Skip to content

Shiley 5SCTRecalls & reported problems

Covidien LP · Adult Tracheostomy Tube Cuffed,Single Cannula · Class 2

All models of Shiley (387)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20884522031209), received between Feb 25, 2025 and Mar 3, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

23 other recalls cover devices cleared under the same FDA submission (K962173); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 3, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THERE WAS A NUB OR SMALL LUMP ON TRACH TUBE. THE DOCTOR HAD TO CUT THE NUB OR SMALL LUMP IN ORDER TO BE ABLE TO USE IT. THE NUB OR LUMP REPORTEDLY APPEARS TO BE AT THE SIX O'CLOCK POSITION ON THE TRACH. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21511411 ↗
  • MalfunctionFeb 25, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THERE WAS A NUB OR SMALL LUMP AT THE BOTTOM OF THE TRACH TUBE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21465775 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K962173. One submission often covers many models, so check the product and lot details in each record.

  • Z-2174-2012 · Jul 18, 2012 · Terminated
    Shiley Reusable Cannula Low Pressure Cuffed Tracheostomy Tubes, model numbers (Worldwide) 8LPC, 8FEN; (Outside U.S.) 8FENJ, 8FENJ-S, 8LPC-S, 8FEN-S. Product Usage: The product
  • Z-1470-2010 · Apr 13, 2010 · Terminated
    Shiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 4 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA
  • Z-1467-2010 · Apr 13, 2010 · Terminated
    Shiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 6 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa
  • Z-1480-2010 · Apr 13, 2010 · Terminated
    Shiley Low Pressure Cuffed Tracheostomy Tube, 8 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
  • Z-1463-2010 · Apr 13, 2010 · Terminated
    Shiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 6 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended f
  • Z-1466-2010 · Apr 13, 2010 · Terminated
    Shiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 4 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa
  • Z-1479-2010 · Apr 13, 2010 · Terminated
    Shiley Low Pressure Cuffed Tracheostomy Tube, 6 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
  • Z-1477-2010 · Apr 13, 2010 · Terminated
    Shiley Fenestrated Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 8 FEN-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton,
  • Z-1471-2010 · Apr 13, 2010 · Terminated
    Shiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 6 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA
  • Z-1473-2010 · Apr 13, 2010 · Terminated
    Shiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 10 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, C
  • Z-1462-2010 · Apr 13, 2010 · Terminated
    Shiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 4 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended f
  • Z-1481-2010 · Apr 13, 2010 · Terminated
    Shiley Low Pressure Cuffed Tracheostomy Tube, 10 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
  • Z-1489-2010 · Apr 13, 2010 · Terminated
    Shiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 6 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend
  • Z-1472-2010 · Apr 13, 2010 · Terminated
    Shiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 8 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA
  • Z-1490-2010 · Apr 13, 2010 · Terminated
    Shiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 8 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend

…and 8 more.

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page