Shiley 6.5PLCF (UDI 20884521779942)Recalls & reported problems

Covidien LP · Pediatric Tracheostomy Tube Long with TaperGuard Cuff · Class 2

Quick summary

This Shiley 6.5PLCF is a long pediatric tracheostomy tube with a tapered cuff, made by Covidien. The 43 reports mostly describe the cuff losing air so the tube's seal fails, and the tube coming out of the patient unexpectedly, sometimes after the neck flange or inflation line breaks or splits. Seven injuries were reported and no deaths. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 20884521779942), received between Mar 2, 2022 and Mar 9, 2026 — classified as 7 injuries, 35 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 4 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem27
  • Gas/Air Leak13
  • Material Split, Cut or Torn2
  • Human-Device Interface Problem1
  • Manufacturing, Packaging or Shipping Problem1
  • Material Separation1
  • Patient-Device Incompatibility1

Patient problems

  • Unintended Extubation24
  • No Clinical Signs, Symptoms or Conditions17
  • Hypoventilation4
  • Insufficient Information4
  • Low Oxygen Saturation4
  • Choking3
  • Discomfort1
  • Sleep Dysfunction1

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 9, 2026Gas/Air Leak

    IT WAS REPORTED THAT THERE WAS AN AIR LEAKAGE FROM THE CUFF AFTER TWO DAYS OF USE BY THE PATIENT, AND THE TUBE WAS REMOVED OR CAME OUT FROM THE PATIENT. FROM THE INFLATION LINE, AIR WAS TRIED TO INSERT, BUT THE AIR DID NOT GO IN. REPLACEMENT OF THE TUBE WAS DONE.

    Patient problems: Unintended Extubation

    MAUDE report 24551496 ↗
  • MalfunctionDec 12, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING PRE-USAGE TESTING, THE TRACH TUBE'S CUFF HAD AN AIR LEAK. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23794695 ↗
  • MalfunctionAug 19, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT PRIOR TO USAGE, A HOLE WAS FOUND FROM THE BASE OF THE TRACH TUBE'S INFLATION LINE, AS ILLUSTRATED IN THE ATTACHMENT. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22844258 ↗
  • MalfunctionJul 2, 2025Human-Device Interface Problem; Human-Device Interface Problem

    IT WAS REPORTED THAT THE INSERTION AND REMOVAL OF THE OBTURATOR WERE HARD PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22391561 ↗
  • MalfunctionMar 11, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT DURING USE, THE TRACHEOSTOMY TUBE WAS DISLODGED FROM A PATIENT AT HOME. IT WAS UNCERTAIN IF IT WAS OUTWARD APPEARANCE OR DAMAGE TO THE CUFF. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 21575083 ↗
  • MalfunctionJan 10, 2025Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, DURING USE, THE CUFF DID NOT INFLATE EVEN WHEN AIR WAS ADDED. THE TUBE WAS REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation

    MAUDE report 21132269 ↗
  • MalfunctionOct 30, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S BALLOON BURST AFTER 48 HOURS OF CHANGED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation

    MAUDE report 20569154 ↗
  • MalfunctionOct 18, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF HAD PROBLEM WITH INFLATING DURING THE INFLATION TEST PRIOR TO THE PLACEMENT OF THE CANNULA. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20482315 ↗
  • MalfunctionOct 18, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF HAD PROBLEM WITH INFLATING DURING THE INFLATION TEST PRIOR TO THE PLACEMENT OF THE CANNULA. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20482269 ↗
  • InjurySep 3, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, DURING USE, THE TRACHEOSTOMY TUBE'S PILOT VALVE WAS LEAKING THROUGH THE VENTILATOR CIRCUIT AND WAS REPLACED THREE TIMES WITHIN TWO WEEKS. THE PATIENT EXPERIENCED CHOKING, SALIVATION, AND SATURATION DROPPED.

    Patient problems: Choking; Low Oxygen Saturation; Unintended Extubation

    MAUDE report 20127920 ↗
  • InjurySep 3, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, DURING USE, THE TRACHEOSTOMY TUBE'S PILOT VALVE WAS LEAKING THROUGH THE VENTILATOR CIRCUIT AND WAS REPLACED THREE TIMES WITHIN TWO WEEKS. THE PATIENT EXPERIENCED CHOKING, SALIVATION, AND SATURATION DROPPED.

    Patient problems: Choking; Low Oxygen Saturation; Unintended Extubation

    MAUDE report 20127858 ↗
  • InjurySep 3, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, DURING USE, THE TRACHEOSTOMY TUBE'S PILOT VALVE WAS LEAKING THROUGH THE VENTILATOR CIRCUIT AND WAS REPLACED THREE TIMES WITHIN TWO WEEKS. THE PATIENT EXPERIENCED CHOKING, SALIVATION, AND SATURATION DROPPED.

