Skip to content

HIP PRESERVATION SYSTEM SIGNATURE SERIES HPS-HB11Recalls & reported problems

Conmed Corporation · HPS PREBENT SPHERICAL BUR, 5.5 MM X 19 CM · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20845854080518), received between Oct 7, 2019 and Jan 19, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 3 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation5

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2024Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE HPS-HB11, 5.5MM HPS PREBENT SPHERICAL BUR WAS BEING USED ON (B)(6) 2023 DURING A HIP ARTHROSCOPY LABRAL REPAIR AND FAI DEBRIDEMENT AND "DURING HIP ARTHROSCOPY 2 X HPS-HB11 FELL APART DURING OPERATION - HEAD FELL OFF AND THE BLACK PLASTIC SHEATH ALSO CAME OFF. THE ITEMS THEN NEEDED TO BE FISHED OUT OF THE HIP. MANUAL RETRIEVAL OF BUR AND SHEATH FROM JOINT¿. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE SAME DEVICE OF A DIFFERENT LOT NUMBER. THERE WAS A 30 MINUTE DELAY TO THE PROCEDURE. THERE WAS NO INJURY OR IMPACT TO THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18545901 ↗
  • MalfunctionDec 20, 2023Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE HPS-HB11, 5.5MM HPS PREBENT SPHERICAL BUR, WAS BEING USED DURING AN HIP ARTHROSCOPY PROCEDURE ON (B)(6) 2023 WHEN IT WAS REPORTED, ¿THE BURR HEAD BROKE OFF IN THE PATIENT'S JOINT, AND THE DOCTOR WAS ABLE TO RETRIEVE IT." THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO REPORTED DELAY. A SECOND BUR, HPS-HB11, 5.5MM HPS PREBENT SPHERICAL BUR, WAS USED AS A CONCOMITANT DEVICE; HOWEVER, THE SURGEON FELT THE BLADE WAS NOT WORKING PROPERLY AND WAS CONCERNED THAT THE BURR HEAD WOULD COME APART AGAIN. THERE WAS NO REPORT OF THE ALTERNATE DEVICE HAVING ANY BREAKAGE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18362910 ↗
  • MalfunctionDec 6, 2023Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, HPS-HB11, 5.5MM HPS PREBENT SPHERICAL BUR, WAS BEING USED DURING A HIP ARTHROSCOPY, CAM RESECTION ON (B)(6) 2023, AND ¿WE HAD 2 HPS-HB11 BURS MALFUNCTION IN THE SAME CASE THIS MORNING. THE FIRST BUR WAS ¿SKIPPING OR STALLING¿ WHILE THE FOOT PEDAL WAS COMPLETELY PRESSED. WE SWITCHED THE BUR OUT WITH ANOTHER ONE OF THE SAME LOT NUMBER AND IT BEGAN DOING THE SAME THING. THE SURGEON KEPT USING IT AND THE HEAD OF THE BUR BROKE OFF INSIDE THE HIP ON THE CAM RESECTION. HE WAS ABLE TO REMOVE IT QUICKLY WITH NO SIGNIFICANT DELAY.¿ THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE SAME DEVICE. FURTHER ASSESSMENT CONFIRMED THAT THERE WAS NO INJURY, MEDICAL OR SURGICAL INTERVENTION FOR THE PATIENT OR USER, AND NO EXTENDED HOSPITAL STAY WAS REQUIRED FOR THE PATIENT. THE PATIENT IS REPORTEDLY DOING WELL. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18277197 ↗
  • MalfunctionDec 5, 2023Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE HPS-HB11, 5.5MM HPS PREBENT SPHERICAL BUR WAS BEING USED ON (B)(6) 2023 AND "WE HAD AN ISSUE TODAY WITH A NEW CONMED SURGEON. WE HAD 2 OF THE PRODUCTS ABOVE BREAK IN ONE CASE BACK-TO-BACK. WAS USING THE BURR IN ACCORDANCE TO HOW IT IS INTENDED TO BE USED AND HAD TO OPEN UP OTHER COMPANIES BURRS INSTEAD OF OURS TO FINISH THE CASE. BOTH WERE THE SAME LOT NUMBER¿. PER FURTHER ASSESSMENT IT WAS FOUND THAT "THE DEVICE DID BREAK INTO THE SURGICAL FIELD AND ALL COMPONENTS WERE REMOVED FROM THE FIELD." THERE WAS NO REPORT OF IMPACT/INJURY, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION TO THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18265674 ↗
  • MalfunctionOct 7, 2019Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT HPS-HB11 DEVICE WAS BEING USED DURING A HIP SURGERY ON (B)(6) 2019. AFTER 3-5 SECONDS THERE WAS A NOISE AND THEN THE DEVICE WAS STUCK. IT WAS REPORTER DURING ASSESSMENT QUESTIONING THAT "THERE WAS SOME PLASTIC ABRASION IN THE JOINT" THAT WAS REMOVED BY THE SURGEON. THERE WAS NO INJURY TO THE PATIENT OR USER. THERE WAS NO REPORT OF MEDICAL INTERVENTION OR HOSPITALIZATION. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9161238 ↗

FDA clearance and classification

The FDA regulates product code GFF (bur, surgical, general & plastic surgery) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HIP PRESERVATION SYSTEM SIGNATURE SERIES HPS-HB11

Has the HIP PRESERVATION SYSTEM SIGNATURE SERIES HPS-HB11 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20845854080518), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HIP PRESERVATION SYSTEM SIGNATURE SERIES HPS-HB11?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Oct 7, 2019 and Jan 19, 2024. In those reports, the most common specific device problem is “Material Fragmentation” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HIP PRESERVATION SYSTEM SIGNATURE SERIES HPS-HB11?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HIP PRESERVATION SYSTEM SIGNATURE SERIES HPS-HB11 FDA approved?

Devices of this type (product code GFF) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Conmed Corporation.

Who makes the HIP PRESERVATION SYSTEM SIGNATURE SERIES HPS-HB11, and is it still sold?

The FDA's device database lists Conmed Corporation as the company responsible for it (brand HIP PRESERVATION SYSTEM SIGNATURE SERIES, model HPS-HB11). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other BUR, SURGICAL, GENERAL & PLASTIC SURGERY devices

All BUR, SURGICAL, GENERAL & PLASTIC SURGERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page