WECK IPN915189Recalls & reported problems

Teleflex Incorporated · Class 1

Quick summary

This Teleflex Weck surgical clip applier is a hand-held tool that places surgical clips. Its 221 reports often describe the applier jaws being misaligned, bent, or breaking, and sometimes parts detaching. There were 3 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
221
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 221 adverse event reports that list this exact device (DI 24026704696813), received between Mar 3, 2021 and Aug 14, 2023 — classified as 3 injuries, 218 malfunctions.

The most frequently reported specific device problem is “Failure to Align” (114 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Failure to Align114
  • Break28
  • Material Twisted/Bent27
  • Detachment of Device or Device Component14
  • Activation, Positioning or Separation Problem12
  • Loose or Intermittent Connection7
  • Material Separation4
  • Activation Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions217
  • Insufficient Information2
  • Foreign Body In Patient1
  • Hemorrhage/Blood Loss/Bleeding1
  • Perforation of Vessels1

Recent adverse event reports

The 25 most recent of 221 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE JAWS OF THE APPLIER WERE MISALIGNED DURING A SURGERY. THEREFORE, ANOTHER APPLIER WAS USED TO COMPLETE THE PROCEDURE. NOTHING FELL/REMAINED IN THE PATIENT." NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17546166 ↗
  • MalfunctionAug 1, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE APPLIER JAWS WERE FOUND MISALIGNED BEFORE USE. THEREFORE, A NEW UNIT WAS USED INSTEAD." NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17434549 ↗
  • MalfunctionJul 28, 2023Detachment of Device or Device Component; Material Twisted/Bent; Detachment of Device or Device Component; Material Twisted/Bent

    IT WAS REPORTED THAT "THE JAWS OF THE APPLIER GOT BENT AND THE PIVOT PIN GOT DETACHED FROM THE APPLIER DURING USE. THEREFORE, THE APPLIER WAS REPLACED WITH A NEW ONE TO COMPLETE THE PROCEDURE. NOTHING FELL/REMAINED IN THE PATIENT AND NO PATIENT INJURY OCCURRED". THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17418056 ↗
  • MalfunctionJul 27, 2023Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE JAWS OF THE APPLIER GOT BENT AND THE PIVOT PIN GOT DETACHED FROM THE APPLIER DURING USE. THEREFORE, THE APPLIER WAS REPLACED WITH A NEW ONE TO COMPLETE THE PROCEDURE. NOTHING FELL/REMAINED IN THE PATIENT AND NO PATIENT INJURY OCCURRED. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17409683 ↗
  • MalfunctionJul 25, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE JAWS OF THE APPLIER WERE FOUND MISALIGNED BEFORE USE. THEREFORE, A NEW UNIT WAS USED INSTEAD". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17389814 ↗
  • MalfunctionJul 19, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT THE JAWS OF THE APPLIER WERE FOUND MISALIGNED BEFORE USE. THEREFORE, A NEW UNIT WAS USED INSTEAD. NO PATIENT IN VOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17355239 ↗
  • MalfunctionJul 17, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE JAWS OF THE APPLIER WERE FOUND MISALIGNED DURING A SURGERY. THEREFORE, THE APPLIER WAS REPLACED WITH A NEW ONE TO COMPLETE THE PROCEDURE. NO CLIP FELL REMAINED IN THE PATIENT." NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17336819 ↗
  • MalfunctionJul 13, 2023Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE JAWS OF THE APPLIER WERE FOUND BENT BEFORE USE. THEREFORE, A NEW UNIT WAS USED INSTEAD. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17321397 ↗
  • MalfunctionJul 13, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT THE JAWS WERE FOUND MISALIGNED WHEN CHECKING IT AT THE HOSPITAL SUPPLY ROOM. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17321345 ↗
  • MalfunctionJul 13, 2023Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT "THE USER WAS UNABLE TO LOAD A CLIP TO THE APPLIER DURING USE. THEREFORE, ANOTHER APPLIER WAS USED TO COMPLETE THE PROCEDURE. NO CLIP FELL/REMAINED IN THE PATIENT. ON CHECKING THE APPLIER, THE JAWS WERE FOUND BENT." NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17318144 ↗
  • MalfunctionJul 10, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE USER COULD NOT LIGATE A CLIP PROPERLY DURING THE PRETEST BEFORE USE. ON CHECKING, THE JAWS OF THE APPLIER WERE FOUND MISALIGNED". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17291690 ↗
  • MalfunctionJul 5, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE JAWS OF THE APPLIER WERE FOUND MISALIGNED BEFORE USE". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17258360 ↗
  • MalfunctionJul 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT "THE SCREW OF THE HANDLE GOT DETACHED DURING USE. THEREFORE, THE APPLIER WAS REPLACED WITH A NEW ONE TO COMPLETE THE SURGERY. NOTHING FELL/REMAINED IN THE PATIENT". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17248744 ↗
  • MalfunctionJun 30, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT THE JAWS OF THE APPLIER WERE FOUND MISALIGNED DURING A SURGERY. THEREFORE, ANOTHER APPLIER WAS USED TO COMPLETE THE PROCEDURE. NO CLIP FELL/REMAINED IN THE PATIENT NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17238475 ↗
  • MalfunctionJun 30, 2023Mechanics Altered; Mechanics Altered

