Skip to content

Level 1 D-60HLRecalls & reported problems

Icu Medical, Inc. · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20695085406003), received between Aug 16, 2021 and May 19, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak4
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION/LEAKING SET RECENTLY. THE LEAKING OCCURRED WHERE THE SPIKE MET THE TUBING WHERE YOU SPIKE THE BLOOD PRODUCT. THEY WERE NOT ABLE TO SAVE THE TUBING SINCE IT HAD BLOOD PRODUCT WITHIN THE LINE. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE OUTCOME WAS RESOLVED; A NEW TUBING WAS USED. THERE WAS A PATIENT INVOLVEMENT, AND NO PATIENT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22047524 ↗
  • MalfunctionJan 5, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT BLOOD LEAKED OUT OF THE INNER TUBE AND MIXED WITH THE FLUID IN THE OUTER TUBE AND CONTAMINATED THE LEVEL 1 AS SOME BLOOD WAS FOUND IN THE RESERVOIR. HAVE TRIED TROUBLESHOOTING WITH ANOTHER TUBING SET (USING RED FOOD DYE) TO SEE IF THE RED DYE ENDED UP IN THE RESERVOIR. IT DID NOT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18457121 ↗
  • MalfunctionJan 5, 2023Break; Break

    IT WAS REPORTED THAT THE CAREGIVER ATTEMPTED TO LOAD AND UTILIZE TUBING FOR THE DEVICE. SHORTLY AFTER SETUP AND BLOOD CONNECTED, THE CAREGIVER NOTICED THAT THE TUBING HAS SNAPPED JUST ABOVE THE FILTER. THIS INCIDENT DELAYED CARE DUE TO THE FACT THAT ADDITIONAL TUBING WAS NEEDED AND SET UP AGAIN. EVENTUALLY PATIENT RECEIVED TRANSFUSION. THERE WAS NO PATIENT HARM REPORTED, ALTHOUGH THE SIGNIFICANT DELAY IN TRANSFUSION COULD HAVE POTENTIALLY CAUSED ISSUES. PATIENT REQUIRED ADDITIONAL CRYSTALLOID BEFORE BLOOD COULD BE ADMINISTERED. PATIENT WAS STABILIZED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16108927 ↗
  • MalfunctionAug 16, 2021Fluid/Blood Leak; Fluid/Blood Leak

    INFORMATION RECEIVED A SMITHS MEDICAL FLUID WARMING/LEVEL 1 TRAUMA FAST FLOW DISPOSABLES D-60HL WAS LEAKING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12328561 ↗
  • MalfunctionAug 16, 2021Fluid/Blood Leak; Fluid/Blood Leak

    INFORMATION RECEIVED A SMITHS MEDICAL FLUID WARMING/LEVEL 1 TRAUMA FAST FLOW DISPOSABLES D-60HL WAS LEAKING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12328488 ↗

Other DEVICE, WARMING. BLOOD AND PLASMA devices

All DEVICE, WARMING. BLOOD AND PLASMA devices →