Level 1 DI-60HLRecalls & reported problems
Icu Medical, Inc. · Class 2
Quick summary
This ICU Medical Level 1 DI-60HL warmer warms blood and plasma before transfusion. Its 151 adverse event reports include no injuries and 1 death, and many describe fluid leaking from tubing connections, filters and the heat exchanger, and one hole that sprayed fluid. Reports fell to 4 in the past 12 months from 9 before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 150
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 150 adverse event reports that list this exact device (DI 20695085406010), received between Nov 1, 2019 and Apr 8, 2026 — classified as 1 death, 149 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (72 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak72
- Leak/Splash21
- Connection Problem12
- Infusion or Flow Problem7
- Obstruction of Flow6
- Break4
- Defective Component4
- Deformation Due to Compressive Stress4
Patient problems
- No Clinical Signs, Symptoms or Conditions119
- No Consequences Or Impact To Patient13
- Insufficient Information9
- No Patient Involvement7
- No Information2
- Death1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 25 most recent of 150 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Fluid/Blood Leak
IT WAS STATED IN THE REPORT THAT THE CHEMICAL LEAKAGE OCCURRED DURING PRIMING. THE EVENT OCCURRED IN JULY-2024. THE ACTUAL ITEM IS AVAILABLE FOR INVESTIGATION. THIS OCCURRED BEFORE PATIENT USE/DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24817721 ↗MalfunctionFeb 11, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THE LEAKAGE FROM THE FILTER WAS CONFIRMED DURING PATIENT USE. THE EVENT OCCURRED DURING PATIENT USE/ADMINISTRATION AT FACILITY. THERE WAS NO REPORTED PATIENT HARM/ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24316931 ↗MalfunctionNov 12, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE DEVICE EXHIBITED A HOLE NEAR THE LOWER CLAMP THAT CAUSED TRANSFUSION TO SPRAY OUT OF THE HOLE. THERE WAS PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23535770 ↗MalfunctionAug 15, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS STATED THAT THE LIQUID WAS LEAKING, FROM ALUMINUM PART. THE EVENT OCCURRED DURING PATIENT USE/ADMINISTRATION AT FACILITY. THERE WAS NO REPORTED PATIENT HARM/ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22816130 ↗MalfunctionAug 7, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER RETURNED THE DEVICE FOR "LEAKAGE OF WATER FROM THE CIRCUIT," DURING DEVICE ANALYSIS, IT WAS FOUND THAT THE DEVICE WAS LEAKING FROM THE INFUSATE TUBING OUTLET AT THE BOTTOM OF THE HEAT EXCHANGER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22737240 ↗MalfunctionJul 15, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE DEVICE EXHIBITED LEAKAGE FROM THE TUBE CONNECTION. THERE WAS NO PATIENT INVOLVEMENT, NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22511306 ↗MalfunctionJun 20, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER RETURNED THE PRODUCT FOR WATER IS LEAKING FROM THE ALUMINUM CONNECTION, DURING DEVICE EVALUATION, IT WAS DETERMINED THAT A LEAK WAS OBSERVED AT THE INFUSATE TUBING OUTLET AT THE BOTTOM OF THE HEAT EXCHANGER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22302903 ↗MalfunctionJun 9, 2025Fluid/Blood Leak; Device Ingredient or Reagent Problem; Fluid/Blood Leak; Device Ingredient or Reagent Problem
IT WAS STATED THAT LIQUID LEAKAGE WAS OBSERVED DURING THE CIRCULATING. THE EVENT OCCURRED BEFORE PATIENT USE/DURING PRIMING. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22170982 ↗MalfunctionMay 1, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE DEVICE HAD A LEAKAGE FOUND IN THE TUBE. THE EVENT OCCURRED BEFORE OPENING/DURING OPENING/DURING UNPACKING AT FACILITY. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21942438 ↗MalfunctionMar 24, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT LEAKAGE WAS FOUND ON BLUE D-60 MANIFOLD AT THE BEGINNING OF USE. THE PROBLEM WAS RESOLVED BY REPLACING WITH A NEW SET. THE EVENT OCCURRED WHILE IN USE WITH PATIENT AT THE HOSPITAL. THERE WAS NO MEDICAL INTERVENTION REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21679472 ↗MalfunctionDec 9, 2024Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS LEAKAGE FOUND ON THE TUBING SYSTEM AT THE BEGINNING OF USE (LESS THAN 3 MINUTES). THEN THE TUBE DETACHED FROM THE OUTER WHITE TUBE. THIS PROBLEM WAS RESOLVED BY REPLACING WITH A NEW SET. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878893 ↗MalfunctionDec 6, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WATER LEAKS FROM THE CIRCUIT. THE EVENT OCCURRED DURING PRIMING AT THE FACILITY; NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20865561 ↗MalfunctionSep 17, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT LEAKAGE OCCURRED DURING PRIMING. THIS OCCURRED BEFORE PATIENT USE/DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20237584 ↗MalfunctionAug 22, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WITH THE DEVICE THE PART WHERE THE GAS-VENT FILTER AND THE BLOOD TRANSFUSION ROUTE (A TUBING) WERE CONNECTED WAS LEAKING. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20033058 ↗MalfunctionAug 8, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT LIQUID LEAKED FROM THE GRAY TIP. THIS OCCURRED DURING PATIENT USE/ADMINISTRATION. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19939476 ↗MalfunctionJul 17, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WHEN THE DEVICE WAS TRANSFUSING BLOOD TO THE PATIENT, BLOOD WAS LEAKING OUT FROM THE CONNECTION OF THE UPPER TUBE THAT IS COMING OUT OF THE SIDE OF THE FILTER WITH THE GAS VENT IN THE CIRCUIT. THE CIRCUIT WAS CHANGED OUT WITH A NEW ONE AND THEN IT WORKED FINE. THERE WAS PATIENT INVOLVEMENT, BUT NO HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19766863 ↗MalfunctionJul 2, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING PRIMING THE DEVICE WAS LEAKING MEDICAL SOLUTION FROM THE GRAY TIP. NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19656843 ↗MalfunctionJun 27, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BLOOD LEAKED FROM THE CONNECTION OF THE FLOW ROUTE IN THE DIRECTION OF THE AIR FILTER. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19626123 ↗MalfunctionMay 21, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS A CIRCULATING WATER LEAK. THIS OCCURRED BEFORE PATIENT USE/DURING PRIMING AT FACILITY. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19363866 ↗MalfunctionMay 20, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT WHEN THE PRODUCT WAS USED, BLOOD LEAKAGE FROM THE CIRCUIT WAS CONFIRMED. ADVERSE PATIENT EFFECTS ARE UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 19351926 ↗MalfunctionApr 25, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE DEVICE LEAKED FLUID ONTO THE FLOOR. THE TUBING WAS CUT OFF DURING THE PROCESS BECAUSE IT WAS CONSIDERED DANGEROUS. THIS OCCURRED DURING INFUSION. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19181222 ↗MalfunctionApr 11, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE CHEMICAL SOLUTION LEAKED WHEN USING THE PRODUCT. THE EVENT OCCURRED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19087605 ↗MalfunctionApr 2, 2024Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED THAT "WHEN THE PRODUCT WAS USED, IT WAS OBSERVED THAT IT WAS DIFFICULT TO PASS THE INFUSION SOLUTION." THE EVENT OCCURRED IN LATE FEB-2024. THIS OCCURRED DURING PATIENT USE. NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19020851 ↗MalfunctionMar 27, 2024Break; Separation Problem; Break; Separation Problem
IT WAS REPORTED THAT AFTER PRIMING, THE FIRST RBC (BLOOD PRODUCT) WAS ADMINISTERED UNDER PRESSURE, AND WHILE CONNECTING THE SECOND RBC TO THE SPIKE OF THE CIRCUIT, THE OPERATORS HAND SLIPPED AND SEPARATED THE SPIKE NEEDLE WHILE IT WAS STILL CONNECTED TO THE RBC. IT IS THOUGHT THAT THE WEIGHT OF ONE RBC WAS APPLIED TO THE UPPER TIP OF THE DRIP CHAMBER, CAUSING IT TO BREAK AND BREAK OFF. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18987896 ↗MalfunctionJan 5, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT BLOOD FOR A TRANSFUSION LEAKED. THE LEAK APPEARED TO HAVE COME FROM AROUND THE JOINT BETWEEN THE FILTER BODY WITH THE GAS VENT AND THE TUBE. AFTER THAT, A NEW TUBE WAS USED AND THERE WAS NO PROBLEM. THIS OCCURRED DURING PATIENT USE. NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18458246 ↗
Other DEVICE, WARMING. BLOOD AND PLASMA devices
- Level 1 DI-100Icu Medical, Inc. · DEVICE, WARMING. BLOOD AND PLASMA3 deaths131 reports
- Level 1 D-100Icu Medical, Inc. · DEVICE, WARMING. BLOOD AND PLASMA85 reports
- Level 1 DI-300Icu Medical, Inc. · DEVICE, WARMING. BLOOD AND PLASMA44 reports
- Level 1 D-300Icu Medical, Inc. · DEVICE, WARMING. BLOOD AND PLASMA16 reports
- Level 1 D-60HLIcu Medical, Inc. · DEVICE, WARMING. BLOOD AND PLASMA5 reports
- Level 1 Y-INJIcu Medical, Inc. · DEVICE, WARMING. BLOOD AND PLASMA2 reports
- Level 1 PF-1Icu Medical, Inc. · DEVICE, WARMING. BLOOD AND PLASMA1 report