SURGAIRTOME TWO 00505800100Recalls & reported problems
Conmed Corporation · SURGAIRTOME TWO HANDPIECE · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20845854018924), received between Mar 4, 2022 and Dec 18, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K781979); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Overheating of Device1
Patient problems
- Insufficient Information1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 18, 2023Overheating of Device; Overheating of Device
A SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, 00505800100, SURGAIRTOME TWO HANDPIECE, WAS BEING USED DURING AN UNNAMED PROCEDURE ON (B)(6) 2023, AND THE ¿OVERHEATED SURGAIRTOME RESULTED IN A BURN TO THE PATIENTS INNER LIP¿. THE CUSTOMER CONFIRMED ¿IT WAS A FIRST DEGREE BURN. THEY TREATED IT WITH OINTMENT, NO SURGICAL INTERVENTION. THE PATIENT IS CURRENTLY ¿OK¿. THE SURGERY WAS COMPLETED SUCCESSFULLY.¿. THE 00137501200 WILL BE LISTED AS A CONCOMITANT DEVICE AND THERE IS NO ALLEGATION AGAINST IT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: Superficial (First Degree) Burn
MAUDE report 18344277 ↗MalfunctionMar 4, 2022Break
A BURR SNAPPED IN HALF AT THE STERILE FIELD DURING SURGERY DUE TO A BROKEN LOCKING MECHANISM ON A SURGAIRTOME.
Patient problems: Insufficient Information
MAUDE report 13666307 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K781979. One submission often covers many models, so check the product and lot details in each record.
- Z-0711-2008 · Aug 9, 2007 · TerminatedConMed Linvatec, Ultrapower Burs, Linvatec Corporation, Largo, FL 33773
- Z-0041-05 · Aug 26, 2004 · TerminatedUltra Power Burs, Sterile, Single-Use.
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, PNEUMATIC POWERED & ACCESSORY/ATTACHMENT devices
- NA 310.221Synthes GmbH · Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment123 reports
- NA 310.23Synthes GmbH · Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment50 reports
- NA 310.15SSynthes GmbH · Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment1 recall0 reports
- NA 310.19Synthes GmbH · Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment21 reports
- NA 310.15Synthes GmbH · Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment18 reports
- NA 310.31Synthes GmbH · Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment15 reports
- NA 310.16Synthes GmbH · Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment9 reports
- NA 310.634Synthes GmbH · Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.