Bausch + Lomb SE5523WVRecalls & reported problems
Bausch & Lomb Incorporated · 23 ga Combined Wide-Field Elite Pack · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20757770059694), received between Apr 10, 2023 and Jul 10, 2024 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Dull, Blunt” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Aug 8, 2018Terminated
The back cap separates from the body of the vitrectomy cutter.
Z-3275-2018 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Dull, Blunt2
- Leak/Splash1
- Particulates1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Retinal Detachment2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 10, 2024Particulates; Particulates
IT WAS REPORTED THAT THE SURGEON REMOVED A PARTICLE FROM THE EYE DURING A PROCEDURE. THE PARTICLE WAS NOTICED COMING DOWN THE IRRIGATION SIDE AND OUT THE IRRIGATION HOLE ON THE LEFT WHILE USING A QUAD TECHNIQUE FOR PHACOEMULSIFICATION. THE SURGEON HALTED THE PROCEDURE, REMOVED THE PHACOEMULSIFICATION TOOL, AND EXTRACTED THE PARTICLE WITH FORCEPS. NO HARM TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19715578 ↗InjuryApr 11, 2024Leak/Splash; Leak/Splash
THE USER FACILITY IN THE UNITED KINGDOM REPORTED THAT DURING RETINAL DETACHMENT SURGERY, THREE TROCARS WERE LEAKING. THE PRESSURE WAS SET AT 30MMHG. SUTURES WERE REQUIRED AS THE WOUNDS DID NOT SEAL PROPERLY AT THE TROCAR ENTRY SIGHT. THE PROCEDURE WAS EXTENDED BY 15-20 MINUTES. THERE WAS NO FURTHER MEDICAL INTERVENTION NOR IMPACT OR CLINICAL CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19093113 ↗InjuryApr 10, 2023Dull, Blunt; Dull, Blunt
THE USER FACILITY IN FRANCE REPORTED THE INSTRUMENTS WERE NOT SHARP ENOUGH TO INSERT COMPLETELY. REPORTEDLY, THIS LED TO CHOROIDAL DETACHMENT INTRAOPERATIVELY. ANOTHER SATELLITE KIT (WITH SCLEROTOME AND TROCAR) WAS USED. THE PROCEDURE WAS DELAYED BY ONE HOUR AND THE PATIENT REQUIRED SEVERAL ADDITIONAL SCLEROTOMIES. THIS REPORT IS FOR THE SECOND PATIENT.
Patient problems: Retinal Detachment
MAUDE report 16707510 ↗InjuryApr 10, 2023Dull, Blunt; Dull, Blunt
THE USER FACILITY IN FRANCE REPORTED THE INSTRUMENTS WERE NOT SHARP ENOUGH TO INSERT COMPLETELY. REPORTEDLY, THIS LED TO CHOROIDAL DETACHMENT INTRAOPERATIVELY. ANOTHER SATELLITE KIT (WITH SCLEROTOME AND TROCAR) WAS USED. THE PROCEDURE WAS DELAYED BY ONE HOUR AND THE PATIENT REQUIRED SEVERAL ADDITIONAL SCLEROTOMIES. THIS REPORT IS FOR THE FIRST PATIENT.
Patient problems: Retinal Detachment
MAUDE report 16707449 ↗
Other Instrument, Vitreous Aspiration And Cutting, Ac-Powered devices
- CONSTELLATION 8065751150Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered1 recall1,246 reports
- CONSTELLATION 8065753041Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered503 reports
- Centurion, OZil 8065751761Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered475 reports
- UNITY 00380650002969Alcon Laboratories, Inc. · Instrument, Vitreous Aspiration And Cutting, Ac-Powered271 reports
- Ozil 8065750469Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered261 reports
- Infiniti 8065752160Alcon Laboratories, Inc. · Instrument, vitreous aspiration and cutting, ac-powered54 reports
- Bausch + Lomb SE5525WVBausch & Lomb Incorporated · Instrument, Vitreous Aspiration And Cutting, Ac-Powered1 recall2 reports
- ALCON 00380650003935Alcon Laboratories, Inc. · Instrument, Vitreous Aspiration And Cutting, Ac-Powered1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page