Shiley 4.5PDCRecalls & reported problems
Covidien LP · Pediatric Tracheostomy Tube Cuffed · Class 2
Quick summary
This Shiley 4.5PDC is a cuffed pediatric tracheostomy tube made by Covidien. The 31 reports mostly describe the cuff failing to inflate or deflate properly and leaking air, along with connectors coming loose and some discoloration; several patients came out of the tube and needed it replaced or reintubated. There are 13 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 20884522029114), received between Jan 10, 2018 and Jun 14, 2023 — classified as 13 injuries, 18 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (10 reports).
8 other recalls cover devices cleared under the same FDA submission (K955680); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Use of Device Problem10
- Gas/Air Leak6
- Inflation Problem6
- Material Discolored4
- Connection Problem2
- Contamination1
- Contamination /Decontamination Problem1
- Contamination of Device Ingredient or Reagent1
Patient problems
- Extubate13
- No Clinical Signs, Symptoms or Conditions9
- Unintended Extubation8
- No Consequences Or Impact To Patient3
- No Patient Involvement2
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 14, 2023Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE CUFF OF THE TRACH CANNULA WAS NO LONGER INFLATING WITHIN 30 DAYS OF USE. THE PATIENT HAD A REPLACEMENT OF THE TUBE.
Patient problems: Unintended Extubation
MAUDE report 17131394 ↗MalfunctionNov 10, 2022Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE'S CUFF HAD A LEAK. RECANNULATION WAS DONE.
Patient problems: Unintended Extubation
MAUDE report 15772149 ↗MalfunctionMay 7, 2022Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING PRE-CHECK TEST, IT WAS FOUND THAT THE CUFF WILL NOT HOLD AIR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14317410 ↗MalfunctionApr 15, 2022Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE INFLATION LINE OF THE DEVICE WAS FOUND TO BE CURVED AND THE CUFF WOULD NOT INFLATE OR DEFLATE BECAUSE AIR COULD NOT BE INJECTED WITH SYRINGE. THE SALES REP CHECKED IT AND AIR COULD BE INJECTED NORMALLY. THE DEVICE HAS BEEN REPLACED. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14118626 ↗MalfunctionFeb 25, 2022Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, IT WAS EASIER TO COME OFF FROM THE CONNECTOR THAN USUAL WHEN CONNECTED TO THE CATHETER MOUNT. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation
MAUDE report 13603488 ↗MalfunctionFeb 25, 2022Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, IT WAS EASIER TO COME OFF FROM THE CONNECTOR THAN USUAL WHEN CONNECTED TO THE CATHETER MOUNT. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Extubation
MAUDE report 13603442 ↗MalfunctionAug 31, 2021Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE HAD AN AIR LEAK FROM THE CONNECTION PORT OF THE CUFF AND THE INFLATION LINE. THERE WAS NO PROBLEM IN THE PRE-USE INSPECTION, BUT AFTER SEVERAL DAYS OF USE, THE CUFF DID NOT DEFLATE PROPERLY, AND IT WAS CONFIRMED. THE PATIENT REQUIRED AN EXCHANGE AT THE EMERGENCY OUTPATIENT DEPARTMENT. THE SAME EVENT OCCURRED TWICE.
Patient problems: Unintended Extubation
MAUDE report 12399867 ↗MalfunctionAug 31, 2021Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE HAD AN AIR LEAK FROM THE CONNECTION PORT OF THE CUFF AND THE INFLATION LINE. THERE WAS NO PROBLEM IN THE PRE-USE INSPECTION, BUT AFTER SEVERAL DAYS OF USE, THE CUFF DID NOT DEFLATE PROPERLY, AND IT WAS CONFIRMED. THE PATIENT REQUIRED AN EXCHANGE AT THE EMERGENCY OUTPATIENT DEPARTMENT. THE SAME EVENT OCCURRED TWICE.
Patient problems: Unintended Extubation
MAUDE report 12393842 ↗MalfunctionJul 23, 2021Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE DEVICE HAD A LEAK FROM THE INFLATION LINE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12217305 ↗MalfunctionJul 23, 2021Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, AIR LEAKAGE OCCURRED FROM THE CUFF ABOUT 10 DAYS AFTER USE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12217052 ↗MalfunctionJul 23, 2021Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, AIR LEAKAGE OCCURRED FROM THE CUFF ABOUT 10 DAYS AFTER USE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12217040 ↗MalfunctionJul 23, 2021Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE DEVICE HAD A LEAK FROM THE INFLATION LINE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12217001 ↗MalfunctionApr 15, 2021Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE DEVICE HAD A CUFF INFLATION/DEFLATION ISSUE. THE CUFF WOULD ONLY INFLATE WHEN THE SYRINGE WAS CONNECTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11675619 ↗InjuryOct 31, 2020Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE DEVICE WAS DISCONNECTED IN THE INFLATION LINE. THE INFLATION LINE GOT DISCONNECTED JUST BY PULLING A LITTLE IMMEDIATELY AFTER THE TUBE EXCHANGE. THE PATIENT WAS RECANNULATED.