    IT WAS REPORTED THAT THE USER WAS UNABLE TO LIGATE A CLIP DURING A SURGERY, AS THE APPLIER JAW PIN WAS LOOSENED. THEREFORE, THE APPLIER WAS REPLACED WITH A NEW ONE TO COMPLETE THE SURGERY. NO CLIP FELL/REMAINED IN THE PATIENT. NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17238327 ↗
  • MalfunctionJun 29, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT THE APPLIER JAWS WERE FOUND MISALIGNED WHEN IT WAS CHECKED BEFORE USE. THEREFORE, A NEW UNIT WAS USED INSTEAD. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17230834 ↗
  • MalfunctionJun 12, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT PRIOR TO USE, THE APPLIER JAWS WERE FOUND MISALIGNED. AS A RESULT, A NEW APPLIER WAS USED IN THE PROCEDURE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17108745 ↗
  • MalfunctionJun 6, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE APPLIER JAWS WERE FOUND MISALIGNED DURING THE PRETEST BEFORE USE. THEREFORE, A NEW UNIT WAS USED INSTEAD". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17070719 ↗
  • MalfunctionMay 30, 2023Failure to Align; Failure to Align

    IT IS REPORTED THAT PRIOR TO USE, "APPLIER IS MISALIGNED". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17019233 ↗
  • MalfunctionMay 30, 2023Failure to Align; Failure to Align

    IT IS REPORTED THAT PRIOR TO USE, THE "APPLIER IS MISALIGNED". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17019140 ↗
  • MalfunctionMay 26, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT PRIOR TO USE, THE "APPLIER IS MISALIGNED". NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 17014828 ↗
  • MalfunctionMay 26, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT PRIOR TO USE THE "APPLIER IS MISALIGNED". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17011838 ↗
  • MalfunctionMay 26, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT PRIOR TO USE THE "APPLIER IS MISALIGNED". NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17011787 ↗
  • MalfunctionMay 25, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT PRIOR TO USE, THE "APPLIER IS MISALIGNED". NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 17007069 ↗
  • MalfunctionMay 17, 2023Failure to Align; Failure to Align

    IT WAS REPORTED THAT "THE JAWS OF THE APPLIER WERE FOUND MISALIGNED DURING A SURGERY. THEREFORE, ANOTHER APPLIER WAS USED TO COMPLETE THE PROCEDURE. NO CLIP FELL/REMAINED IN THE PATIENT". THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16947641 ↗

Other Applier, surgical, clip devices

All Applier, surgical, clip devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.