7162140 ↗