N/A 1882569HS (UDI 20721902978009)Recalls & reported problems
Medtronic Xomed, Inc. · BUR 1882569HS 3PK CVD DIAMOND DCR 2.5MM · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20721902978009), received between Mar 31, 2017 and Feb 21, 2020 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Overheating of Device2
- Use of Device Problem2
- Break1
- Material Fragmentation1
- Material Separation1
- Unintended Movement1
Patient problems
- No Consequences Or Impact To Patient4
- Radiation Exposure, Unintended1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryFeb 21, 2020Material Separation; Material Separation
A HEALTHCARE PROVIDER REPORTED VIA MANUFACTURER REPRESENTATIVE THAT THE BUR HEAD BROKE OFF INSIDE THE PATIENT. THE HCP USED A DCR BUR AT 30,000 RPM INSTEAD OF THE INDICATED 12,000 RPM. THE BUR HEAD WAS REMOVED, AN X-RAY WAS PERFORMED TO CHECK THAT NO METAL REMAINED IN THE PATIENT. THE PROCEDURE WAS DELAYED FOR 60 MINUTES. THE BUR WAS STUCK INSIDE THE HANDPIECE. A DIFFERENT HANDPIECE WAS USED TO FINISH THE PROCEDURE. THE PATIENT WAS NOT INJURED.
Patient problems: Radiation Exposure, Unintended
MAUDE report 9738993 ↗MalfunctionAug 15, 2018Break; Break
THE BUR BROKE AFTER 20 MINUTES OF USING. THERE WAS NO PATIENT IMPACT OR INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7784038 ↗MalfunctionJul 11, 2018Overheating of Device; Use of Device Problem; Unintended Movement; Use of Device Problem
DURING USE, THE BUR OVER-HEATED AND THE SHAFT WOBBLED. THIS IS A BRAND NEW BUR. THERE WAS NO PATIENT IMPACT OR INJURY. THE PROCEDURE WAS COMPLETED WITH BACK-UP PRODUCT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7678912 ↗MalfunctionJul 11, 2018Overheating of Device; Use of Device Problem; Use of Device Problem
DURING USE, THE BUR OVER-HEATED AND THE SHAFT WOBBLED. THIS IA A BRAND NEW BUR. THERE WAS NO PATIENT IMPACT OR INJURY. THE PROCEDURE WAS COMPLETED WITH BACK-UP PRODUCT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7678876 ↗MalfunctionMar 31, 2017Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT A HIGH SPEED BUR BROKE OFF 2CM FROM THE TIP AND INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED WITHOUT COMPLICATIONS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6450382 ↗
Other BUR, EAR, NOSE AND THROAT devices
- Quadcut® 1884380HRMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall7 reports
- Silver Bullet® 1882905HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall3 reports
- Diego Elite MBUR6060FRCVGyrus ACMI, LLC · BUR, EAR, NOSE AND THROAT1 recall2 reports
- Quadcut® 1884380HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall1 report
- RAD® 1882916Medtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall1 report
- N/A 1884032Medtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall0 reports
- DIEGO ELITE MBUR5060FRCVGyrus ACMI, LLC · BUR, EAR, NOSE AND THROAT1 recall0 reports
- Tricut® 1884033HREMedtronic Xomed, Inc. · BUR, EAR, NOSE AND THROAT1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page