Shiley 6.5PLCRecalls & reported problems
Covidien LP · Pediatric Tracheostomy Tube Long Cuffed · Class 2
Quick summary
This Shiley pediatric long cuffed tracheostomy tube is placed in the airway to help children breathe. Reports mostly describe the cuff not inflating or losing air during use, with the tube then removed and replaced, and some mention unusual shape or fit. The records include 24 reports with no deaths and 10 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 20884522029060), received between Aug 10, 2017 and Apr 15, 2024 — classified as 10 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (9 reports).
8 other recalls cover devices cleared under the same FDA submission (K955680); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Inflation Problem9
- Gas/Air Leak5
- Use of Device Problem5
- Defective Component2
- Human-Device Interface Problem1
- Leak/Splash1
- Obstruction of Flow1
Patient problems
- Extubate9
- Unintended Extubation6
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient3
- No Information2
- No Patient Involvement2
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 15, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF PRESSURE WAS UNCONTROLLABLE DURING USE. THE TRACH TUBE WAS REMOVED AND REPLACED.
Patient problems: Unintended Extubation
MAUDE report 19105781 ↗MalfunctionFeb 2, 2023Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE PILOT BALLOON WOULD NOT INFLATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16284895 ↗MalfunctionJun 9, 2022Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE TRACH'S BALLOON (CUFF) SEAL LOSES AIR AFTER 1 OR 2 DAYS OF USE AND MUST BE REPLACED. THE PATIENT WAS ON VENTILATOR 12 HOURS, OVERNIGHT. THE CUFF WAS TESTED PRIOR TO USE. THE CUFF INFLATED BUT LOST PRESSURE. IT WAS NOTED THAT THE CUFF WALLS FEEL THINNER. THE PATIENT'S AIRWAY ANATOMY WAS NOT UNUSUAL AND HAD BEEN USING THE SAME TRACH FOR 20 YEARS. THE TUBE WAS NOT LUBRICATED BEFORE INSERTION. THE TRACH TUBES WAS CLEANED WITH SOAP AND WATER. THERE WAS NO PHYSICAL/WATER DAMAGE.
Patient problems: Unintended Extubation
MAUDE report 14648940 ↗MalfunctionJun 9, 2022Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE DEVICE CUFF WOULD NOT HOLD AIR AND CUFF WALLS FELT THINNER. IT WORKED INITIALLY BUT AFTER 1 OR 2 DAYS THE CUFF NO LONGER HOLD THE AIR AND MUST BE REPLACED. PATIENT WAS RECANNULATED.
Patient problems: Unintended Extubation
MAUDE report 14647142 ↗MalfunctionJun 9, 2022Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE DEVICE CUFF WOULD NOT HOLD AIR AND CUFF WALLS FELT THINNER. IT WORKED INITIALLY BUT AFTER 1 OR 2 DAYS THE CUFF NO LONGER HOLD THE AIR AND MUST BE REPLACED. PATIENT WAS RECANNULATED.
Patient problems: Unintended Extubation
MAUDE report 14647019 ↗MalfunctionDec 6, 2021Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING THE CHECK OF THE CUFF, THE CUFF WOULD NOT INFLATE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12933219 ↗MalfunctionJan 4, 2021Defective Component; Defective Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE ANGLE OF THE CANNULA WAS DIFFERENT AND IT HURT THE PATIENT. RECANNULATION WAS REQUIRED.
Patient problems: Unintended Extubation
MAUDE report 11108710 ↗InjuryDec 1, 2020Defective Component; Defective Component
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE'S SHAPE AND CURVE WAS NOT USUAL. THE PATIENT REQUIRED RECANNULATION.
Patient problems: Unintended Extubation
MAUDE report 10927284 ↗MalfunctionOct 26, 2020Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING USE, WHEN AIR WAS INJECTED TO THE DEVICE, THE CUFF WITHERED. AIR WAS INJECTED SEVERAL TIMES ABOUT EVERY 2 HOURS, BUT AIR WOULD COME OUT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10732915 ↗MalfunctionSep 30, 2020Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE UNIT CUFF DID NOT INFLATE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10609957 ↗MalfunctionJul 28, 2020Obstruction of Flow; Obstruction of Flow
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE HAD A SUCTION TUBE THAT COULD GET NOTHING THROUGH. THERE WAS NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10331298 ↗InjuryOct 18, 2019Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, ONE HOUR AFTER THE START OF USING THE PRODUCT, THE DEVICE HAD AN AIR LEAK FROM THE CUFF. IT WAS REPORTED THAT RECANNULATION OF A REPLACEMENT TUBE WAS REQUIRED.