Patient problems: Extubate; Unintended Extubation
MAUDE report 10767495 ↗InjuryOct 31, 2020Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, TWO WEEKS AFTER PLACEMENT, THE DEVICE HAD AIR LEAK. THE DEVICE WAS REPLACED. UPON CHECKING THE PRODUCT, CUFF LEAKAGE WAS FOUND BUT THE LEAKAGE LOCATION COULD NOT BE IDENTIFIED. THE PATIENT WAS REINTUBATED.
Patient problems: Extubate; Unintended Extubation
MAUDE report 10767478 ↗InjuryJul 31, 2020Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE'S INFLATION LINE CAME OFF FROM THE CONNECTION PART OF THE CANNULA. THE PATIENT HAD A REPLACEMENT OF THE TUBE.
Patient problems: Extubate
MAUDE report 10352829 ↗InjuryApr 20, 2020Material Discolored; Material Discolored
ACCORDING TO THE REPORTER, 14 DAYS AFTER THE START OF USING, THE DEVICE¿S TUBE TIP WAS COLORED DARK THAT LOOKED LIKE A MOLD AFTER THE TRACHEOSTOMY CANNULA REPLACEMENT AND IT WAS OCCURRED 2 CASES CONSECUTIVELY. THERE WAS NO PATIENT INJURY.
Patient problems: Extubate
MAUDE report 9979455 ↗InjuryMar 26, 2020Material Discolored; Material Discolored
ACCORDING TO THE REPORTER, 14 DAYS AFTER THE START OF USING, THE DEVICE¿S TUBE TIP WAS COLORED DARK THAT LOOKED LIKE A MOLD AFTER THE TRACHEOSTOMY CANNULA REPLACEMENT AND IT WAS OCCURRED 2 CASES CONSECUTIVELY. THERE WAS NO PATIENT INJURY.
Patient problems: Extubate
MAUDE report 9886667 ↗InjuryMar 26, 2020Material Discolored; Material Discolored
ACCORDING TO THE REPORTER, 14 DAYS AFTER THE START OF USING, THE DEVICE¿S TUBE TIP WAS COLORED DARK THAT LOOKED LIKE A MOLD AFTER THE TRACHEOSTOMY CANNULA REPLACEMENT AND IT WAS OCCURRED 2 CASES CONSECUTIVELY. THERE WAS NO PATIENT INJURY.
Patient problems: Extubate
MAUDE report 9886637 ↗InjuryNov 26, 2019Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER DURING USE, THE DEVICE¿S INFLATION LINE WAS DISCONNECTED FROM THE TUBE AROUND 20 DAYS AFTER EXCHANGE. THE PATIENT WAS REINTUBATED FOR PERIODIC EXCHANGE TO SOLVE THE PROBLEM.
Patient problems: Extubate
MAUDE report 9380215 ↗MalfunctionOct 14, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem
ACCORDING TO THE REPORTER, DURING DEVICE REPLACEMENT, A BLACK STAIN WAS FOUND TO BE ADHERED TO THE CONNECTOR PART OF THE EXTUBATED TUBE. IT WAS INDICATED THAT IT WAS A MOLD LIKE STAIN WAS FOUND TO BE ADHERED. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9186626 ↗InjurySep 26, 2019Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE DEVICE HAD AN AIR LEAK. IT WAS INDICATED THAT THE EVENT OCCURRED 30 DAYS AFTER OF USING THE DEVICE. THE PATIENT WAS REINTUBATED.
Patient problems: Extubate
MAUDE report 9122883 ↗MalfunctionApr 18, 2019Material Discolored; Material Discolored
ACCORDING TO THE REPORTER, THE INSIDE OF THE SLIP JOINT HAD TURNED BLUE. THERE WAS NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8527161 ↗InjuryJan 7, 2019Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, THE DEVICE AIR LEAKAGE BECAME NOTICEABLE FROM ABOUT ONE WEEK AFTER THE REPLACEMENT. THE SAME EVENT OCCURRED SEVERAL TIMES WITHIN SEVERAL MONTHS. THE PATIENT WAS REINTUBATED. CUSTOMER REPORTED THAT THERE WAS NO PATIENT INJURY.
Patient problems: Extubate
MAUDE report 8223543 ↗InjuryJan 7, 2019Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, DURING USE, AIR LEAKAGE WAS NOTICED ABOUT ONE WEEK AFTER THE REPLACEMENT. THE CUSTOMER REPORTED PATIENT WAS RECANNULATED.
Patient problems: Extubate
MAUDE report 8223540 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K955680. One submission often covers many models, so check the product and lot details in each record.
- Z-1497-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 5.5 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up
- Z-1495-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 5.5 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a
- Z-1499-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 6.5 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up
- Z-1492-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 4.0 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a
- Z-1494-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 5.0 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a
- Z-1493-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 4.5 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a
- Z-1496-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 5.0 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up
- Z-1498-2010 · Apr 13, 2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 6.0 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.