Skip to content

DLP™ 77014Recalls & reported problems

Medtronic, Inc. · CANNULA 77014 ART ONE PIECE 14FR 5PK 17L · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20763000091136), received between Sep 25, 2024 and Sep 2, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K024069); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 12, 2024Open, Classified

    Incorrect labeling for seven manufactured lots of certain models of Arterial Cannulae.

    Z-0984-2025 · Root cause: Packaging · FDA record ↗

Most reported problems

Device problems

  • Device Markings/Labelling Problem1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT UPON OPENING THE PACKAGING OF THE DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA, THE PLASTIC MANDREL DID NOT REACH THE TIP OF THE CANNULA, WHICH MADE CANNULATION OF THE AORTA DIFFICULT. THE USE OF THE DEVICE WAS UNSPECIFIED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22931776 ↗
  • MalfunctionSep 25, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    MEDTRONIC RECEIVED INFORMATION THAT AT AN UNSPECIFIED TIME, THIS DLP PEDIATRIC ONE-PIECE ARTERIAL CANNULA WAS FOUND TO BE MARKED AS A 14FR AND WAS INDEED IN A BOX LABELLED 14FR, HOWEVER THE ACTUAL SIZE OF THE DEVICE WAS 12FR. THE MANDREL WAS INDEED 14FR. USE OF DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20299651 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K024069. One submission often covers many models, so check the product and lot details in each record.

  • Z-2738-2026 · May 25, 2026 · Open, Classified
    Medtronic DLP One-Piece Pediatric Arterial Cannula, Product Number/CFN 77010; Cardiopulmonary bypass vascular cannula

Other Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices

All Catheter, cannula and tubing, vascular, cardiopulmonary bypass devices →