Shiley 6.5PELF (UDI 20884521779904)Recalls & reported problems
Covidien LP · Pediatric Tracheostomy Tube Long, Cuffless · Class 2
All models of Shiley (387)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 20884521779904), received between Nov 10, 2021 and Dec 12, 2023 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (5 reports).
1 other recall cover devices cleared under the same FDA submission (K122531, K182861); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking5
- Use of Device Problem3
- Defective Component2
- Defective Device1
Patient problems
- Insufficient Information5
- No Clinical Signs, Symptoms or Conditions4
- Low Oxygen Saturation1
- Unintended Extubation1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2023Physical Resistance/Sticking; Physical Resistance/Sticking
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S OBTURATOR WAS HARD AND DIFFICULT TO REMOVE BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18313894 ↗MalfunctionDec 11, 2023Defective Component; Defective Device; Defective Component; Defective Device
ACCORDING TO THE REPORTER, THE "OPTULATOR" OR THE STYLET OF THE TRACHEOSTOMY TUBE FITTED TOO TIGHTLY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18302555 ↗MalfunctionAug 7, 2023Use of Device Problem; Defective Component; Use of Device Problem; Defective Component
ACCORDING TO THE REPORTER, THE ANGLE OF THE TRACHEOSTOMY TUBE WAS TIGHT, AND THERE WAS RESISTANCE WHEN INSERTING AND REMOVING THE OBTURATOR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17477163 ↗MalfunctionOct 31, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
ACCORDING TO THE REPORTER, WHEN CHECKING THE TRACHEOSTOMY TUBE PRIOR TO THE PLANNED TUBE REPLACEMENT, THE INTRODUCER WAS VERY DIFFICULT TO REMOVE. TRIED A SECOND TUBE FROM THE SAME LOT NUMBER, BUT IT WAS ALSO DIFFICULT TO MOVE. HOWEVER, ABLE TO MANAGE TO SAFELY AND SUCCESSFULLY PERFORM THE TUBE CHANGE. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15705123 ↗MalfunctionAug 5, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THERE WAS SIGNIFICANT DIFFICULTY IN REMOVING THE "INTRODUCER" FROM THE TUBE PRIOR TO THE TUBE CHANGE. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Insufficient Information
MAUDE report 15180177 ↗MalfunctionJul 27, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
ACCORDING TO THE REPORTER, DURING USE, THERE WAS A SIGNIFICANT DIFFICULTY IN REMOVING THE "INTRODUCER" FROM THE TUBE PRIOR TO THE TUBE CHANGE.
Patient problems: Insufficient Information
MAUDE report 15117327 ↗MalfunctionJul 27, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
ACCORDING TO THE REPORTER, DURING USE, THERE WAS A SIGNIFICANT DIFFICULTY IN REMOVING THE ""INTRODUCER"" FROM THE TUBE PRIOR TO THE TUBE CHANGE.
Patient problems: Insufficient Information
MAUDE report 15113074 ↗MalfunctionJul 27, 2022Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING USE, THERE WAS A SIGNIFICANT DIFFICULTIES IN REMOVING THE INTRODUCER FROM THE TUBE PRIOR TO THE TUBE CHANGE.
Patient problems: Insufficient Information
MAUDE report 15112621 ↗InjuryNov 10, 2021Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE DEVICE'S FLANGE EYELET HAD A BREAKAGE. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Insufficient Information; Low Oxygen Saturation; Unintended Extubation
MAUDE report 12786920 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122531, K182861. One submission often covers many models, so check the product and lot details in each record.
- Z-1951-2015 · May 8, 2015 · TerminatedShiley Neonatal, Pediatric and Long Pediatric Tracheostomy Tube with TaperGuard Cuff. Intended for use in providing tracheal access for airway managements.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page