Skip to content

N/A 1882569HS (UDI 20721902978009)Recalls & reported problems

Medtronic Xomed, Inc. · BUR 1882569HS 3PK CVD DIAMOND DCR 2.5MM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20721902978009), received between Mar 31, 2017 and Feb 21, 2020 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Overheating of Device2
  • Use of Device Problem2
  • Break1
  • Material Fragmentation1
  • Material Separation1
  • Unintended Movement1

Patient problems

  • No Consequences Or Impact To Patient4
  • Radiation Exposure, Unintended1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 21, 2020Material Separation; Material Separation

    A HEALTHCARE PROVIDER REPORTED VIA MANUFACTURER REPRESENTATIVE THAT THE BUR HEAD BROKE OFF INSIDE THE PATIENT. THE HCP USED A DCR BUR AT 30,000 RPM INSTEAD OF THE INDICATED 12,000 RPM. THE BUR HEAD WAS REMOVED, AN X-RAY WAS PERFORMED TO CHECK THAT NO METAL REMAINED IN THE PATIENT. THE PROCEDURE WAS DELAYED FOR 60 MINUTES. THE BUR WAS STUCK INSIDE THE HANDPIECE. A DIFFERENT HANDPIECE WAS USED TO FINISH THE PROCEDURE. THE PATIENT WAS NOT INJURED.

    Patient problems: Radiation Exposure, Unintended

    MAUDE report 9738993 ↗
  • MalfunctionAug 15, 2018Break; Break

    THE BUR BROKE AFTER 20 MINUTES OF USING. THERE WAS NO PATIENT IMPACT OR INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7784038 ↗
  • MalfunctionJul 11, 2018Overheating of Device; Use of Device Problem; Unintended Movement; Use of Device Problem

    DURING USE, THE BUR OVER-HEATED AND THE SHAFT WOBBLED. THIS IS A BRAND NEW BUR. THERE WAS NO PATIENT IMPACT OR INJURY. THE PROCEDURE WAS COMPLETED WITH BACK-UP PRODUCT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7678912 ↗
  • MalfunctionJul 11, 2018Overheating of Device; Use of Device Problem; Use of Device Problem

    DURING USE, THE BUR OVER-HEATED AND THE SHAFT WOBBLED. THIS IA A BRAND NEW BUR. THERE WAS NO PATIENT IMPACT OR INJURY. THE PROCEDURE WAS COMPLETED WITH BACK-UP PRODUCT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7678876 ↗
  • MalfunctionMar 31, 2017Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT A HIGH SPEED BUR BROKE OFF 2CM FROM THE TIP AND INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED WITHOUT COMPLICATIONS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6450382 ↗

Other BUR, EAR, NOSE AND THROAT devices

All BUR, EAR, NOSE AND THROAT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page