Skip to content

DLP™ 77010 (UDI 20763000091112)Recalls & reported problems

Medtronic, Inc. · CANNULA 77010 ART ONE PIECE 10FR 5PK 17 · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 20763000091112), received between Aug 31, 2023 and Jun 2, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Misassembled During Installation” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K024069); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 19, 2024Open, Classified

    Potential for unsealed sterile packing.

    Z-1793-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Misassembled During Installation5
  • Contamination3
  • Crack2
  • Material Separation2
  • Mechanical Problem2
  • Material Puncture/Hole1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 2, 2026Material Separation

    IT WAS REPORTED THAT PRIOR TO USE OF THE DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULAE, DUST AND PARTICLES WERE FOUND INSIDE THE STERILE PACKAGE AND ONE OF THE CANNULAE HAD LOOSE CAPS. THE DEVICES WERE NOT USED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25360877 ↗
  • MalfunctionMay 23, 2025Contamination; Misassembled During Installation; Contamination; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULAE, IT WAS REPORTED THAT THERE WAS DIRT IN THE POUCHES. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE STERILE BARRIERS WERE STILL SEALED WHEN THE FOREIGN MATERIAL (FM) WAS IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22077230 ↗
  • MalfunctionMay 23, 2025Contamination; Misassembled During Installation; Contamination; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULAE, IT WAS REPORTED THAT THERE WAS DIRT IN THE POUCHES. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE STERILE BARRIERS WERE STILL SEALED WHEN THE FOREIGN MATERIAL (FM) WAS IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22077192 ↗
  • MalfunctionMay 23, 2025Contamination; Misassembled During Installation; Contamination; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF THESE DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULAE, IT WAS REPORTED THAT THERE WAS DIRT IN THE POUCHES. THE DEVICES WERE REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE STERILE BARRIERS WERE STILL SEALED WHEN THE FOREIGN MATERIAL (FM) WAS IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22077146 ↗
  • MalfunctionOct 8, 2024Mechanical Problem; Misassembled During Installation; Mechanical Problem; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA, IT WAS REPORTED THAT THE CUSTOMER FOUND THAT THE TUBE CORE AND THE CANNULA BODY DID NOT MATCH AFTER UNPACKING. THE TUBE CORE WAS TOO SHORT FOR THE TIP TO COME OUT. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20395362 ↗
  • MalfunctionAug 28, 2024Mechanical Problem; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA, IT WAS REPORTED THAT THE TIP OF THE CANNULA COULD NOT BE PULLED. THE DEVICE WAS REPLACED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TIP OF THE CANNULA COULD NOT BE PULLED OUT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20083385 ↗
  • MalfunctionAug 23, 2024Obstruction of Flow; Misassembled During Installation; Obstruction of Flow; Misassembled During Installation

    MEDTRONIC RECEIVED INFORMATION THAT AT AN UNSPECIFIED TIME, THIS DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA HAD A CAP OF THE INTRODUCER THAT WAS FITTED BACKWARDS; HENCE, IT DOES NOT OCCLUDE THE CANNULA. NO PATIENT WAS HURT, AS THE DEFECT WAS NOTICED TIMELY. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20046007 ↗
  • MalfunctionJun 18, 2024Material Separation; Material Separation

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA, IT WAS REPORTED THAT THE CUSTOMER OBSERVED THAT THE DEVICE ACCESSORIES HAD FALLEN OFF WHEN THE HOSPITAL WAS CHECKING THE DEVICE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE DEVICES WERE DAMAGED, THE INNER CORE FRACTURES HAD FALLEN OFF ON EACH DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19557163 ↗
  • MalfunctionNov 28, 2023Crack; Crack

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THIS DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA IT WAS NOTED THAT THE CANNULA HAD A TEAR. THE TEAR WAS NOTICED WHEN THE LIVER FROM THE PUMP HAD BEEN REMOVED. THE DEVICE WAS REPLACED. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18215145 ↗
  • MalfunctionAug 31, 2023Crack; Material Puncture/Hole; Crack; Material Puncture/Hole

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA, THE CUSTOMER REPORTED A CUT IN THE FRONT OF THE CANNULA WHICH WAS OBSERVED AFTER STARTING BYPASS. SEE ATTACHED PHOTOS. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.  THE CANNULA WAS POSSIBLY TIED TOO TIGHT BY THE SURGEON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17660934 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K024069. One submission often covers many models, so check the product and lot details in each record.

  • Z-2738-2026 · May 25, 2026 · Open, Classified
    Medtronic DLP One-Piece Pediatric Arterial Cannula, Product Number/CFN 77010; Cardiopulmonary bypass vascular cannula
  • Z-0984-2025 · Dec 12, 2024 · Open, Classified
    DLP Pediatric One-Piece Arterial Cannulae, Model Numbers 77008 (8 Fr) and 77014 (14 Fr)

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices

All Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.