    Patient problems: Choking; Low Oxygen Saturation; Unintended Extubation

    MAUDE report 20127849 ↗
  • MalfunctionMar 15, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S INFLATION LINE BROKE INTO PIECES AFTER IT WAS PULLED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18917783 ↗
  • MalfunctionMar 6, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE CONNECTION PART WITH THE TUBE OF THE INFLATION LINE WAS WOBBLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18852545 ↗
  • InjuryJan 12, 2024Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE CUFF OF THE FIRST TRACHEOSTOMY TUBE HAD AN AIR LEAK. AN EMERGENCY RESPONSE WAS PROVIDED BY INJECTING AIR. EVEN WHEN AIR WAS INJECTED, THE CUFF PRESSURE DID NOT GO UP, SO THE PATIENT WAS TAKEN FOR AN EMERGENCY MEDICAL CONSULTATION AND TRACHEOSTOMY TUBE REPLACEMENT WAS PERFORMED. ON THE SECOND DEVICE, IT WAS STATED THAT THERE WERE SEVERAL EMERGENCY VISITS IN THE FACILITY DUE TO CUFF LEAKS, AND THE PREVIOUS DEFECTIVE PRODUCT WAS DAMAGED ON THE INFLATOR LINE PART. ALSO, THERE WERE TIMES WHEN THE PATIENT PLAYED WITH THE LINES, SO NOW THE INFLATOR LINES ARE TAPED AND FIXED TOGETHER. A CUFF LEAK WAS DISCOVERED WHEN A SOUND WAS HEARD LEAKING THROUGH THE VOCAL CORDS. RESPIRATORY MANAGEMENT WAS PERFORMED.

    Patient problems: Hypoventilation; Unintended Extubation

    MAUDE report 18508826 ↗
  • InjuryJan 12, 2024Use of Device Problem; Gas/Air Leak; Use of Device Problem; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE CUFF OF THE FIRST TRACHEOSTOMY TUBE HAD AN AIR LEAK. AN EMERGENCY RESPONSE WAS PROVIDED BY INJECTING AIR. EVEN WHEN AIR WAS INJECTED, THE CUFF PRESSURE DID NOT GO UP, SO THE PATIENT WAS TAKEN FOR AN EMERGENCY MEDICAL CONSULTATION AND TRACHEOSTOMY TUBE REPLACEMENT WAS PERFORMED. ON THE SECOND DEVICE, IT WAS STATED THAT THERE WERE SEVERAL EMERGENCY VISITS IN THE FACILITY DUE TO CUFF LEAKS, AND THE PREVIOUS DEFECTIVE PRODUCT WAS DAMAGED ON THE INFLATOR LINE PART. ALSO, THERE WERE TIMES WHEN THE PATIENT PLAYED WITH THE LINES, SO NOW THE INFLATOR LINES ARE TAPED AND FIXED TOGETHER. A CUFF LEAK WAS DISCOVERED WHEN A SOUND WAS HEARD LEAKING THROUGH THE VOCAL CORDS. RESPIRATORY MANAGEMENT WAS PERFORMED.

    Patient problems: Hypoventilation; Unintended Extubation

    MAUDE report 18508820 ↗
  • MalfunctionJan 12, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE CUFF OF THE TRACHEOSTOMY TUBE DEFLATED EVEN AFTER AIR WAS INSERTED DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18506573 ↗
  • MalfunctionJan 9, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF "BURST" SPONTANEOUSLY THREE TO FOUR DAYS DURING USE. AFTER THE CUFF "BURST," MORE AND MORE SECRETIONS BEGAN TO APPEAR, WHICH WENT DOWN THE PATIENT'S THROAT INSTEAD OF THROUGH THE TUBE. THE PATIENT WAS VERY TIRED AND COULD NOT SLEEP AT NIGHT BECAUSE SECRETIONS KEPT APPEARING. THE PATIENT'S TUBE WAS REPLACED, AND HAD BEEN USING CUFFED TUBES FOR SOME TIME FROM A DIFFERENT MANUFACTURER.

    Patient problems: Sleep Dysfunction; Unintended Extubation

    MAUDE report 18472984 ↗
  • MalfunctionDec 21, 2023Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF WAS LEAKING AIR EVEN RE-INJECTING AIR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18373788 ↗
  • MalfunctionDec 8, 2023Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF HAD AN AIR LEAK AFTER SEVERAL DAYS OF PATIENT USAGE. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 18296939 ↗
  • MalfunctionNov 16, 2023Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF WAS INFLATED BUT THE CUFF PRESSURE CANNOT BE MAINTAINED DURING USE. THE PATIENT HAD A REPLACEMENT OF THE TUBE.

    Patient problems: Unintended Extubation

    MAUDE report 18150581 ↗
  • MalfunctionNov 13, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE CUFF OF THE TRACHEOSTOMY TUBE HAD A LEAK AND DID NOT INFLATE DURING TESTING. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18124878 ↗
  • MalfunctionOct 18, 2023Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE TRACH'S CUFF WAS FOUND NO PRESSURE AFTER 30 DAYS OF PATIENT USE, SAME SITUATION EVEN WHEN AIR WAS REINSERTED. THE PATIENT HAD A REPLACEMENT OF THE TUBE.

    Patient problems: Unintended Extubation

    MAUDE report 17961792 ↗
  • MalfunctionOct 4, 2023Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE TRACH TUBE'S CUFF WAS NOT INFLATING AFTER A MONTH OF PATIENT USAGE. THE PATIENT HAD A REPLACEMENT OF THE TUBE.

    Patient problems: Unintended Extubation

    MAUDE report 17873038 ↗
  • MalfunctionSep 21, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THERE WAS A CRACK FOUND IN THE TRACHEOSTOMY TUBE'S INFLATION LINE THAT CAUSED THE DEFLATION OF THE CUFF AFTER A FEW DAYS OF PATIENT USE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17788914 ↗

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.