Skip to content

Medline MDSM611950Recalls & reported problems

Medline Industries, Inc. · ELECTRODE,ECG,STRESS,FOAM,50PK · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20884389830465), received between Jun 12, 2024 and Sep 26, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Adhesive Too Strong” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adhesive Too Strong1
  • Appropriate Device Problem Term/Code Not Available1
  • High Readings1
  • Loss of or Failure to Bond1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Rash1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 26, 2026Adhesive Too Strong

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906334 ↗
  • MalfunctionSep 21, 2026Loss of or Failure to Bond

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50845181 ↗
  • MalfunctionSep 14, 2026High Readings

    It was reported that the issue was "inaccurate readings and spikes."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50761421 ↗
  • InjuryJun 12, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE FACILITY ON 05/14/24 THE PATIENT EXPERIENCED AN "INTENSELY PRURITIC VESICULAR RASH WHERE MEDLINE ECG ELECTRODE WERE PLACED".

    Patient problems: Rash

    MAUDE report 19519440 ↗

FDA clearance and classification

The FDA regulates product code DRX (electrode, electrocardiograph) under 21 CFR 870.2360 as a Class II device. 21 CFR 870.2360 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Medline MDSM611950

Has the Medline MDSM611950 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20884389830465), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Medline MDSM611950?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jun 12, 2024 and Sep 26, 2026. In those reports, the most common specific device problems are “Adhesive Too Strong” (1), “High Readings” (1) and “Loss of or Failure to Bond” (1); the most common patient problem is “Rash” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Medline MDSM611950?

None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Medline MDSM611950 FDA approved?

Devices of this type (product code DRX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Medline Industries, Inc.

Who makes the Medline MDSM611950, and is it still sold?

The FDA's device database lists Medline Industries, Inc. as the company responsible for it (brand Medline, model MDSM611950). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Electrode, electrocardiograph devices

All Electrode, electrocardiograph devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page