TEMPO AQUA 452414H0Recalls & reported problems

Cordis Corporation · CATH TEMPO AQUA 4FVER135°100CM · Class 2

All models of TEMPO AQUA (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032008843), received between Jan 30, 2019 and Mar 27, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, A 4F 100CM VERTEBRAL (VERT) 135-DEGREE CURVE TEMPO AQUA DIAGNOSTIC CATHETER WAS PREPARED AND CONNECTED TO A POWER INJECTOR. THERE WAS PATENT FLOW CONFIRMED. A CONTRAST POWER INJECTION WAS PERFORMED AT 4ML/SECOND FOR 24ML TOTAL WITH 1 SECOND RATE RISE AT 400 POUNDS PER SQUARE INCH (PSI) IN THE RIGHT INTERNAL CAROTID ARTERY. CONTRAST COULD BE SEEN BRIEFLY DURING THE POWER INJECTION AND THEN DISAPPEARED. UPON INVESTIGATION OF THE DEVICE, IT WAS DISCOVERED THAT THE CATHETER WAS SEPARATED. AN IMAGE OF THE CATHETER WAS PROVIDED WHICH SHOWS THE LUER HUB AND STRAIN RELIEF ARE COMPLETELY SEPARATED FROM THE MAIN BODY OF THE CATHETER. A SMALL PORTION OF THE SEPARATED CATHETER WAS STICKING OUT FROM THE PATIENT AND WAS ABLE TO BE IMMEDIATELY CLAMPED AND REMOVED. THERE WERE NO NOTED COMPLICATIONS UPON CHART REVIEW AND NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND THERE WAS NO DIFFICULTY FLUSHING THE CATHETER PRIOR TO INSERTION. A 0.035" ANGLED NON-CORDIS HYDROPHILIC GUIDEWIRE WAS USED TO DELIVER THE CATHETER PRIOR TO THE ANGIOGRAM. THERE WAS NO RESISTANCE OR FRICTION BETWEEN THE NON-CORDIS GUIDEWIRE AND CATHETER. THERE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21710551 ↗
  • MalfunctionJan 30, 2019Crack; Crack

    AS REPORTED, THE 4F TEMP AQUA (4FVER135 DEGREE 100CM) DIAGNOSTIC CATHETER WAS FOUND DAMAGED WHILE REMOVED IT FROM STERILE BAG. THE DAMAGE CONTAIN CRACK OF 1-CENTIMETERS (CM) NEAR THE HUB. THERE WAS NO PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8289440 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →