Skip to content

LIGACLIP LX107Recalls & reported problems

Ethicon Endo-Surgery, LLC · Ligaclip Clip Applier Small 19cm · Class 1

Quick summary

This Ethicon Ligaclip clip applier is a surgical tool used to place small metal clips during surgery. Its 105 reports, 2 of them injuries, often describe clips that did not form properly, jaws that did not line up, or jaws crossing like scissors instead of applying clips. There were no deaths, and no recalls name this device. Reports rose to 14 in the last 12 months from 4 in the previous 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
105
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 105 adverse event reports that list this exact device (DI 20705036012853), received between Mar 9, 2018 and Sep 16, 2026 — classified as 2 injuries, 103 malfunctions.

The most frequently reported specific device problem is “Failure to Form Staple” (55 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 7 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 6 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Form Staple55
  • Failure to Align21
  • Mechanical Problem13
  • Appropriate Device Problem Term/Code Not Available6
  • Unintended Ejection6
  • Detachment of Device or Device Component5
  • Difficult to Open or Close5
  • Defective Device4

Patient problems

  • No Consequences Or Impact To Patient51
  • No Clinical Signs, Symptoms or Conditions44
  • No Patient Involvement6
  • Hemorrhage/Blood Loss/Bleeding3
  • No Known Impact Or Consequence To Patient2
  • Cardiac Arrest1
  • Failure to Anastomose1

Recent adverse event reports

The 25 most recent of 105 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Failure to Form Staple; Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50788300 ↗
  • MalfunctionJul 6, 2026Failure to Form Staple

    IT WAS REPORTED THAT A CLIP APPLIER IS SCISSORING DURING CABG PROCEDURE. NO ADVERSE IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25716871 ↗
  • MalfunctionJun 15, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING A CARDIOVASCULAR PROCEDURE, CLIP APPLIERS ARE SCISSORING DURING CASES. THERE IS NO PATIENT HARM THAT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505608 ↗
  • MalfunctionMay 29, 2026Difficult to Open or Close

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, IT WAS UNCLEAR WHETHER THE ISSUE OCCURRED PRIOR TO OR DURING THE PROCEDURE. THE FIRST DEVICE WAS REPORTED TO HAVE A TIP DEFECT, AND THE SECOND DEVICE WOULD NOT OPEN. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25335509 ↗
  • MalfunctionMay 22, 2026Failure to Form Staple

    IT WAS REPORTED TO THE SALES REP THAT DURING A CABG PROCEDURE, THE SURGEON REPORTED SCISSORING WHEN ATTEMPTING TO APPLY CLIPS. WHEN CLIPPING DIRECTLY ON THE TISSUE. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25256213 ↗
  • MalfunctionMay 22, 2026Failure to Form Staple

    IT WAS REPORTED TO THE SALES REP THAT DURING A CABG PROCEDURE, THE SURGEON REPORTED SCISSORING WHEN ATTEMPTING TO APPLY CLIPS. WHEN CLIPPING DIRECTLY ON THE TISSUE. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25256194 ↗
  • MalfunctionMay 19, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING A CARDIOVASCULAR PROCEDURE, CLIP APPLIERS ARE SCISSORING DURING CASES. THERE IS NO PATIENT HARM THAT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25218353 ↗
  • MalfunctionMay 14, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING A CARDIOVASCULAR PROCEDURE CLIP APPLIERS ARE SCISSORING DURING CASES. THERE IS NO PATIENT HARM THAT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25174889 ↗
  • MalfunctionMay 14, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING A CARDIOVASCULAR PROCEDURE, CLIP APPLIERS ARE SCISSORING DURING CASES. THERE IS NO PATIENT HARM THAT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25174869 ↗
  • MalfunctionFeb 10, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, FUNCTIONAL ISSUES WITH ALL FIVE CLIP APPLIERS IN THE LOT. SPECIFICALLY, THE LAPRA-TY DEVICES DID NOT SEAT PROPERLY ON THE JAWS AND WOULD NOT CLOSE OR LOCK AS INTENDED. IN ADDITION, BOTH THE HEM-O-LOCK BLUE CLIP APPLIERS AND THE WHITE CLIP APPLIERS WERE UNABLE TO HOLD THE CLIPS SECURELY AND WERE NOTED TO BE MISALIGNED. NO PATIENT CONSEQUENCE WERE REPORTED IN ASSOCIATION WITH THESE FAILURES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24305465 ↗
  • MalfunctionFeb 10, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, FUNCTIONAL ISSUES WITH ALL FIVE CLIP APPLIERS IN THE LOT. SPECIFICALLY, THE LAPRA-TY DEVICES DID NOT SEAT PROPERLY ON THE JAWS AND WOULD NOT CLOSE OR LOCK AS INTENDED. IN ADDITION, BOTH THE HEM-O-LOCK BLUE CLIP APPLIERS AND THE WHITE CLIP APPLIERS WERE UNABLE TO HOLD THE CLIPS SECURELY AND WERE NOTED TO BE MISALIGNED. NO PATIENT CONSEQUENCE WERE REPORTED IN ASSOCIATION WITH THESE FAILURES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24305396 ↗
  • MalfunctionJan 14, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THAT ON THREE DEVICES, THE SURGEON EXPERIENCED A SCISSORING ISSUE WITH THE CLIP APPLIER. WHEN THE DEVICE WAS CLOSED, THE RIDGE SEEMS OFF. THERE WERE NO PATIENT ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24068054 ↗
  • MalfunctionJan 14, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THAT ON THREE DEVICES, THE SURGEON EXPERIENCED A SCISSORING ISSUE WITH THE CLIP APPLIER. WHEN THE DEVICE WAS CLOSED, THE RIDGE SEEMS OFF. THERE WERE NO PATIENT ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24068032 ↗
  • MalfunctionJan 14, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THAT ON THREE DEVICES, THE SURGEON EXPERIENCED A SCISSORING ISSUE WITH THE CLIP APPLIER. WHEN THE DEVICE WAS CLOSED, THE RIDGE SEEMS OFF. THERE WERE NO PATIENT ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24067981 ↗
  • MalfunctionJun 10, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING A THYROIDECTOMY SURGERY. AFTER CLAMPED THE CLIPS AND CLIP COULD NOT CLOSE. CHANGED ANOTHER ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED, NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22178980 ↗
  • MalfunctionMar 19, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, DEVICE BECOMES DISTORTED AND DOES NOT FULFILL ITS FUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21647222 ↗
  • MalfunctionMar 4, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIP IS NOT PERFORMING CORRECTLY. THERE WERE NO PATIENT CONSEQUENCE REPORTED. DEVICE IS AVAILABLE FOR RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21518038 ↗
  • InjuryNov 19, 2024Mechanical Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, DEVICE HAS ASYMMETRICAL JAWS AND THE CLIPS ARE NOT FORMED PROPERLY.

    Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20725331 ↗
  • MalfunctionOct 24, 2024No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available; No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIP APPLIER IS CUTTING THE TISSUE WHEN THE JAW IS CLOSED, WHICH IS CAUSING LEAKAGE AND BLEEDING. PROCEDURE DELAYED 20 MINUTES. A NEW CLIP APPLIER WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20529859 ↗
  • MalfunctionJul 17, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS WOULD NOT BECOME SECURE AND WOULD SLIP OFF. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19763638 ↗
  • MalfunctionJul 17, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS WOULD NOT BECOME SECURE AND WOULD SLIP OFF. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19763606 ↗
  • InjuryApr 17, 2024Failure to Form Staple; Adverse Event Without Identified Device or Use Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A CABG PROCEDURE THAT THE CLIPS ARE EITHER SCISSORING OR FALLING OUT OF THE APPLIER BEFORE THEY CAN BE APPLIED. POST OP OF THE PROCEDURE, PATIENT DEVELOPED A BLEED WHICH WAS NOTICED BY THEIR HEMOGLOBIN DROPPING. PATIENT WAS BROUGHT BACK TO THE OR AND CLIPPED THE BLEEDING VESSEL WHICH HAD PREVIOUSLY RECEIVED A SCISSORED CLIP. PATIENT DOING FINE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 19125237 ↗
  • MalfunctionApr 11, 2024Unintended Ejection; Failure to Form Staple; Unintended Ejection; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS ARE NOT CLIPPING PROPERLY AND ARE NOT EFFECTIVE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19091794 ↗
  • MalfunctionMar 20, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS ARE NOT CAPTURING PROPERLY ON THE TISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18941540 ↗
  • MalfunctionMar 7, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING A MYOCARDIAL REVASCULARIZATION WITH BYPASS PROCEDURE, DEVICE DOES NOT COMPLETE THE NECESSARY ADJUSTMENT OF CLIPS. THE CLIP DOES NOT PROPERLY GRASP THE VESSEL. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18857137 ↗

Other Applier, surgical, clip devices

All Applier, surgical, clip devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.