CD Horizon™ Astute™ Spinal System 5630724Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SET SCREW 5630724 BRKOFF 6.0PEEK STERBX4 · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20763000313344), received between Aug 18, 2021 and Sep 13, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Material Deformation1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 13, 2021Material Deformation; Scratched Material; Material Deformation; Scratched Material
INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA FIELD REPRESENTATIVE REGARDING PATIENT WITH SYMPTOMS OF CSR INVOLVED IN PSF (POSTERIOR SPINAL FUSION) PROCEDURE. LEVELS IMPLANTED: C2-T2. IT WAS REPORTED THAT RIGHT TAPERED ROD WAS BROKEN. NO BROKEN FRAGMENTS LEFT IN THE PATIENT'S BODY. THERE WAS NO DELAY IN OVERALL PROCEDURE TIME. PATIENT WAS NOT HOSPITALIZED PROLONG AS A RESULT OF THIS EVENT. REPORTED THEY MIGHT HAVE REMOVED THE NUT AND ROD FOR PERFORMING DECOMPRESSION ADDITIONALLY, AND THE RIGHT ROD WAS BROKEN WHEN PREPARING. PRODUCT WAS USED CORRECTLY ACCORDING TO THE DIRECTIONS GIVEN IN THE IFU/LABELING. REPORTED EVENT WAS REVISION SURGERY FOR CSR. INITIAL SURGERY WAS PERFORMED WITH NON-MDT PRODUCT. PROCEDURE INVOLVED IN INITIAL SURGERY WAS C3/4: PSF FOR PLASTY HIRAYAMA DISEASE. ADDITIONAL TREATMENT: FORAMINOTOMY C5 /6 /7 LEFT WAS PERFORMED. ADDITIONAL COMMENTS ABOUT THE EVENT: (B)(6) 2007 HIRAYAMA DISEASE. C3/4: PSF PERFORMED WITH NON-MDT PRODUCT (B)(6) 2021. CSR C2-T2: PSF NON-MDT PRODUCT REMOVAL COULD NOT BE PERFORMED. THE RIGHT HEAD CAME OFF ON ONE SIDE, AND THERE WAS NO CHOICE BUT TO LEAVE PRODUCT ON THE LEFT AND HYBRID WITH VS. ON (B)(6) 2021, RECEIVED ADDITIONAL
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12460843 ↗MalfunctionAug 18, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROFESSIONAL (HCP) VIA THE MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED AFTER THE INITIAL L1/2/3 PSF PROCEDURE DONE ON (B)(6) 2018 IN A PATIENT DIAGNOSED WITH SPINAL CORD TUMOR. IT WAS REPORTED THAT DURING A REGULAR INSPECTION ON 26 JULY 2021 IT WAS CONFIRMED THAT THE SET SCREW ON THE LEFT SIDE OF L1 BACKED HAD BACKED OUT. A REVISION SURGERY IS SCHEDULED FOR (B)(6) 2021 TO EXPLANT THE PRODUCT. NO PATIENT INJURY / COMPLICATION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12335237 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page