Kangaroo 775759Recalls & reported problems
Cardinal Health, Inc. · 924 Safety Screw Spike Set · Class 2
Quick summary
This Cardinal Health Kangaroo enteral feeding set has 40 adverse event reports. They mostly describe leaking or splashing feeding solution from the spike connection or tubing, sometimes with tubes disconnecting, cracking, or coming loose, with no reported patient harm. There were no deaths and no injuries among the 40 malfunction reports. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 40
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 20884521155739), received between Jun 15, 2017 and Oct 16, 2019 — classified as 40 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (22 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak22
- Leak/Splash17
- Crack1
- Detachment of Device or Device Component1
- Disconnection1
- Material Puncture/Hole1
Patient problems
- No Consequences Or Impact To Patient21
- No Known Impact Or Consequence To Patient13
- No Patient Involvement7
Recent adverse event reports
The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE NURSE HEARD A DRIPPING SOUND AND FOUND THAT THE TUBE HAD FALLEN OUT OF THE BAG. ADDITIONAL INFORMATION RECEIVED FROM THE INITIAL REPORTER STATED THAT THE FEEDING SET TUBE BECAME DISCONNECTED FROM THE ADAPTER THAT ATTACHES TO THE BAG.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9196856 ↗MalfunctionOct 4, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THE FEEDING SET WAS FILTERING AND POLLUTING WHICH WAS NOTICED DURING USE. IT WAS FURTHER CLARIFIED THAT THE FEEDING SET WAS LEAKING AND THE FOOD WAS GETTING ON THE OUTSIDE OF THE FEEDING SET.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9156653 ↗MalfunctionOct 4, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT A FAILURE WAS NOTICED DURING USE. WHEN THE PUNCH WAS CONNECTED TO THE NUTRITION, THE PURPLE CONNECTOR IMMEDIATELY BEGINS TO FILTER THE NUTRITION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9156257 ↗MalfunctionOct 4, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THE CONTENT FILTRATED BETWEEN THE HOSE CONNECTION AND THE PROXIMAL PORT OF THE NUTRITION. THE CUSTOMER FURTHER CLARIFIED THAT THE FEEDING SET WAS LEAKING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9155762 ↗MalfunctionOct 3, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THE EQUIPMENT WAS INSTALLED IN A GLYCERNA BOTTLE AND WAS OBSERVED TO LEAK. THE CUSTOMER REPORTED THAT THE FAULT OCCURRED AT THE THREADED SPIKE CONNECTION TO THE FEEDING BOTTLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9153730 ↗MalfunctionOct 3, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTS THAT NUTRITION LEAKS INTO THE TOP PACKAGING. THE CUSTOMER FURTHER REPORTED THAT THE FAULT OCCURS AT THE THREADED SPIKE CONNECTION WHEN THE FEED IS PASSING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9153568 ↗MalfunctionOct 3, 2019Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
THE CUSTOMER REPORTED WHEN THEY OPENED THE PUNCH DEVICE, THEY FOUND A FISSURE THROUGH THE CONNECTION PART AND A LEAKAGE WAS DETECTED DURING USE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9152532 ↗MalfunctionOct 3, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE THREADED PUNCH CONNECTION IS LEAKING THE NUTRITION. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9150729 ↗MalfunctionOct 3, 2019Leak/Splash; Leak/Splash
THE CUSTOMER REPORTED THAT THE THREADED PUNCH CONNECTION WHERE THE FEED PASSES LEAKED WHEN THE FEED BEGINS TO FILTER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9150711 ↗MalfunctionOct 3, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE FEED IS LEAKING THE NUTRITION. ANOTHER SET WAS REQUESTED AND THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9150592 ↗MalfunctionOct 3, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE THREADED PUNCH CONNECTION IS LEAKING THE NUTRITION. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9150374 ↗MalfunctionOct 2, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THERE WAS A NUTRITION LEAK AT THE PURPLE CONNECTOR.