Skip to content

Shiley 4.5PEF (UDI 20884521101583)Recalls & reported problems

Covidien LP · Pediatric Tracheostomy Tube Cuffless · Class 2

All models of Shiley (387)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 20884521101583), received between Aug 28, 2020 and Oct 8, 2024 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Human-Device Interface Problem” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K122531, K182861); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Human-Device Interface Problem4
  • Appropriate Device Problem Term/Code Not Available1
  • Patient Device Interaction Problem1
  • Use of Device Problem1

Patient problems

  • Unintended Extubation6
  • Bacterial Infection3
  • Extubate1
  • No Clinical Signs, Symptoms or Conditions1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 8, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, AFTER 3 DAYS OF PATIENT USE, THE TRACHEOSTOMY TUBE'S NECK FLANGE FIXING HOLE WAS FOUND BROKEN AS SHOWN IN THE ATTACHED IMAGE. THE TUBE WAS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation

    MAUDE report 20395564 ↗
  • MalfunctionJan 16, 2024Human-Device Interface Problem; Appropriate Device Problem Term/Code Not Available; Human-Device Interface Problem; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE REPORTER, THE PATIENT MISTAKENLY KICKED AND APPLIED TENSION TO THE TRACHEOSTOMY TUBE WITH THEIR OWN FOOT, AND SELF- EXTUBATION MISTAKENLY OCCURRED TWO TO THREE DAYS AFTER INTUBATION. WHEN THAT TUBE WAS CHECKED, IT WAS CONFIRMED THAT THERE WAS A CRACK IN ONE OF THE PINHOLES OF THE NECK FLANGE. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 18518518 ↗
  • InjuryApr 20, 2022Human-Device Interface Problem; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, THE CLEAR FLANGE OF THE DEVICE WAS COLLECTED OF A MILKY FILM COATING AND THEN WOULD FLAKE OFF. THE DEVICE HAD A LOT MORE MUCOUS PLUGGING THAN NORMAL. THE PATIENT HAD A REPLACEMENT OF THE TUBE.

    Patient problems: Bacterial Infection; Unintended Extubation

    MAUDE report 14160418 ↗
  • InjuryApr 20, 2022Human-Device Interface Problem; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, THE CLEAR FLANGE OF THE DEVICE WAS COLLECTED OF A MILKY FILM COATING AND THEN WOULD FLAKE OFF. THE DEVICE HAD A LOT MORE MUCOUS PLUGGING THAN NORMAL. THE PATIENT HAD A REPLACEMENT OF THE TUBE.

    Patient problems: Bacterial Infection; Unintended Extubation

    MAUDE report 14160342 ↗
  • InjuryApr 20, 2022Human-Device Interface Problem; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, THE CLEAR FLANGE OF THE DEVICE WAS COLLECTED OF A MILKY FILM COATING AND THEN WOULD FLAKE OFF. THE DEVICE HAD A LOT MORE MUCOUS PLUGGING THAN NORMAL. THE PATIENT HAD A REPLACEMENT OF THE TUBE.

    Patient problems: Bacterial Infection; Unintended Extubation

    MAUDE report 14160275 ↗
  • InjuryAug 28, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, DURING USE, THE PATIENT HAD AN ALLERGIC REACTION AROUND THE STOMA FROM THE DEVICE. THE PATIENT WAS RECANNULATED.

    Patient problems: Extubate; Skin Inflammation/ Irritation; Unintended Extubation

    MAUDE report 10465489 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122531, K182861. One submission often covers many models, so check the product and lot details in each record.

  • Z-1951-2015 · May 8, 2015 · Terminated
    Shiley Neonatal, Pediatric and Long Pediatric Tracheostomy Tube with TaperGuard Cuff. Intended for use in providing tracheal access for airway managements.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates