TEMPO AQUA 452414H0Recalls & reported problems
Cordis Corporation · CATH TEMPO AQUA 4FVER135°100CM · Class 2
All models of TEMPO AQUA (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032008843), received between Jan 30, 2019 and Mar 27, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, A 4F 100CM VERTEBRAL (VERT) 135-DEGREE CURVE TEMPO AQUA DIAGNOSTIC CATHETER WAS PREPARED AND CONNECTED TO A POWER INJECTOR. THERE WAS PATENT FLOW CONFIRMED. A CONTRAST POWER INJECTION WAS PERFORMED AT 4ML/SECOND FOR 24ML TOTAL WITH 1 SECOND RATE RISE AT 400 POUNDS PER SQUARE INCH (PSI) IN THE RIGHT INTERNAL CAROTID ARTERY. CONTRAST COULD BE SEEN BRIEFLY DURING THE POWER INJECTION AND THEN DISAPPEARED. UPON INVESTIGATION OF THE DEVICE, IT WAS DISCOVERED THAT THE CATHETER WAS SEPARATED. AN IMAGE OF THE CATHETER WAS PROVIDED WHICH SHOWS THE LUER HUB AND STRAIN RELIEF ARE COMPLETELY SEPARATED FROM THE MAIN BODY OF THE CATHETER. A SMALL PORTION OF THE SEPARATED CATHETER WAS STICKING OUT FROM THE PATIENT AND WAS ABLE TO BE IMMEDIATELY CLAMPED AND REMOVED. THERE WERE NO NOTED COMPLICATIONS UPON CHART REVIEW AND NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND THERE WAS NO DIFFICULTY FLUSHING THE CATHETER PRIOR TO INSERTION. A 0.035" ANGLED NON-CORDIS HYDROPHILIC GUIDEWIRE WAS USED TO DELIVER THE CATHETER PRIOR TO THE ANGIOGRAM. THERE WAS NO RESISTANCE OR FRICTION BETWEEN THE NON-CORDIS GUIDEWIRE AND CATHETER. THERE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21710551 ↗MalfunctionJan 30, 2019Crack; Crack
AS REPORTED, THE 4F TEMP AQUA (4FVER135 DEGREE 100CM) DIAGNOSTIC CATHETER WAS FOUND DAMAGED WHILE REMOVED IT FROM STERILE BAG. THE DAMAGE CONTAIN CRACK OF 1-CENTIMETERS (CM) NEAR THE HUB. THERE WAS NO PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8289440 ↗
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.