Patient problems: Extubate
MAUDE report 9206505 ↗MalfunctionJul 30, 2019Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE CUFF DID NOT INFLATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8842806 ↗InjuryJun 14, 2019Leak/Splash; Leak/Splash
ACCORDING TO THE REPORTER, DURING HOSPITALIZATION, THE DEVICE HAD AN AIR LEAKAGE AND WAS NOTICED DUE TO AIR LEAKING SOUND AND THE PILOT BALLOON SEEMED IMMEDIATELY DEFLATED ALTHOUGH THERE WAS AN AIR INJECTED. THE CUSTOMER INDICATED THAT THERE WAS A REPLACEMENT OF THE TUBE DONE TO THE PATIENT.
Patient problems: Extubate
MAUDE report 8701186 ↗MalfunctionFeb 20, 2019Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, THE DEVICE COULD NOT MAINTAIN FULL INFLATION. THERE WAS NO REPORTED PATIENT INVOLVEMENT AND OUTCOME.
Patient problems: No Information
MAUDE report 8354216 ↗MalfunctionFeb 20, 2019Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, THE DEVICE COULD NOT MAINTAINED FULL INFLATION. THERE WAS NO REPORTED PATIENT INVOLVEMENT AND OUTCOME.
Patient problems: No Information
MAUDE report 8353732 ↗InjuryNov 27, 2018Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE'S CUFF COULD NOT HOLD AIR AND SHRUNK QUICKLY ALTHOUGH THE AIR WENT IN. IT WAS ALSO REPORTED THAT WHEN THE CUFF WAS DEGAS AND CONTINUOUSLY APPLY NEGATIVE PRESSURE, IT WAS INFLATED ADVERSELY. IT WAS INDICATED THAT THE PATIENT HAD A REPLACEMENT OF THE TUBE.
Patient problems: Extubate
MAUDE report 8107880 ↗InjuryJun 26, 2018Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, WHEN THE DEVICE WAS CHECKED BY CUFF PRESSURE GAUGE, THE AIR DID NOT ENTER THE CUFF. IT WAS REPORTED THAT AFTER EXTUBATION, IT WAS CONFIRMED THAT THE AIR ENTERED THE CUFF.
Patient problems: Extubate
MAUDE report 7637703 ↗InjuryApr 3, 2018Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, DURING USE, THE CUFF WAS NOTED TO BE DEFLATING AND THE TUBE AT THE NECK END APPEARED DIFFERENT. IT WAS RE PORTED THAT THE TUBE WAS REPLACED, BUT THE REPLACEMENT TUBE¿S CUFF DID NOT MAINTAIN FULL INFLATION AND HAD AIR LEAKS NEAR THE CUFF VALVE.
Patient problems: Extubate
MAUDE report 7391811 ↗InjuryMar 28, 2018Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, DURING USE, THE CUFF WAS NOTED TO BE DEFLATING AND THE TUBE AT THE NECK END APPEARED DIFFERENT. IT WAS RE PORTED THAT THE TUBE WAS REPLACED, BUT THE REPLACEMENT TUBE¿S CUFF DID NOT MAINTAIN FULL INFLATION AND HAD AIR LEAKS NEAR THE CUFF VALVE.
Patient problems: Extubate
MAUDE report 7379174 ↗InjuryMar 26, 2018Human-Device Interface Problem; Human-Device Interface Problem
ACCORDING TO THE REPORTER, THE TAPE WAS PULLED OFF THE CUFF INFLATION DEVICE AND IT SNAPPED OFF COMPLETELY. THEY QUICKLY PERFORMED AN EMERGENCY TRACHEOSTOMY CHANGE AND IT WAS SUCCESSFUL. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Extubate
MAUDE report 7369896 ↗InjuryFeb 9, 2018Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, THE CUFF WAS DEFLATED SO AN EMERGENCY TRACHEOSTOMY CHANGE WAS PERFORMED. IT WAS REPORTED THAT THE DEFECTIVE PRODUCT IS UNAVAILABLE.
Patient problems: Extubate
MAUDE report 7260731 ↗InjuryFeb 9, 2018Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, THE PATIENT WAS DISTRESSED AND COUGHING, IT WAS REALIZED THAT THE CUFF WAS DEFLATED. IT WAS REPORTED THAT THE PATIENT¿S MOTHER PERFORMED AN EMERGENCY TRACHEOSTOMY CHANGE USING A SPARE TUBE OF THE SAME SIZE. IT WAS REPORTED THAT THE DEFECTIVE PRODUCT IS UNAVAILABLE.
Patient problems: Extubate
MAUDE report 7260521 ↗MalfunctionAug 10, 2017Inflation Problem; Inflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT PRIOR TO USE ON A PATIENT, AIR LEAK WAS FOUND. CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6787087 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K955680. One submission often covers many models, so check the product and lot details in each record.
- Z-1493-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 4.5 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a
- Z-1496-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 5.0 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up
- Z-1492-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 4.0 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a
- Z-1499-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 6.5 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up
- Z-1494-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 5.0 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a
- Z-1497-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 5.5 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up
- Z-1495-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 5.5 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a
- Z-1498-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 6.0 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.