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9148135 ↗MalfunctionOct 2, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THERE WAS A LEAKAGE OF NUTRITION AT THE SPIKE CONNECTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9147910 ↗MalfunctionOct 2, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THE THREADED PUNCH CONNECTION IS LEAKING AT THE SCREW ADAPTER. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9147679 ↗MalfunctionOct 2, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THE THREADED PUNCH CONNECTION IS LEAKING WHEN THE FEED IS PASSING. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9147514 ↗MalfunctionSep 30, 2019Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component
THE CUSTOMER REPORTED THE THREADED PUNCH CONNECTION IS LEAKING AND DETACHING WHEN THE NUTRITION IS FLOWING MAKING THE NUTRITION LEAK. THIS IS OCCURRING AT THE CONNECTION OF THE SET TO THE FEEDING BOTTLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9134608 ↗MalfunctionSep 30, 2019Leak/Splash; Leak/Splash
THE CUSTOMER REPORTED THAT THE THREADED PUNCH CONNECTION WHERE THE FEED PASSES LEAKED WHEN THE FEED BEGINS TO FILTER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9134242 ↗MalfunctionSep 27, 2019Leak/Splash; Leak/Splash
THE CUSTOMER REPORTED THAT THE THREADED PUNCH CONNECTION WHERE THE FEED PASSES LEAKED WHEN THE FEED BEGINS TO FILTER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9131959 ↗MalfunctionSep 27, 2019Leak/Splash; Leak/Splash
THE CUSTOMER REPORTED THAT THE THREADED PUNCH CONNECTION WHERE THE FEED PASSES LEAKED WHEN THE FEED BEGINS TO FILTER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9131520 ↗MalfunctionSep 24, 2019Leak/Splash; Leak/Splash
THE CUSTOMER REPORTS THE FEEDING TUBES ARE LEAKING AT SPIKE CONNECTION, OCCURRING WITH MULTIPLE SETS. THE LEAK IS AT THE SPIKE PROXIMAL TO BAG.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9111464 ↗MalfunctionSep 23, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THERE WAS LEAKING WHICH CAUSED NUTRITION LOST ON THE PATIENT THAT OCCURRED IN THE THREADED SPIKE CONNECTION WHEN THE FEED WAS PASSING THAT BEGUN TO PRIME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9106332 ↗MalfunctionSep 20, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT THERE WAS A LEAKAGE OF NUTRITION HENCE AT THE SPIKE CONNECTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9100684 ↗MalfunctionSep 18, 2019Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
THE CUSTOMER REPORTED THAT THE TWO FEEDING SETS HAD HOLES AND WERE LEAKING AT THE SCREW SPIKE CONNECTOR. THIS HAPPENED DURING PRIMING. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9081729 ↗MalfunctionJun 19, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTS THAT THE FEED IS LEAKING AT THE PORT THAT CONNECTS TO TUBE FEED PORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8712416 ↗MalfunctionJun 19, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTS THAT THE SET WAS SPIKED INTO THE BAG AND THE CONNECTION WAS BROKEN CAUSING THE FLUID TO LEAK DOWN THE LINE. THE BAG AND SPIKE SET WAS REPLACED AND THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8712107 ↗
Other PUMP, INFUSION, ENTERAL devices
- Kangaroo 775100Cardinal Health, Inc. · PUMP, INFUSION, ENTERAL988 reports
- Kangaroo 382400Cardinal Health, Inc. · PUMP, INFUSION, ENTERAL677 reports
- Kangaroo 383400Cardinal Health, Inc. · Pump, Infusion, Enteral1 death659 reports
- Infinity Enteral Feeding Pump INFKIT2Zevex International, Inc. · Pump, Infusion, Enteral2 deaths407 reports
- Infinity Enteral Feeding Pump INFP01Zevex International, Inc. · Pump, Infusion, Enteral1 death207 reports
- Perfusor® 8713030UB. Braun Medical Inc. · Pump, infusion, enteral2 recalls3 deaths76 reports
- Kangaroo 775659Cardinal Health, Inc. · PUMP, INFUSION, ENTERAL1 death139 reports
- Perfusor® 8713031UB. Braun Medical Inc. · PUMP, INFUSION, ENTERAL1 recall21 deaths87 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.